Prevention of Future Deaths reports · 2016
Regulation 28 report to prevent future deaths, reference 2016-0112, written 17 Mar 2016. A coroner writes one of these when an inquest reveals a risk that could cause further deaths unless something changes.
| Date of report | 17 Mar 2016 |
|---|---|
| Reference | 2016-0112 |
| Deceased | Jacqueline Scott |
| Coroner | Angela Hodes |
| Coroner area | London Inner (West) |
| Category | Hospital Death (Clinical Procedures and medical management) related deaths |
| Source | judiciary.uk record · original PDF |
| Responses published | 2 |
Text recovered by OCR from a scanned PDF. OCR is imperfect: check anything you rely on against the source PDF. Reproduced verbatim, including the scan's own layout.
REGULATION 28: REPORT TO PREVENT FUTURE DEATHS REGULATION 28 REPORT TO PREVENT FUTURE DEATHS THIS REPORT IS BEING SENT TO: 1. Miles Scott Chief Executive St Georges University Hospitals NHS Foundation Trust Blackshaw Rd London SW17 0QT Managing Director Healthcare UK&l Phillips Healthcare Philips Centre Guildford Business Park Guildford Surrey GU2 8XH The Rt Hon Jeremy Hunt MP Secretary of State for Health Department of Health Richmond House 79 Whitehall London SW1A 2NS CORONER lam Angela Hodes, Assistant Coroner, for the coroner area of Inner West London 2 | CORONER’S LEGAL POWERS | make this report under paragraph 7, Schedule 5, of the Coroners and Justice Act 2009 and regulations 28 and 29 of the Coroners (Investigations) Regulations 2013. [HYPERLINKS] 3 | INVESTIGATION and INQUEST An investigation into the death of Jacqueline Emma Brown Scott age 55 was commenced on 10 April 2015. The investigation concluded at the end of the inquest on 2 February 2016. The conclusion of the inquest was that Mrs Scott died of natural causes on 31 March 2015 and the medical cause of death was recorded as Type II respiratory failure (treated with complications March 2015) Hypoventilation syndrome and Bronchopneumonia. Hypertensive heart disease also contributed to her death. 4 | CIRCUMSTANCES OF THE DEATH On 31 March 2015 Mrs Scott was admitted to hospital extremely unwell with worsening shortness of breath and reduced mobility. She had a history of severe sleep apnoea and morbid obesity. She was put on Philips Respironics Trilogy 202 machine BIPAP (Bilevel Positive Airway Pressure) also known as non-invasive ventilation (NIV) machine in A&E and later in the day was transferred later to the Acute Dependency Unit (ADU) on Richmond Ward for NIV and ward based care. Although the BIPAP machine was attached to the mains power socket of bed 5 in the ADU, the BIPAP was in fact running on battery as it was not known that there was a failure of power to the bed 5 power sockets and consequently to the BIPAP machine. Eventually the battery ran out and the machine stopped leaving Mrs Scott in a very vulnerable position and she died shortly thereafter. CORONER’S CONCERNS During the course of the inquest the evidence revealed matters giving rise to concern. In my opinion there is a risk that future deaths will occur unless action is taken. In the circumstances it is my statutory duty to report to you. The MATTERS OF CONCERN are as follows. — (1) The Philips Respironics Trilogy 202 (BIPAP) (i)The BIPAP Trilogy 202 machine had a subtle visual display symbol which denoted when the machine was running on battery power. That symbol is not visible if there are many alert alarms as the alarm messages fill up the screen as they come in pushing the earlier alerts (including low battery) off the screen. In Mrs Scott’s case there had been 17 alert alarms in the space of 50 minutes. (ii) The BIPAP Trilogy 202 Machine has the same alarm sound for battery depletion as for circuit disconnect (where for example the face mask slipped) which was the more usual and expected reason for an alarm and these two factors separately and together did not have any feature of urgent warning to alert staff to battery depletion. (2) Staff training (iii) Staff who were experienced and trained on the BIPAP machine did not appear to be trained to be alert to the situation or to the significance of a battery symbol showing on the machine when the machine was plugged in to the mains or to any particular alarm which denoted battery depletion rather than mask slippage. (3) | The provision of power to the ward (iv) Richmond ward ADU beds was designated as a category 4 area which in this case meant there was no isolated power supply (IPS) provided to the ward notwithstanding life-saving equipment was routinely used. (v) Hospital Technical Memoranda (HTM) 06-01 Part A provides advice and guidance and a benchmark standard for electrical installation, maintenance and safety etc in healthcare premises. It is a matter of concern that there is a conflict of advice between clause 4.22 and Clause 6.62. Clause 4.22 states: “Clinical treatment and patient safety may be compromised ( but not endangered) by any interruption of electrical supply “ whereas Clause 6.62 states: “/n clinical risk Category 4 and 5 areas the patient environment should have at least two IPS circuits at the bedhead” (vi)There was no system or check that would alert ward staff to the failure of mains power in any particular area. (4) Notification to estates management (vii) The crash bell for bed bay 5 did not work when the emergency arose. However estates management had been notified some days earlier of the broken patient call in the same bay. This was of concern as both emergency bells were on the same circuit and not fixed until 2 April 2014 when by chance the failure of electricity was identified. ACTION SHOULD BE TAKEN In my opinion action should be taken to prevent future deaths and | believe you AND/OR your organisations have the power to take such action. YOUR RESPONSE You are under a duty to respond to this report within 56 days of the date of this report, namely by 14 April 2016. I, the coroner, may extend the period. Your response must contain details of action taken or proposed to be taken, setting out the timetable for action. Otherwise you must explain why no action is proposed. COPIES and PUBLICATION | have sent a copy of my report to the Chief Coroner and to the following Interested persons: (address supplied) Hempsons Hempsons House 40 Villiers Street London WC2N 6NJ (for St Georges University Hospital NHS Foundation Trust) | have also sent it to CQC, HSE, MHRA and NHS ENGLAND who may find it useful or of interest. | am also under a duty to send the Chief Coroner a copy of your response. The Chief Coroner may publish either or both in a complete or redacted or summary form. He may send a copy of this report to any person who he believes may find it useful or of interest. You may make representations to me, the coroner, at the time of your response, about the release or the publication of your response by the Chief Coroner. 17™ March 2016 Angela Hodes Assistant Coroner Westminster Coroner’s Court, 65, Horseferry Road, London SW1P 2ED
2 responses published against this report on judiciary.uk. A response is a body's written reply to the coroner's concerns; publication is at the discretion of the Chief Coroner's office, so an absent response does not mean nobody replied.
From the Lord Prior of Brampton
Ps Parliamentary Under Secretary of State for NHS Productivity (Lords)
Department
of Health
Richmond House
79 Whitehall
Ms Angela Hodes, Assistant Coroner London
Westminster Coroner’s Court SW1A 2NS
65 Horseferry Road P|
London
~ . SWIP 2ED
\ 04 MAY 2016
{ M, [ote
Thank you for your letter of 17 March 2016, following the inquest
into the death of Jacqueline Scott. I was sorry to hear of her death
and wish to extend my condolences to her family.
You raise four main concerns in this case regarding:
e the BIPAP Trilogy 202 machine and it’s visual display
symbols and alert alarms;
e staff training on use of the BIPAP machine;
e the provision of electrical power to the ward; and
e notification to and response from the hospital’s estates
management concerning a failed emergency call bell
Many of these issues need to be addressed by Phillips Healthcare and
the NHS Foundation Trust. However, I do acknowledge your
concerns about the design of the BiPAP machine and the safe
provision of power supply which I will address.
My officials contacted colleagues at NHS England who have advised
that the design of some non-invasive ventilation (NIV) systems and
other critical devices could be improved by having safety features
which warn staff of delivery problems, such as disconnection and
power failure. NHS England has a close working relationship with
the Medicines and Healthcare Regulatory Agency (MHRA) and is
able to share such concerns with them. Their collective power to
improve the design of medical devices is often limited however, as
manufacturers are not required to make changes to a product if it
meets the relevant regulatory requirements (the Trilogy 202
ventilator is CE marked to show compliance with the Medical
Devices Directive). In order to impose manufacturers to make such
improvements would require a change to the UK and EU
requirements.
NHS England has been able to raise awareness of unintentional
interruption of this type of therapy however, and on 13 February
2015, it issued a Patient Safety Alert concerning the risk of severe
harm and death from unintentional interruption of non-invasive
ventilation. This publication was a joint effort with the Medicines
and Healthcare Regulatory Agency (MHRA) and served as a
reminder of the correct procedures and the importance of familiarity
with the ventilators and disposables used.
A copy of the alert can be found at:
https://www.england.nhs.uk/2015/02/psa-niv/psa-niv/
MHRA was made aware of the events outlined in your letter in April
2015 and submitted a report to the coroner which confirmed the
manufacturer had examined the device and found no faults.
The Phillips model of ventilator has features to alert users to the
status of the power supply. There is an LED that lights up on the unit
when on mains power, and icons on the display to show whether the
device is currently drawing from mains or battery power. There are
also multiple visual and audible alarms to alert users to the depletion
of battery power. On inspection by the manufacturer, the alarms
were seen and heard to function correctly. The device log showed
that on the day of the event, the ventilator had alarmed as expected to
alert users to the depleting battery power.
MHRA has conducted a search of its adverse incident database
which has not revealed any similar reported incidents (involving the
ventilator being used on battery power until it fully depleted) for this
model of ventilator. Philips has reported that it is aware of one
similar event reported to them in 2011, which occurred in the USA.
This involved a device which “was being used on DC power and an
AC source was unavailable when the batteries depleted”, but MHRA
is not aware of the detailed circumstances.
MHRA has confirmed that this model of ventilator was first placed
on the UK market in July 2010. At the time of its report to the
coroner, the manufacturer had advised that a total of 501 Trilogy 202
ventilators have been sold in the UK, with a further 1,582 sold in the
ae
Department
of Health
rest of Europe and 5,799 in the rest of the world. Given the number
of ventilators of this model that have been sold, the available
evidence indicates that battery power depletion with no corrective
action taken by the user is not currently a widespread problem with
this model of ventilator.
I am aware that Phillips Electronics has already provided you with a
detailed response to your Regulation 28 letter and has concluded that
no changes to the design of the Device, including the alarm system,
visual and audible alerts, and the instructions for use are required to
prevent future deaths. However, MHRA is planning to ask Philips
Electronics to consider the visibility of alarms and battery icons in
terms of usability in the next design review as part of their ongoing
post market surveillance.
With regard to issues around the provision of power to the ward, you
quote from our Department’s Hospital Technical Memoranda (HTM)
06-01and consider that there is a conflict of advice in Part A between
clauses 4.22 and 6.62.
HTM 06-01 Electrical services supply and distribution - Part A:
design considerations, provides best practice guidance and should be
read in conjunction with the Institution of Engineering and
Technology (IET) Wiring Regulations, British Standard 7671. The
HTM sets out the application of BS7671 in the specific, unique
context of healthcare.
Moreover, the text at paragraph 4.22 relating to Category 4 —
Patients in special medical locations should be read in a holistic
manner in conjunction with the guidance provided within other
sections of the HTM and BS7671. If this task is undertaken, there is
no conflict of advice between 4.22 and 6.62 of the HTM.
By way of illustration, the HTM contains the following guidance
which it is imperative to follow:
2.6 It is recommended that designers and stakeholders review
Chapter 4 as well as Chapter 6 for all projects.
4.17 While it is not intended to be absolute, this section should be
sufficient to prompt the necessary discussion at all stages of the
design process. The categories given are intended to demonstrate a
range of patient risk from an electrical fault or loss of electrical
supply
4.18 Consideration of the categories in Figure 6 should establish a
minimum acceptable risk option at the point of treatment or care.
For the purpose of this guidance, the patient levels described are not
intended to be exhaustive, but rather an aid to consider the issues.
The definition relating to Medical Location (Chapter 1) (location
intended for the purpose of diagnostic treatment (including cosmetic)
or monitoring a patient under medical supervision), identifies those
locations where discontinuity of the electrical supply can cause
danger to life.
* Group 0 Medical locations where no applied parts are
intended to be used.
* Group 1 Medical locations where discontinuity of the
electrical supply is not a risk to human life (unless the location
is part of a Group 2 location).
* Group 2 Medical locations where discontinuity of the
electrical supply can cause danger to life.
Should a patient’s life be endangered by a discontinuity of supply,
then a Group 2 definition may be required to be applied to the
Category 4 patient clinical risk.
The HTM contains the following guidance with respect to Group 2
areas:
16.37 IEC 60364-7-710 and BS 7671 require Group 2 areas to have
at least two separate socket-outlet sub circuits at each patient
treatment location (for example bedhead or theatre pendant). This
applies to Group I areas also. This can be achieved from a single
IPS unit with an integral single-phase distribution board. The
resilience would be further enhanced if the IPS had dual 100%-rated
isolation transformers serving different integral distribution boards.
Such arrangements would provide an N+I resilient IPS isolation
transformer as defined in paragraphs 6.8-6. 14.
The HTM also provides guidance on the provision of audible and
visual alarms in relation to interruptions to power supply failures and
the need to provide indication at the nurse’s station for the relevant
medical area.
ae
Department
of Health
For the reasons set out above, the Department does not consider that
there is a conflict of advice in the HTM. However, the concerns you
raise are noted and the relevant sections of the HTM will be
considered and reviewed, as part of the wider technical guidance
programme, to determine if there is a need for greater clarity.
Similarly, the issue of alerts will be considered, although the HTM is
considered to provide adequate guidance on this issue.
I hope that this reply is helpful and I am grateful to you for bringing
the circumstances of Mrs Scott’s death to my attention.
ides be
DAVID PRIOR =
St George's University Hospitals Nh NHS Foundation Trust St George’s University Hospitals NHS Foundation Trust 11th May 2016 Blackshaw Road Ms Angela Hodes London SW17 0QT Tel: 020 8725 4555 Assistant Coroner for Inner West London Westminster Coroners Court 65 Horseferry Road London SW1P 2ED Dear Ms Hodes | am writing in response to the Regulation 28 report that you issued to St. George’s University Hospitals NHS Foundation Trust on 17/03/2016 regarding the death of Mrs. Jacqueline Scott. In relation to the matters of concern you raised the following points: 1) The Philips Respironics Trilogy 202 (BIPAP) 2) 3) i) ii) The BIPAP Trilogy 202 machine had a subtle visual display symbol which denoted when the machine was running on battery power. That symbol is not visible if there are many alert alarms as the alarm messages fill up the screen as they come in pushing the earlier alerts (low battery) off the screen. The BIPAP Trilogy 202 Machine has the same alarm sound for battery depletion as for circuit disconnect (where for example the face mask slipped) which was the more usual and expected reason for an alarm and these two factors separately and together did not have any feature of urgent warning to alert staff to battery depletion. Staff Training ii) Staff who were experienced and trained on the BIPAP machine did not appear to be trained to be alert to the situation or to the significance of a battery symbol showing on the machine when the machine was plugged in to the mains or to any particular alarm which denoted battery depletion rather than mask slippage. The provision of power to the ward iv) v) Richmond ward ADU beds was designated as a category 4 area which in this case meant there was no isolated power supply (IPS) provided to the ward not withstanding life-saving equipment was routinely used. Hospital Technical Memoranda (HTM) 06-01 Part A provides advice and guidance and benchmarking standards for electrical installation, maintenance and safety etc in healthcare premises. It is a matter of concern that there is a Chairman: Sir David Henshaw Chief Executive: Prof Simon MacKenzie conflict of advice between clause 4.22 and Clause 6.62. Clause 4.22 states: “ clinical treatment and patient safety may be compromised (but not endangered) by any interruption of electrical supply” whereas clause 6.62 states: “In clinical risk Category 4 and 5 areas the patient environment should have at least two IPS circuits at the bedhead” vi) There was no system or check that would alert ward staff to the failure of mains power in any particular area. 4) Notification to estates management vii) The crash bell for bed 5 did not work when the emergency arose. However estates management had been notified some days earlier of the broken patient call in the same bay. This was of concern as both emergency bells were on the same circuit and not fixed until 2 April 2014 when by chance the failure of electricity was identified. In response to these concerns for ease | will address each of these points individually. Point 1 - The Philips Respironics Trilogy 202 (BIPAP) Ms Louise Best has provided a written response on behalf of Phillips on the 12" April 2016. Point 2 — Staff Training During the Serious Incident investigation, the panel reviewed the training for nursing staff provided for non-invasive ventilation (NIV) and the particular medical device (Trilogy ventilator). They found that training was overseen by a nurse consultant and a clinical nurse specialist in respiratory medicine. There were appropriate clinical competencies developed as part of the Trust’s NIV policy. The nurses had undergone either formal (e.g. classroom based) training or informal on the job training by the specialist nurses, practice educators and senior staff on the respective ward and in ED. The panel was satisfied during the investigation that all nursing staff involved in the care of Mrs Scott were appropriately trained on the particular machine and NIV therapy. There was also access to senior help and a practice educator was available on the day when the incident happened. The nurse directly involved in the incident was a trained ITU nurse of six years, who had additional training on the NIV and the Trilogy machine. The panel found, however, that there was no consistent documentary evidence of the training that staff had received. The panel therefore concluded that staff training was appropriate in this case, but recommended an action in the SI report to complete a training needs analysis to establish the exact number of other staff that had been trained and to improve documentation of this training. Following this incident staff working in this area have undergone a period of retraining by Phillips which includes the points raised in point 1. These key aspects of training have further been incorporated into the training delivered by the practice educator and specialist nurses. The unit has further employed a dedicated practice educator to work with staff in ADU. All staff other than two new starts have now received training on NIV and this device and competencies are held within the department. Point 3 - The provision of power to the ward The Trust engaged an external expert to advise the organisation on the areas where the electrical infrastructure requires upgrading to comply with HTM 06-01. This advice has Chairman: Sir David Henshaw Chief Executive: Prof Simon MacKenzie formed the basis of the design of a UPS/ IPS back up system in the Richmond ADU area. The work has now been put to a tender process with the summer 2016 set as the date for completion of this work due to the co-ordination that is required with clinical areas for this work to be completed. The Trust has set aside sufficient funds for this work to be completed and once installed the designed UPS/IPS back up system will alert staff of any failure in the electricity supply to the mains sockets. | would be happy to update you on the progress of this work following the tendering process and on completion of the installation. Point 4 - Notification to estates management Call bell repairs are part of the top priority (Priority 1) schedule. Calls rated at Top priority are attended within 3 hours of the call being logged. Timely Response to Priority1 calls forms part of the Key Performance Indicators for the Estates department. The electrical circuits in the Richmond ADU area are to be upgraded in line with the response under section 5 part 3 of the section 28 notice. Therefore the installation of the UPS/IPS system following the tendering process will ensure that this requirement is met. Additionally following this incident the nursing staff within this area undertake checks of emergency call bell systems twice weekly to ensure they are operational. This is recorded on department checking sheets which are held locally and any faults reported to estates. | hope that the above actions address your concerns and provide reassurance that the Trust is fully committed to providing our patients with the highest level of care and to this end we are continually seeking to improve patient safety. If you require clarification on any of the above actions, | would be very happy to respond further. Jennie'Hall Chief Nurse and Director of Infection Prevention and Control Chairman: Sir David Henshaw Chief Executive: Prof Simon MacKenzie
See every Prevention of Future Deaths report matching Hospital Death (Clinical Procedures and medical management) related deaths, and how often a new one appears.
What would an alert for this have sent me? Search the full text
Free to try — the preview shows the real matches and how many arrived in the last 12 months. Your first email alert is free.
These reports are published by the Chief Coroner's office at judiciary.uk and are © Crown copyright. The text here is reproduced from the published PDF so it can be searched. If something on this page is wrong, or you are a person named in it and want it reviewed, email drcjar@gmail.com and we will act promptly.