Regulation 28: REPORT TO PREVENT FUTURE DEATHS 

REGULATION 28 REPORT TO PREVENT DEATHS 

THIS REPORT IS BEING SENT TO:  
1  MEDTRONIC  
2  ROYAL BERKSHIRE NHS FOUNDATION TRUST  
3  BERKSHIRE INTEGRATED CARE BOARD  
4  MEDICATION AND HEALTHCARE PRODUCTS REGULATORY AGENCY  
5  NHS ENGLAND   

1  CORONER 

I am Katy Thorne KC, Assistant Coroner for the coroner area of Berkshire. 

2  CORONER’S LEGAL POWERS 

I make this report under paragraph 7, Schedule 5, of the Coroners and Justice Act 2009 and 
Regulations 28 and 29 of the Coroners (Investigations) Regulations 2013. As set out in the 
case of R (
issuing of a Regulation 28 report entails no more than the coroner bringing some information 
regarding a public safety concern to the attention of the recipient. The report is not punitive 
in nature. 
INVESTIGATION  

) v Assistant Coroner for East London, the 

3 

On 23 November 2020, I commenced an investigation into the death of Devon Drew Turner 
aged 16 weeks. The investigation concluded at the end of the inquest on 16 August 2023. 
The conclusion of the inquest was natural causes. The medical cause of death was Sudden 
Unexpected Death in an Infant with Trisomy 9 Mosaicism (SUDI with Mosaic Trisomy 9) 

4  CIRCUMSTANCES OF THE DEATH 

1. Devon Drew Turner was born on 17 January 2022 at Basingstoke and North Hampshire 
Hospital. He died on 10 May 2022 at Royal Berkshire Hospital of Sudden Unexpected Death in 
an Infant with Trisomy 9 Mosaicism. (SUDI with Mosiac Trisomy 9)  
2. Devon was born with a number of abnormalities due to a rare chromosomal disorder, 
Mosaic Trisomy 9 which had caused him serious health vulnerability, and particularly with his 
respiratory system, throughout his short life. He had been admitted to the Paediatric 
Intensive Care Unit twice.  
3. On 21 April 2022 he was discharged home to live with his parents and needed respiratory 
support on a CPAP machine throughout the night and for long periods during the day and 
monitoring on a SATS machine. Throughout Devon’s time at home he was visited by the 
community nurses every 2-3 days and the family were supported by the specialist nurses at 
UHS. 
4. On 30 April 2022 he was admitted to Basingstoke and North Hampshire Hospital with an 
increasing need for CPAP but was discharged the following day at 12.50pm as he remained 
stable.  
5. On 10 May 2022 Devon was given his 1am feed and the family settled down to sleep. At 
4.30am Devon’s mother woke to find him unresponsive. An ambulance was called and he was 
taken to the Royal Berkshire Hospital but despite the best efforts of the medical teams he 
could not be revived and he died the same morning. Despite extensive post-mortem 
examination by a number of specialists it was not possible to come to any firm conclusion 
about what specifically caused his death, and the cause of death was given of SUDI with 
Trisomy 9 such that it was possible to conclude that it was a natural death and was linked 
directly to his diagnosis of trisomy 9.  

Regulation 28 – Before Inquest 
Document Template Updated 30/07/2020 

 
 
 
 
 
 
 
 
 
 
 
 
 
 6. On examination of the CPAP machine that was maintaining his breathing, no faults were 
identified and it was confirmed that the only interventions by human hand were consistent 
with the parents’ account, namely that he was given his feed at 1pm, the parents had gone 
to sleep and had woken again at 4.30am to find Devon unresponsive, the CPAP machine 
continuing to blow. The CPAP machine recorded at 4.30am when they disconnected him from 
it and turned it off as they called 999.  
7. By contrast the SATS machine, which was also examined and confirmed to be functioning, 
was set to alarm when Devon’s oxygen levels fell below 90%, but despite his oxygen being 
recorded as falling below this level, an alarm was not heard by the parents. There are some 
anomalies in the evidence relating to the SATs machine.  
8. Firstly the SATS machine technician provided evidence that the data revealed that the 
machine was not used until 3 May 2022. The evidence suggested that it was the same 
machine that had been used by the family consistently since 25 April 2022.  
9. Secondly the SATS machine technician provided evidence that the data showed that from 
1.44am to 2.00am the SATS machine would have been sounding an alarm that could be 
silenced temporarily by pressing a button but that the alarm would restart after 60 seconds, 
sounding again. The alarm sound on this machine was set on maximum. A second alarm 
would have been sounding at 2.00am due to a loss of pulse. 

5  CORONER’S CONCERNS 

During the course of the investigation my inquiries revealed matters giving rise to concern. In 
my opinion there is a risk that future deaths could occur unless action is taken.  In the 
circumstances it is my statutory duty to report to you. 

The MATTERS OF CONCERN are as follows. –  

According to the evidence heard at the inquest:  
1. Devon was discharged home on 21 April 2022 with a SATS machine from University 
Hospital Southampton (UHS). However, as the family lived in the Royal Berkshire NHS 
Foundation Trust area, and they used a different brand of SATS machine to that issued from 
UHS, this first machine was changed to a different second SATS machine the following day on 
22 April 2022.  
2. The family had been trained to use the UHS issued SATS machine.  

Matter of Concern 1:  
3. On 25 April 2022 this second SATS machine, stopped working. It had some sort of error 
message on the screen. The parents contacted the community nurse team and Devon’s 
mother had to attend the hospital that night to pick up a new SATS machine.  
4. Therefore the first concern is regarding the reliability of such SATS machines sent home 
with vulnerable patients.  
5. If a safety plan is put in place which includes the use of a SATS machine to monitor the 
wellbeing of the patient, and that machine is less than at least 99% reliable, or has software 
issues, or is unsuitable for home use, or is complicated to use, this may result in a false 
sense of security from the family who will rely on the machine to alert them if the oxygen 
levels drop below a certain figure.  
6. Therefore there is a risk that future deaths will occur if hospital trusts do not ensure that 
the SATS machines are reliable and easy to use and that parents are trained to use the 
particular brand they are issued with.  
7. Action should be taken by Royal Berkshire NHS Foundation Trust to identify what 
happened in this instance and to ensure that such events are avoided. This may involve the 
local Integrated Care Board and/or MHRA.  

Matter of Concern 2:  
8. The third machine was issued to the family on 25 April 2022, which remained with them 
until Devon’s death. It was manufactured by Medtronic (Model number MBH1920704). It is 
not known if this was the same brand as the second SATS machine mentioned above.  
9. The concern is that this Medtronic machine was also either not functioning reliably or was 
not suitable for the home environment.  

Regulation 28 – Before Inquest 
Document Template Updated 30/07/2020 

 
 
 
 
 
 
 
 10. On 10 May 2022 Devon was given his 1am feed and the family settled down to sleep. At 
4.30am Devon’s mother woke to find him unresponsive. The CPAP machine log confirms 
those two times.  
11. The SATS machine was seized by the police on 10 May 2022.  
12. Medtronic technician, 
append to this Report, in which he states that: 

 made a statement dated 6/7/2022 which I 

• The testing of the SATS machine revealed that it was functioning properly 
• The data from the machine shows that from approximately 1.44am Devon’s oxygen 
saturations started to drop and continued to fall until approximately 2.00am when 
there was no pulse.  
• The alarm would have been sounding throughout that time and that at the loss of 
pulse, at 2.00am a further alarm would have sounded.  
• The machine alarm was at maximum volume  
• There is a silence button on the machine but that it only silences the alarm for 60 
seconds, after which interval the alarm would sound again throughout the whole 
period that the saturation levels were low.  

13. Given the crucial role a SATS machine has in monitoring a vulnerable baby at home, with 
non-medically trained carers who need to sleep, there should be no doubt but that the alarm 
sound will sound reliably, that its volume will wake sleeping exhausted parents and should 
provide an accurate log of events.  
14. The conclusions that the data from the SATS machine appear to provide do not accord 
with the other evidence provided to the inquest about the events in question. That gives 
cause for concern that either the analysis of the SATS machine has not been accurate or the 
SATS machine has not correctly recorded the data or that this SATS machine is not a suitable 
device for use at home.  
15. For example, the community nurses would have noticed if the SATS machine was not 
working in the first weeks of Devon being at home and yet it has recorded that it was not 
used at all until 3 May 2022.  
16. Secondly in order for the SATS machine data to have been correct about events of 10 
May 2022, either both the parents would have had to sleep through an alarm sounding for at 
least 15 minutes at full volume as well as a second alarm after 15 minutes at the loss of 
pulse, or the parents would have had to have used the silence button on the alarm system 
every 60 seconds for that 15 minute period.  
17. The SATS machine silence button was not within reach of Devon’s mother unless she sat 
up and reached up and over baby Devon in his cot. Devon’s father would have had to get out 
of bed altogether. A copy of the police photographs of the bedroom on 10 May 2022 is 
attached to this report. It is inconceivable that these concerned and careful parents, who had 
taken such an active role in Devon’s care, would have turned off this alarm in this way 
several times and failed to notice his respiratory distress, and when he was found 
unresponsive, would have forgotten that they had silenced the alarm in this way.  Therefore 
either the SATS machine did not function as it should or was not sufficiently loud to wake 
either of these careful parents.  

18. The potential for future deaths is that future parents will also rely on the SATS 
machine to alert them to a vulnerable child stopping breathing and taking action will not be 
alerted because either the alarm is too quiet, or it cuts out automatically before waking the 
parents, or it simply does not sound at all.  
19. Action needs to be taken by the Trust, Integrated Care Board, MHRA, NHS England and 
Medtronic to investigate these events and establish whether the Medtronic SATS alarm may 
be insufficiently loud or may be unreliable. 

6  ACTION SHOULD BE TAKEN 

In my opinion action should be taken to prevent future deaths and I believe you (and/or your 
organisation) have the power to take such action. 

7  YOUR RESPONSE 

You are under a duty to respond to this report within 56 days of the date of this report,  
namely by 13/10/2023.   

Regulation 28 – Before Inquest 
Document Template Updated 30/07/2020 

 
 
 
 
 
 
 Your response must contain details of action taken or proposed to be taken, setting out the 
timetable for action.  Otherwise you must explain why no action is proposed. 

8  COPIES and PUBLICATION 

I have sent a copy of my report to the Chief Coroner and to the following Interested Persons      

• The family of Devon Turner 
• UHS NHS Trust 

I have also sent it to         

• Medtronic  
• Royal Berkshire NHS Foundation Trust 
• Berkshire Integrated Care Board  
• Medication And Healthcare Products Regulatory Agency  
• NHS England  

who may find it useful or of interest. 

I am also under a duty to send the Chief Coroner a copy of your response.  

The Chief Coroner may publish either or both in a complete or redacted or summary form. He 
may send a copy of this report to any person who he believes may find it useful or of 
interest. You may make representations to me, the coroner, at the time of your response, 
about the release or the publication of your response by the Chief Coroner.  
 Dated: 18/08/2023 

9 

Katy Thorne KC 
Assistant Coroner for  
Berkshire 

Regulation 28 – Before Inquest 
Document Template Updated 30/07/2020

Responses

3 responses published against this report on judiciary.uk. A response is a body's written reply to the coroner's concerns; publication is at the discretion of the Chief Coroner's office, so an absent response does not mean nobody replied.

Response from Buckinghamshire Oxfordshire and Berkshire West Integrated Care Board (PDF)
Sandford Gate, Second Floor  
East Point Business Park  
Oxford  
OX4 6LB 

Ms K Thorne 
HM Coroners court  
Reading Town Hall 
Blagrave Street 
Reading  
RG1 1QH 

12th October 2023 

Dear Ms Thorne 

Re: Regulation 28 Report to Prevent Future Deaths – Devon Drew Turner who died on 
10 May 2022 

Thank  you  for  your  Report  to  Prevent  Future  Deaths  (hereafter  “Report”)  dated  18 August 
2023 concerning the death of Devon Drew Turner who died on 10 May 2022. In advance of 
responding to the specific concerns raised in your Report, I would like to express my deep 
condolences to Devon’s family and loved ones. Buckinghamshire Oxfordshire and Berkshire 
West ICB (BOB ICB) are keen to assure the family and the coroner that the concerns raised 
about Devon’s care have been listened to and reflected upon. 

I have liaised with our colleagues who represent system partners. On 13th May a Joint Agency 
Response meeting held with all stakeholders, I have assurance that actions were identified 
and completed with a copy of the minutes being sent to the coroner’s office for consideration. 
Subsequently,  on  8th  June  a  Child  Death  Review  meeting  was  convened  with  all  partner 
organisations involved in Devon’s care, and at the time of his death.  

Although Berkshire Healthcare NHS Foundation Trust were not invited to participate in the 
inquest, nor cited in the Regulation 28 report, I have sought clarification from them as part of 
this reply, as they sit within the Buckinghamshire, Oxfordshire and Berkshire Integrated Care 
System and they supplied the equipment cited in the report to the parents of the deceased.  

I have raised questions with Berkshire Healthcare NHS Foundation Trust directly regarding 
the provision of equipment, training, service, and maintenance process and I have enclosed 
their response at Appendix 1 for completeness.  

The  engineer  from the manufacture  of the  SATS  machine  detailed their  findings  during the 
inquest  of  which  is  included  within  the  Report.  I  have  contacted  the  Medicines  Healthcare 
Regulatory Authority (MHRA) by both telephone and follow-up email, I am cognisant that there 
are defined processes for direct escalations of concerns regarding pharmaceutical equipment. 
The organisation confirmed that they are aware of the incident from the Regulation 28 Report 
and  are  in  receipt  of  the  manufacturer’s  investigation  into  the  Medtronic  equipment  (serial 
number MBH1920704).  Which was examined by the manufacturer with the police present, 
and it was confirmed that no fault was found with the device. The alarm volume was set at its 

www.bucksoxonberksw.icb.nhs.uk 

 
 
 
 
 
 
 
 
 
 
 
 
 maximum. The MHRA is continuing to work with the manufacturer to ensure this matter is fully 
investigated and action taken as needed. 

Furthermore, the MHRA detail that the operator’s manual for the device indicates it is intended 
for home use when used as an adjunct in patient assessment and is to be used in conjunction 
with monitoring for clinical signs and symptoms. There are several warnings in the operating 
manual  regarding  the  alarms  including  for  example  ensuring  the  speaker  is  clear  of  any 
obstruction to prevent an inaudible alarm tone. Also, to keep patients under close surveillance 
when monitoring as it is possible, although unlikely, that radiated electromagnetic signals from 
sources external to the patient and the monitoring system can cause inaccurate measurement 
readings and not to rely entirely on the monitoring system's readings for patient assessment. 
The team confirmed that they are also in contact with the coroner’s office to determine what 
response  is  required  from  the  MHRA  in  relation  to  their  findings  and  recommendations  for 
actions. The MHRA  keeps the safe  and  effective use  of  all medicines  and medical  devices 
under continuous review.  

To conclude, I am satisfied that all organisations detailed within the Report, have responded 
proportionately. Additionally, that system partners have collectively examined the timeline of 
events with those directly involved supporting the family to care for Devon.  

I would also like to provide further assurances taking place within BOB ICB around the Reports 
to  Prevent  Future  Deaths.  All  reports  received  are  summarised  within  the  Learning  from 
Deaths  system  Network  Group.  This  ensures  that  key  learnings  and  insights  around 
preventable deaths are shared widely across our Health and Social Care organisations and 
helps us pay close attention to any emerging trends or themes that may require further review 
and action. 

Your Regulation 28 Report asked the ICB to take action to identify what happened in relation 
to the reliability and ease of use of the SATS machines being used. The ICB and the MHRA 
have been involved in the oversight of the review of the machine and, following a review of 
the evidence presented at the inquest, the ICB feel that there is no further action needed in 
order to Prevent Future Deaths.  

You also ask that the trusts, ICB, MHRA, NHSE and Medtronic investigate the events leading 
up to the death in relation to the loudness and reliability of the alarm on the Medtronic SATS 
machine. This has been undertaken as described above and the ICB, on behalf of all partners, 
feel that there is no further action to be taken in order to Prevent Future Deaths. 

Thank you for bringing these important patient safety issues to my attention and please do not 
hesitate to contact me should you need any further information. 

Your sincerely 

Chief Nursing Officer 

www.bucksoxonberksw.icb.nhs.uk 

 
 
 
 
 
 
 
 Appendix-Berkshire Healthcare NHS Foundation Trust response to direct questions: 

In relation to the bedside saturation monitor given to the family on 22 April:  

1)  Did the family receive training, in line with recognised Trust processes, to use the 

saturation monitor that was given to them on 22 April?  
The Community Children’s Nurse [CCN] recorded in Devon’s clinical record that she had 
explained to the parent how to use the handheld and bedside oxygen saturation monitors 
on the day she gave them the equipment. This was in line with Trust processes. At the 
first home visit by the CCN all medical equipment in use was reviewed, checked, and 
documented in Devon’s clinical record. Berkshire Healthcare's ‘Acknowledgement and 
terms of Equipment loan’ document was completed and signed by the parent and the 
CCN at the next home visit on 27 April 2022.  

2)  Did the family have a designated contact and/or contact number to use if they 

identified any issues with the saturation monitor?  
During the first home visit on 22 April 2022 the CCN explained the CCN service. She also 
discussed her role as named CCN in supporting Devon’s care in the home. Parents are 
encouraged and advised to contact the CCN team with any queries or concerns they 
have about their child or their child’s care and are given contact numbers for the team. 
There is evidence in the clinical record to show that the family contacted the CCN team 
when they needed advice around Devon’s care, and also when they had concerns about 
the monitor being faulty on 25 April 2022. 

3)  Please confirm the type and nature of the saturation monitor provided to the family 

on 22 April.  
The family were supplied with a tabletop (bedside) Nellcor PM 100N and a handheld 
Nellcor PM10N  

4)  Please confirm whether the saturation monitor supplied to the family on 22 April 
was of a different type to that which had previously been used by the family.  
The saturation monitor was different to the monitor that had been used at University 
Hospitals Southampton [UHS].  

At the discharge planning meeting on 19 April 2022 Devon’s Consultant/ Clinical Nurse 
Specialist team at UHS advised that Devon would require overnight and spot-check 
saturations as per his respiratory care plan. It was agreed that Devon would be 
discharged home with an oxygen saturation monitor from the ward at UHS and that this 
would be replaced with a Nellcor oxygen saturation monitor loaned to him by the CCN 
team whilst specific [Nellcor] equipment was being ordered for him. The agreement was 
that the parent would return the UHS equipment to the ward the following week when he 
returned for a scan. 

5)  Was the saturation monitor supplied to the family by Berkshire Healthcare on 22 

April checked and maintained in line with the Trust's maintenance specifications? 
All equipment supplied to Devon had been checked by the CCN before allocation, was 
within its service dates and had been serviced annually as per manufacturers guidelines.  

6)  Have there been any other reports from families/users or staff of faults with the 
type of saturation monitor supplied to the family on 22 April? If so, what actions 
have been taken to investigate/rectify these?  
Nellcor is the only make of oxygen saturation monitor that the CCN service provide to 
families. The parents’ report of a fault on 25 April regarding the machine supplied to them 
on 22 April is the only instance recorded of a fault of this kind with this type of equipment. 
All reported faults are listened to and acted upon by the team and equipment is sent for 
review and repair, and replacement equipment is issued. The team hold a pool of 
equipment to replace any items that need to be sent for repair or for annual maintenance. 
A clear record is kept of when oxygen saturation monitors have been brought in for 

www.bucksoxonberksw.icb.nhs.uk 

   
 
 
 
 
 
 
 review and repair and also for annual planned preventative maintenance. The team 
always provide families with additional equipment in the home in case of equipment 
developing faults. Devon’s family were provided with a bedside oxygen saturation monitor 
and a handheld portable saturation monitor. 

In relation to the bedside saturation monitor given to the family on 25 April which remained in 
place on 10 May (i.e., at time of death):  

1)  Did the family receive training, in line with recognised Trust processes, to use 

the saturation monitor provided to them on 25 April?  
The Community Children’s Nurse [CCN] recorded in Devon’s clinical record on 25 April 
2022 that the monitor had been replaced and that she had explained to the parent how 
to use the monitor. Berkshire Healthcare’s ‘Acknowledgement and terms of equipment 
loan’ document was completed with details of equipment supplied and was signed by 
Devon’s named CCN and the parent and uploaded to Devon’s clinical record on 27 
April 2022. This document includes an acknowledgement that the equipment is in good 
working order and that the parent understands instructions for use. It also asks users to 
immediately contact a member of Trust staff if the equipment is broken or damaged. 

2)  Did the family have a designated contact and/or contact number to use if they 

identified any issues with the saturation monitor?  
During the first home visit on 22 April 2022 Devon’s named CCN explained the CCN 
service. She also discussed her role as named CCN in supporting Devon’s care in the 
home. Parents are encouraged and advised to contact their CCN or the CCN team with 
any queries or concerns they have about their child or their child’s care and are given 
contact numbers for the team. There is evidence in the clinical record to show that the 
family contacted the named CCN and other members of the CCN team when they 
needed advice around Devon’s care, and also when they had concerns about the initial 
monitor being faulty on 25 April 2022 and the monitor was promptly replaced that day. 

3)  Please confirm the type and nature of the saturation monitor provided to the 

family on 25 April.  
The family were supplied with a tabletop (bedside) Nellcor PM 100N.  

4)  Please confirm whether the saturation monitor supplied to the family on 25 April 
was of a different type to that which had previously been used by the family.  
The saturation monitor supplied on 25 April 2022 was the same make and model as the 
monitor supplied on 22 April 2022.  

5)  Was the saturation monitor supplied to the family by Berkshire Healthcare on 25 

April checked and maintained in line with the Trust's maintenance 
specifications?  
All Oxygen saturation monitors held by the Community Children’s Nursing team have 
annual planned preventive maintenance in line with Berkshire Healthcare’s HS020 
Maintenance and repair of medical devices policy and procedure. The team keep an 
inventory of all medical devices and the dates when equipment needs to be recalled for 
scheduled maintenance. All equipment has a visual check before it is allocated, and it 
is checked to see if it is working correctly before it is set up in the patient’s home or 
given to the patient. 

All equipment supplied to Devon had been checked by the CCN before allocation, all 
were within their service dates and had been serviced annually as per manufacturers 
guidelines.  

6)  Have there been any other reports from families/users or staff of faults with the 
type of saturation monitor supplied to the family on 25 April? If so, what actions 
have been taken to investigate/rectify these?  

www.bucksoxonberksw.icb.nhs.uk 

 
   
 
 
 
 
 
 
 All reported faults in oxygen saturation monitors are acted upon by the team and 
equipment is sent for review and repair, and replacement equipment is issued. The 
team hold a pool of equipment to replace any items that need to be sent for repair or for 
annual maintenance. A clear record is kept of when oxygen saturation monitors have 
been brought in for review and repair and also for annual planned preventative 
maintenance. The team always provide families with additional equipment in the home 
in case of equipment developing faults. Devon’s family were provided with a bedside 
oxygen saturation monitor and a handheld portable saturation monitor. 

www.bucksoxonberksw.icb.nhs.uk
Response from Medtronic (PDF)
Medtronic Limited 
Building 9 Croxley Park 
Hatters Lane 
Watford 
Herts 
WD18 8WW 
United Kingdom 

www.medtronic.com 

Inquest touching the death of Devon Drew Turner 

Medtronic Response to Regulation 28: Report to Prevent Future Deaths dated 18 August 2023 

We  refer  to  the  Inquest  into  the  tragic  death  of  Drew Turner,  Deceased  (“the  Deceased”)  on  16  August  2022, 
subsequent  Regulation 28: Report to Prevent  Future Deaths dated 18  August  2023  and  HM Assistant Coroner’s 
letter dated 9 October 2023 enclosing further documents. We set out below Medtronic’s response. 

Incident as reported to Medtronic Limited and Immediate Actions Taken 

Attendances at Reading University Hospital to download the device memory and perform functional testing of the 
PM100N Bedside Monitor and Nellcor Saturation probe. 

On 11 May 2022, the Adult Nursing Account Manager for Patient Monitoring Solutions at Medtronic Limited (“the 
Medtronic  Adult Nursing Account Manager”) reported internally that she had been informed by a  healthcare 
professional at Reading University Hospital (“the Healthcare Professional”) that a patient who had been monitored 
by a PM100N Bedside Monitor and Nellcor Saturation probe serial number MBH1920704 (“the PM100N device”) 
used in a home environment had died on 10 May 2022, and that the police were involved. 

By email dated 11 May 2022, the Medtronic Adult Nursing Account Manager asked the Healthcare Professional to 
isolate the PM100N device and leave it on charge, informing her that if an investigation was underway, the download 
of data by Medtronic should ideally take place in the presence of the police.  The Medtronic Adult Nursing Account 
Manager also requested that the PM100N device be returned to Medtronic for testing after the download had taken 
place. 

By  email  dated  14  May  2022  to  the  Medtronic  Adult  Nursing  Account  Manager,  the  Healthcare  Professional 
suggested that the police did not need to be present for the download and had asked several questions in respect of 
the PM100N device’s functionality. 

On  17  May  2022,  the  Medtronic  Adult  Nursing  Account  Manager  responded  by  email  to  these  questions  and 
requested a copy of the data downloaded by the hospital on the date of death. The data provided was a pdf version 
of the data rather than the more detailed “raw” excel data available from the PM100N device’s memory. Both the 
pdf and Excel file present the same information in a different format. 

Arrangements  were  made  on  19  May  2022  for  a  further  download  of  the  “raw”  data  from  the  PM100N device 
memory to take place in the presence of the police on 25 May 2022. 

A Business Development Manager at Medtronic Limited (“the Medtronic Business Development Manager”) and 
the Medtronic Adult Nursing Account Manager attended Reading University Hospital on 25 May 2022 to download 
the data from the PM100N device in use by the Deceased at the time of his death. The data obtained was emailed 

1 

 
 
 
 
 
 
 
 
  
 
  
  
  
 
  
 
 
 
 
 
 
  
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 by the Medtronic Business Development Manager to police officer DC 7712 later that day. DC 7712 subsequently 
requested a brief summary of the download from the Medtronic Business Development Manager by email dated 25 
May 2022, who then provided images of the device screen and an explanation of the PM100N device settings to DC 
7712 on 26 May 2022. 

On 7 June 2022, DC 7712 contacted the Medtronic Adult Nursing Account Manager to advise that the police were 
trying to identify a police specialist to carry out a further download of the data to assist “in relation to the alarms 
and if they did sound”, as it was her understanding that the PM100N device would need to be returned to Medtronic 
in the United States for this testing to be done. 

In the absence of the Medtronic Adult Nursing Account Manager, a Regional Sales Manager at Medtronic Limited 
(“the Medtronic Regional Sales Manager”) contacted DC 7712 on 7 June 2022 stating that any further download 
should be carried out by an engineer from Medtronic who would be able to extract the device’s memory data. By 
email dated 9 June 2022, DC 7712 confirmed that a further download was required and that a Medtronic engineer 
could undertake this in the presence of the police as the PM100N device could not be taken out of the UK. 

On 6 July 2022, the Medtronic Business Development Manager re-attended Reading University Hospital with two 
other  personnel  from  Medtronic,  a  service  technician  (“the  Medtronic  Service  Technician”)  and  a  Senior 
Technical  Service  Supervisor  UK  and  Ireland,  Service  and  Repair  (“the  Medtronic  Senior  Technical  Service 
Supervisor”). The Medtronic Service Technician is an Italian national based in the Netherlands.  He travelled to 
the UK on 6 July 2022 to carry out technical checks on the PM100N device and to perform the data download. 
These were done in the presence of DC 7712. 

During the physical examination and functional testing of the PM100N device and subsequent data download, the 
Medtronic Service Technician, whose first language is Italian, explained to DC 7712 to the best of his ability, the 
nature of the checks he was completing and the findings from the data download. He also emailed to DC 7712 a 
copy of the downloaded data from the device memory and an Excel spreadsheet of the same data that day. 

The software used for the download on 6 July 2022 has passed formal verification and validation processes to ensure 
the data downloaded was accurate. 

Findings from the visual examination and functional testing 

Visual examination on 6 July 2022 found no physical damage to the monitor, which was in good condition. The 
device was powered up and no error messages or failures were displayed on the monitor, and it was ready to use. 
All  alarms  were  functioning  according  to  the  specification  and  were  clearly  audible  and/or  visible.  From  the 
downloaded data reviewed, the Medtronic personnel also concluded that the PM100N device was functioning as 
expected at the time of the death of the Deceased. 

Subsequently, between 19 July 2022 and 14 September 2022, Medtronic staff responded via email to certain email 
questions received from DC 7712 about the device alarm settings. 

Information notified to the MHRA. 

Medtronic is dedicated to ensuring the safety, suitability and reliability of all its products and takes all concerns 
potentially associated with its products extremely seriously. 

This incident was initially reported to the Medicines and Healthcare products Regulatory Agency (“MHRA”) on 
20  May  2022  with  follow-up  on  20  August  2022.  Further  correspondence  in  response  to  MHRA  queries  was 
provided on 18 July 2022, 10 October 2023, 27 October 2023 and 17 November 2023. Based on the conclusions 
following testing of the PM100N device on 6 July 2022, shortly after the Deceased’s death, and the review of the 

2 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 downloaded data, it was concluded that no corrective action was required in respect of the PM100N device. 

Additional investigations carried out in order to respond to the Regulation 28 Report 

By letter dated 9 October 2023, HM Assistant Coroner granted Medtronic Limited “Proper Person” status, following 
which  selected  information  and  documentation  was  provided  to  the  company.  This  has  assisted  Medtronic  in 
preparing  this  response.  In  that  letter  HM  Assistant  Coroner  informed  Medtronic  that  she  found  “…that  the 
[PM100N device] did not contribute to the death” and that the Deceased died of natural causes. 

Since receipt of the documentation from HM Assistant Coroner, Medtronic has conducted further investigations into 
the machine’s functionality in order fully to address HM Assistant Coroner’s concerns expressed in the Regulation 
28: Report to Prevent Future Deaths dated 18 August 2023. 

Witness Statements obtained by the police from Medtronic personnel and used in the Inquest. 

We  understand  that  witness  statements  were  obtained  by  the  police  from  the  Medtronic  Business  Development 
Manager  and  from  the  Medtronic  Service  Technician.  It  appears,  although  Medtronic  has  not  had  access  to  a 
transcript of the inquest, nor all of the evidence relied upon, that these witness statements were subsequently relied 
upon by the Court for the purpose of assessing: (a) the functionality of the PM100N device; and (b) the significance 
of  the  data  that  had  been  logged  in  the  PM100N  device  memory,  as  downloaded  by  the  Medtronic  Service 
Technician. 

We note from the Regulation 28 Report that HM Assistant Coroner found anomalies in the evidence relating to the 
PM100N device, namely that “the data revealed that the machine was not used until 3 May 2022” whereas it had 
been in use by the family since 25 April 2022, and secondly that “from 1.44 to 2.00am the [PM100N device] would 
have been sounding an alarm that…was set on maximum” and that a “second alarm would have been sounding at 
2.00am due to a loss of pulse”. For the reasons set out below, the evidence relied on in reaching these conclusions 
was incomplete. 

The  witness  statements  in  question  were  obtained  from  a  business  development  manager  and  from  a  service 
technician  who  is  not  resident  in  the  UK.  While  these  personnel  were  appropriately  qualified  to  perform  the 
requested tasks of download and functional testing, they were not, nor did they purport to be, technical experts on 
the PM100N device nor able to give expert evidence on the significance of the logged data from the memory.  At 
the time the request was made to Medtronic to download the data from the PM100N device memory, and prior to 
the  attendance  of the  Medtronic  personnel  at  Reading  University  Hospital,  Medtronic  personnel  were  under  the 
impression that their role was solely for the purpose of providing access to the data itself.  Had the company been 
made aware that expert witness evidence on the device functionality and the downloaded data would be needed for 
the  purpose  of  expert  evidence  to  be  considered  at  the  Inquest,  the  company  would  have  arranged  for  an 
appropriately  qualified  and  experienced  technical  expert  on  the  product  to  have  provided  such  evidence.  Such 
expertise would have had to have been found within the Medtronic design team based in the USA. 

Neither of the witnesses had any prior experience of giving evidence in inquest proceedings. Neither of them were 
informed by the police about the potential relevance of the witness statements they provided after the data downloads 
nor about the Inquest process. In performing the downloads and testing, and in providing the witness statements, 
they  sought  to  assist  the  police  to  the  best  of  their  ability  without  a  full  appreciation  of  the  use  to  which  the 
information they were providing would be put. The process is summarised as follows. 

On 31 May 2022, DC 7712 contacted Medtronic Business Development Manager by email advising that she would 
draft a witness statement for him based on his email. A draft statement was emailed to him for approval later that 
day. The witness statement was approved by Medtronic Business Development Manager by email dated 7 June 2022 

3 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 and he was asked to sign, scan and return it to DC 7712 the same day.  DC 7712 chased for this on 14 June 2022. 
The MG11 form attached to her email was marked to indicate the statement was not to be used for “civil or other 
proceedings”.  DC7712  printed  the  statement  and  provided  it  to the Medtronic  Business  Development  Manager 
when he attended Reading Hospital for the further download on 6 July 2022. The Medtronic Business Development 
Manager  signed  the  statement  on  the  same  day.  The  data  download  obtained  by  the  Medtronic  Business 
Development Manager on 25 May 2022 was exhibited to this witness statement dated 6 July 2022. 

DC 7712 prepared a witness statement for the Medtronic Service Technician whilst he was testing the PM100N 
device at the Hospital on 6 July 2022. Shortly after the Medtronic Service Technician had completed the testing, 
and before he left the testing room, DC 7712 presented him with a statement that she had typed on her screen. This 
represented her notes of the oral comments the Medtronic Service Technician had made as he was downloading the 
data. DC 7712 asked the Medtronic Service Technician to read and sign the statement then and there, on-screen. He 
was given only limited opportunity to review the statement. He is not a native English speaker, nor a UK resident. 
Medtronic are informed that neither the purpose of the statement, nor the meaning of the statement of truth was 
explained  to  him.  The  Medtronic  Service  Technician  has  informed  Medtronic  that  he  felt  obliged  to  sign  the 
statement as drafted by the police officer, and that he did not appreciate that the statement was intended to stand as 
his sworn evidence in legal proceedings, nor that it would be liable to be relied upon as formal opinion evidence as 
to the functionality of the PM100N device or the data downloaded from the device memory. On a more considered 
reading he has realised that there are certain elements of the statement which may be inaccurate or lost in translation, 
and which we now wish to correct for the Coroner’s record. These are referred to in our response below. 

Save for the  information referred  to  above, Medtronic  was not asked for any further information or evidence  in 
respect of the PM100N device or the downloaded data, nor was it provided with details of the Inquest nor any of the 
evidence to be considered at the Inquest. It was not represented at the Inquest, nor designated an Interested Person. 

The Medtronic Regional Sales Manager re-attended Reading Hospital on 10 January 2024. In the presence of the 
Healthcare  Professional,  he  powered  on  the  machine  and  took  three  photographs  of  the  settings  menu  on  the 
machine. These photographs are enclosed within this document as Appendix A. 

4 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
   
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 Concerns Raised in the Regulation 28 Report to Prevent Future Deaths 

HM Assistant Coroner’s Regulation 28 Report has raised two concerns in respect of the PM100N device as set out 
below: 

1.  The PM100N device: 

a.  Was either not functioning reliably; or 
b.  Was not sufficiently easy to use; and/or 
c.  The alarm is either insufficiently loud to wake the parents; or 
It cuts out automatically before waking the parents; or 
d. 
It did not sound at all. 
e. 

2.  The analysis of the PM100N device has not been accurate, or the device has not correctly recorded the data, 

or the device is not suitable for home use. 

We  note from  HM Assistant  Coroner’s letter  dated  9 October  2023 that  she  did not  conclude that the  PM100N 
device was in fact unreliable, nor that it contributed to the death of Master Turner. 

Background Information (the PM100N device) and investigations undertaken 

Intended purpose of the PM100N device. 

HM Assistant Coroner has been provided with photographs of the PM100N device being used by the Deceased at 
the date of his death. The machine is designed to measure the patient’s pulse rate and the amount of oxygen in the 
blood  (i.e.  oxygen  saturation  levels).  When  either  measurement  goes  below  or  above  a  pre-set  limit  (in  the 
Deceased’s case below 90% (desaturation) or goes above 100%), the monitoring system is designed to warn of this 
by sounding an audible alarm, showing an indicator, and flashing a number. 

The PM100N device is intended for use both in a hospital and (with a slightly adapted range of functionality) a 
home setting in the UK. This enables patients (and their carers) to have the comfort of living in their own homes 
whilst receiving care. From the photographs provided to HM Assistant Coroner by the police, the PM100N device 
was not set to operate in “homecare mode” at the time of death.  This is because the photographs of the PM100N 
device display screen do not show a “house” symbol, which would have been displayed had the home use setting 
been  applied.  Activating  the  “homecare  mode”  setting  is  for  the  healthcare  provider  to  decide  along  with  the 
appropriate settings for the individual patient in order to meet their medical needs (see below). Setting the device to 
“homecare mode” means a patient / user cannot modify the pre-set limits that trigger the alarms, nor can they change 
the  patient  type  selected,  e.g.  paediatric  patient.  The  “homecare  mode”  does  not  prevent  a  patient/user  from 
adjusting the volume level of the alarm or operating the alarm “snooze” function. 

Product information and training provided. 

PM100N devices are supplied by Medtronic to healthcare providers (i.e. hospitals) in the UK. They are supplied 
with an Operator’s Manual and also a Home Use Guide. 

Medtronic offers training on the use of the device to healthcare professionals.  Medtronic does not train individual 
patients/carers on the use of the devices.  We understand that Reading University Hospital provided the device to 
the Deceased’s mother in this case.  Training on the use of the device had been provided to healthcare professionals 
at Reading University Hospital by a Medtronic representative local to the hospital. 

The  training  provided  by Medtronic  includes  details  of  what  is  required to  set  up  the  device including  how the 
settings for an individual patient can be saved through the service menu as an institutional default setting for the 
device. The responsible healthcare professionals at the hospital are then expected to set up the device in accordance 

5 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 with  the  individual  patient’s  requirements,  exercising  their  clinical  judgment  as  to  the  appropriate  settings,  and 
following the Operator’s Manual. 

The healthcare provider is responsible for adjusting the settings on the device before it is provided to a patient or 
carer, so the settings are bespoke to the individual patient’s needs and intended location of use. 

Settings and adjustments 

On 10 January 2024, a photograph was taken for the first time of the “service” settings on the PM100N device (see 
Appendix 1, Photograph 1). These are settings which can only be accessed using an access code. Such codes are 
provided by Medtronic to institutional users only, but not to patients or their carers. Photograph 1 shows that the 
service  settings  in  the  PM100N device  had  been  changed  from  the  original  factory  settings.  It  is  not  known  by 
Medtronic whether or not the service settings have been altered since the death of the Deceased, or whether this 
adjustment took place before the date of death. Medtronic considers it likely that these settings would have been 
altered at the hospital. In any event, they do not accord with the factory settings. 

The treatment of the service settings can only be confirmed by those at the hospital responsible for the custody of 
the PM100N device at all material times, in particular the healthcare provider who set up the PM100N device for 
use by the Deceased. The PM100N device memory data does not include information as to what the settings were 
at any given time. This can only be established by looking at the service settings menu itself. As noted above, the 
status of the settings applied via the service menu was not ascertained prior to the inquest and did not form part of 
the body of evidence considered by HM Assistant Coroner. 

Service Menu Permission to Alter Alarm “priority” and “Power On Settings” 

As indicated above, an institutional user, like a hospital, can change several of the settings for the device’s operation 
for an individual patient from the factory default settings to settings of the hospital’s choice via the settings menu. 
In order to do this, they must have the requisite access code. This code is provided by Medtronic to institutional 
users only, not to patients and/or carers. 

The  product  requirement  document  for  the  device  states  that  it  must  meet  a  minimum  volume  of  45dB  and  a 
maximum volume of 85dB +/-3dB.  The maximum volume setting for the audio alarm is shown on the PM100N 
device display as 8 bars and the minimum of 45dB is shown on the device as 1 bar.  When the devices are supplied 
from the factory by Medtronic to healthcare providers, the audio alarms are set at factory default settings. The audio 
level for any given alarm (whether “low priority”, “medium priority” or “high priority”) in factory settings is “level 
5” (i.e. 5 bars) which is the equivalent of 74.9dB for a high priority alarm, 69.8dB for a medium priority alarm and 
67.1dB for a low priority alarm, according to laboratory tests conducted on the device for the purpose of required 
product standard testing. 

The IEC 60601-1-8 standard governs alarm behaviors of the PM100N device. The devices are tested to this standard 
by specialized labs to ensure that the devices meet the requirements of the standard. 

The  findings  of  the  Medtronic  Service  Technician,  when  undertaking  functional  testing  on  6  July  2022  in  the 
presence of the police, suggest that at least some changes had been made to the settings, after the PM100N device 
had been supplied to the Hospital.  As stated above, the factory default volume setting for the audio alarm is set at 
a level of 5 bars (on a scale of 1-8 bars) by the factory, whereas the Medtronic Service Technician noted on 6 July 
2022 an audio alarm volume setting of 8 bars (i.e. full volume) when the PM100N device was switched on. 

An institutional user, like a hospital, can also re-set the factory settings as to which priority (“low”, “medium” or 
“high”) is allocated to any given audio alarm, provided they have the requisite access code, referred to above. 

6 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 The alarm conditions set by the factory for the device trigger a medium priority alarm whenever the high or low 
pulse rate limits set by the healthcare professional for an individual patient are violated or whenever the high or low 
SpO2 limits set are violated.  At default factory settings this would generate an audio alarm at 69.8dB, which could 
be manually adjusted by a patient/carer (as explained below).  The alarm conditions set by the factory result in a 
high priority alarm being triggered in the event SpO2 loss of pulse occurs.  At factory default settings this would 
sound at level 5 (i.e. 74.9dB).  Based  on the  photographs taken  on 10 January 2024, the alarm  priorities on the 
PM100N device align with factory settings, indicating that they may not have been modified in the service menu. 

It is also possible for an institutional user, holding the requisite access code, to change the “Power On Settings”.  If 
a device is set  to “Factory Defaults”, it will always default to factory settings after each power cycle (i.e. being 
switched “off” and then back “on”).  This can include the audio alarm volume and also alarm priority.  If the “Power 
On Settings” are set to “Last Settings”, the device will always default to the last settings (saved), whatever those 
may be, whenever the device is switched off/on again.  This can include both the audio alarm volume (on a scale of 
1-8 bars) and alarm priority settings (“low”, “medium” or “high”). 

If alternatively, the institutional user sets the device to “Institutional Defaults”, after a power cycle the setting will 
go back to whatever the pre-defined institutional default is that has been set by the hospital or device administrator. 
Once again, this can include both audio alarm volume and alarm priority.  In practice, this means it is open to a 
hospital to set a default audio alarm volume to the maximum 8 bars (this would result in an alarm being triggered at 
82.3dB at “medium” priority alarm setting and 87.6dB at “high” priority alarm setting).  A hospital could also set a 
default alarm priority of “high”.  These defaults could be set for whenever the high or low pulse rate limits set by 
the healthcare professional for the patient are violated or whenever the high or low SpO2 limits set are violated.  In 
both cases, this could then be subject to manual adjustment by the patient / carer during the relevant power cycle. 

Based on the photographs taken on 10 January 2024, the “Power On Settings” on the PM100N device have been 
changed from the factory settings to “Last Settings”.  In practice this would mean that prior to the Deceased’s date 
of  death,  whatever  audio  alarm  volume  setting  had  last  been  applied  (including,  for  example,  through  manual 
adjustment by the user) at the time the PM100N device had last been powered “off”, this would have been the alarm 
volume setting for the PM100N device the last time it was powered “on” for use before the Deceased’s death.  The 
same would apply for the alarm priority setting. 

Service Menu Permission to Deactivate Alarm 

In his witness statement, the Medtronic Service Technician stated that, “the alarm itself cannot be turned off”. This 
is  not  strictly  accurate.  When  the  device  is  in  factory  default  the  user  is  unable  to  turn  the  audible  alarms  off 
completely. The audio alarm function cannot be disabled (i.e. permanently muted) without a code, which Medtronic 
would only provide to the healthcare provider, and not to patients/carers.  To permanently mute the audio alarm the 
user would have to access the service menu (using this code) change the power on settings to “Last Settings” or 
“Institutional Settings” and set “Permission to deactivate audible alarm” to “Yes”. 

The photographs taken on 10 January 2024 indicate that the PM100N device had been changed from factory default 
settings as “Power On Settings” is set to “Last Settings” and the “Permission to deactivate audible alarm” setting is 
set to “Yes”. This means that a user of the machine will, at a point in time which Medtronic cannot determine from 
available data, have changed the settings of the PM100N device from its factory default settings. 

When the “Permission to deactivate audible alarm” setting is set to “Yes”, this means that the user would be able to 
turn off the audible alarms completely. In contrast, when set to “No”, the user cannot completely turn off audible 
alarms and they can instead only be manually turned down to 1 bar (i.e. the lowest alarm volume setting on a scale 
of 1-8 bars). 

7 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
   
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
  
 Both the Operator’s Manual and the Home Use Manual expressly warn users not to silence or decrease the volume 
of the audible alarm if patient safety could be compromised.  It is not known by Medtronic whether or not the Home 
Use Manual was provided for home use in this instance. 

Behaviour of the Alarm 

The audio alarm is intermittent at the frequency seen in the table below (Normal Inter-Burst interval). The Inter-
Burst interval is a standard term used in alarm regulation that describes the initial behaviour of an alarm once it is 
triggered in terms of the number of beeps generated in a fixed amount of time. 

The device has an intelligent alarm system that increases the frequency of the noise bursts after two minutes of the 
alarm being triggered in Normal Inter-Burst interval without acknowledgement. This is shown in the table below 
(Escalated Inter-Burst Interval). 

Volume adjustment for the audio alarm 

Users, including patients and their carers are able to manually adjust the volume of the audio alarm both up and 
down (i.e. between a level of 1 bar to a level of 8 bars), and this can be done simply without needing any code.  The 
Home Use Manual (see excerpt below with our highlighting) explains how the user can turn down the volume: 

There is also a quick reference guide to turning down the alarm volume on the device itself. This can be seen in the 
photographs provided to HM Assistant Coroner by the police. If the volume level is manually turned down by a 
user, it will automatically reset to the pre-set default volume if the device is switched off and on again.  The pre-set 
volume will depend upon the “default” setting applied on the individual device for the particular audio alarm in 
question. 

As indicated above, if an individual device is left at factory default settings, this pre-set volume would be level 5, 
8 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
   
 
   
 
   
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 or a maximum of 69.8dB (when the high/low pulse rate limits were crossed) and 74.9dB (when SpO2 loss of pulse 
occurred). 

As stated above, the photographs taken on 10 January 2024 show that the PM100N device’s “Power On Settings” 
was set to “Last Settings” in the service menu. 

Where “Last Settings” is selected, then the volume level of the audio alarm will revert to whatever was the level 
before the device was switched off and on again (that setting is saved).  This would vary according to the level of 
audio alarm volume selected by the previous user, for example using the manual  adjustment knob, as illustrated 
above in the Home Use Guide. 

The findings of the technical testing undertaken by the Medtronic Service Technician on 6 July 2022 showed the 
PM100N device had an 8-bar volume level when powered on.  Consequently, the factory default settings must have 
been re-set at some point in time prior to that test. The photographs taken on 10 January 2024 confirm that the device 
was set to “Last Settings” default settings.  This would have returned the audio alarm volume to whatever level the 
last user had manually selected, which could be higher or lower than factory settings. This is consistent with the 
volume level observed on the machine when examined on 6 July 2022 which was on the maximum level of “8 bars”, 
after the machine had been switched on for examination. 

Whilst the photographs taken on 10 January 2024 show that the service menu default setting for alarm volume was 
set to “Last Settings”, the raw data downloaded from the PM100N device memory shows that the PM100N device 
was turned off and on again eleven times between the date of the Deceased’s death and the date of functional testing 
by the Medtronic Service Technician on 6 July 2022. Consequently, if the default service menu settings for alarm 
volume had been set to “Last Settings” prior to the Deceased’s death, this would mean that any adjustment made to 
volume, whether intentionally or inadvertently, on any of those eleven occasions, when the PM100N device was 
powered off/on, could have been the volume setting then observed at the date of functional testing by the Medtronic 
Service Technician.  Medtronic does not therefore know whether or not the volume level of the alarm observed by 
the Medtronic Service Technician on 6 July 2022 corresponded to the actual volume level set for the default alarm 
volume at the time of death. 

Device memory and data logging 

The PM100N device generates its data based on data received from a sensor which is attached to the patient. For a 
baby, or small child, the sensor will usually be secured around either the wrist or foot. In order to function, the sensor 
requires a cardiac signal which it picks up from the movement of blood through the body. Once the cardiac signal 
is  detected  and  transmitted  by  the  sensor,  the  device  logs  the  patient’s  pulse  and  oxygen  saturation  every  four 
seconds. 

The PM100N device’s memory logs and stores certain data, but it does not record: (a) whether an alarm has in fact 
sounded at a specific time; or (b) the volume of any audio alarm that may have sounded. There is therefore no data 
memory of when, for how long, and how loudly any given alarm sounded.  A maximum of 88,000 events can be 
stored to memory on the machine. Events are logged every four seconds and therefore the log capacity of the device 
memory is 96 hours in total. There is no reason to believe that the log is incomplete. 

Results of examination and analysis of memory data from the PM100N device 

As stated above, on 25 May 2022 and 6 July 2022, Medtronic personnel downloaded the memory log data from the 
PM100N device in use by the Deceased at the time of his death and provided this to the police.  Additionally, and 
as noted above, on 6 July 2022 a Medtronic Service Technician also completed physical functional checks on the 
machine in the presence of the police.  We understand that the data and findings were all provided by the police to 

9 

 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 HM Assistant Coroner. 

As stated above, visual examination found no cosmetic damage to the monitor, and it was in good condition. The 
PM100N device was powered up, no error messages or failures were displayed on its monitor, and it was ready to 
use. A review of the PM100N device settings on 6 July 2022 found that the alarm volume was set at its maximum 
level of 8 bars (82.3dB) while the pulse and key beep volumes were silenced. This is confirmed in the photographs 
taken on 10 January 2024. 

Alarms were tested and found to be functioning according to the device specification. The device patient mode was 
in paediatric mode and the ‘nurse call’ appeared to function normally (albeit this is a function which can only be 
used with a nurse call infrastructure, usually in a healthcare setting and not in the home). The PM100N device was 
tested with a SpO2 simulator and passed the tests as per manufacturer service instructions. The PM100N device was 
also tested with a spotlight SpO2 functional tester at its maximum values BPM 240 and SpO2 100% and passed all 
tests.  The  PM100N device monitor  was  fully  checked  for functionality  and electrical  safety and passed all  tests 
successfully as per the manufacturer guideline. The conclusions of the Medtronic technical team who reviewed the 
findings and the data downloaded from the PM100N device was that it was functioning correctly and in accordance 
with specification.  We also note in this regard HM Assistant Coroner’s express statement in her letter of 9 October 
2023 that she did not find that the PM100N device was unreliable. 

Response to the Concerns 

We set out below our response to the areas of concern raised by HM Coroner. 

Concern 1: The Medtronic machine was either not functioning reliably or was not sufficiently easy to use 
and/or the alarm is either insufficiently loud to wake the parents or it cuts out automatically before waking 
the parents, or it did not sound at all. 

When the data was downloaded and the PM100N device examined on 25 May 2022 and 6 July 2022, the Medtronic 
Service Technician was able to confirm that the PM100N device was not externally damaged and there were no 
error messages displayed when it was turned on. The PM100N device was installed with the latest software, and the 
battery  was  working  appropriately.  As  set  out  in  detail  above  the  PM100N  device  passed  all  functional  testing 
performed in accordance with manufacturer specification. 

Data “gap” between 25 April 2022 and 3 May 2022 

HM Assistant Coroner raised concerns arising from the apparent lack of data from the machine between 25 April 
2022 and 3 May 2022 during which period the Deceased’s parents stated the PM100N device was in use. The witness 
statement of the Medtronic Service Technician states that he was told by the Healthcare Professional at Reading 
University Hospital that the PM100N device was given to the patient on 25 April 2022. The Medtronic Service 
Technician  has  informed  Medtronic  that  he  has  no  memory  of  the  Healthcare  Professional  being  present  at  the 
download  nor  of  her  providing  such  information  to  him.  He  does  recall  a  lady  at  Reading  University  Hospital 
handing  over  the  PM100N  device  from  storage  for  testing  on  6  July  2022,  but  does  not  know  if  this  was  the 
Healthcare Professional in question.  It is possible that this information was supplied in the draft witness statement 
prepared for the Medtronic Service Technician, but that in the pressure of time to sign the statement, this part of the 
statement was overlooked or misunderstood. 

Medtronic has no information about when the PM100N device was supplied to the parents of the Deceased or about 
the date of first use by the parents. 

A senior technician at Medtronic has conducted a further desktop review of the data downloaded on 6 July 2022 
10 

 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 from the memory of the PM100N device.  That downloaded data does not include data logged before 3 May 2022. 
The reason for this is that a maximum of 88,000 events can be stored to memory on the PM100N device. As stated 
above, events are logged every four seconds and therefore the total memory log capacity at any one time is 96 hours. 
Consequently, by 10 May 2022, any data logged in the PM100N device memory from 25 April 2022 would already 
have been overwritten as oldest data is overwritten first. The absence of logged data in the downloaded date for the 
earlier period up to 3 May 2022 does not therefore indicate that the PM100N device was not in use, not performing 
correctly nor that it was unreliable. Rather it indicates that there was insufficient device memory to retain recorded 
events from before 3 May 2022. 

The witness statement of the Medtronic Service Technician refers to downloading the whole log of the device from 
“15.03.22 to 07.06.22”. This is an error in the statement, stemming from: (a) a typographical error applying the 
number “15” instead of “5”; and (b) the use of United States date convention formatting. Medtronic has since asked 
the Medtronic Service Technician to confirm the correct dates. He confirms that, using the same United States date 
convention format, this should read “05.03.22 to 07.06.22” being 3 May 2022 to 6 July 2022, which is consistent 
with the date range of data obtained from the PM100N device, and the date on which the device data was downloaded 
in front of DC 7712. 

We have again reviewed the data downloaded from the PM100N device from the date of earliest retained recording 
(3 May 2022) to the date of death (10 May 2022). The data collected shows the PM100N device was monitoring the 
Deceased’s pulse and oxygen consistently during this period. 

On the date of death, the downloaded data shows at line 82861 that the Deceased’s oxygen levels started to drop 
below 90% at 01:59:10 and there is reference to desaturation at that point. The oxygen level continued to drop until 
02:14:18 (line A83090) when there was a loss of pulse and no oxygen reading. 

The data thereafter shows that the PM100N device remained on and connected to the Deceased until 04:50:56 on 
10  May  2022  when  the  PM100N  device  was  turned  off.  This  appears  to  coincide  with  the  time  at  which  the 
ambulance was called. 

Operation of the PM100N device audio alarm 

As the device (and the data logged in the device memory) do not specifically log whether or not the alarm sounded, 
nor the volume of the alarm, there is no direct evidence available from the PM100N device’s memory to suggest the 
alarm was not working at the time of death or that the volume was not sufficiently loud.  The fact that the PM100N 
device passed all functional testing and that the alarm functionality was fully operational when the machine was 
switched on again at Reading University Hospital indicates the PM100N device was working in accordance with 
the manufacturer specification. 

As noted above Medtronic does not know at what point the various service menu settings were accessed and changed 
from the original factory default settings. This includes, but is not limited to, the change of the “Power On Settings” 
to  “Last  Settings”, in consequence  of which the  audio  alarm  volume  level  at  the  point the  PM100N  device was 
powered “on” would have defaulted to whatever previous audio alarm level had been saved prior to switching “off” 
the PM100N device.  In addition, Medtronic does not know at what point the “Permission to deactivate audible 
alarm” setting was changed to “Yes”.  Further, Medtronic does not know whether or not a user deactivated the audio 
alarm nor, if so, at what time point this occurred.  The below analysis is therefore set out on the basis of what would 
have occurred had the PM100N device remained in factory default settings. 

Once the oxygen level fell below 90%, this would have triggered an audio alarm until either: (a) the Deceased’s 
oxygen saturation level went above 90%; (b) the oxygen level could not be detected; or (c) the PM100N device was 

11 

 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 turned off. 

The outcome of tests conducted to date indicate that had the device been in factory settings the PM100N device 
would have continued to sound an audio alarm at the volume set on the machine throughout the period from 01:59:10 
until 04:50:56.  As noted above, the data downloaded does not record the sound level of the alarm, and furthermore 
the volume which appeared on first testing the machine in July 2022 cannot be conclusively taken to be the volume 
at which the machine had been set to sound the alarm. This will have depended on the service menu settings, which 
have been changed from factory settings by a user (with the required access code), as well as any adjustments made 
manually to the volume level of the machine by a user.  The alarm can also be temporarily muted manually but only 
for a period of 60 seconds at which point it will then sound again, until manually muted once again. 

As referred to above, the audio alarm functionality could only have been disabled entirely with a specific code. The 
code is intended to be provided to, and kept by, the health care provider only. In the absence of complete alarm 
deactivation, the highest possible volume audio alarm settings for violation of the SpO2 limits and, separately, pulse 
limits set by the Hospital for the Deceased is 82.3dB based on the “medium” priority factory default setting, which 
does not appear to have been changed.  This would have been 87.6dB at the point the “SpO2 loss of pulse” alarm 
was triggered. 

As noted above, the factory default alarm settings for the audio alarm are set to “5 bars” when the device is supplied 
to the healthcare providers though healthcare providers are able to alter the default volume settings as described 
above as appears may have been the case in this instance. The volume of the audio alarm can also be turned down 
manually by a user as explained above in the Home User Manual, which warns against adjusting the volume of the 
alarm to below the level of hearing. 

Concern 2: The analysis of the machine has not been accurate, or the machine has not correctly recorded the 
data, or the machine is not suitable for home use. 

Based on its review of the retained downloaded data and the functional testing performed on the device on 25 May 
2022  and  6  July  2022,  Medtronic  has  not  been  able  to  identify  any  reason  for  doubting  that  the  data  from  the 
PM100N device memory is accurate and that the PM100N device correctly recorded the data. 

Other than the PM100N device time setting running 16 minutes late (which is likely to have been due to an initial 
set up issue by the healthcare provider), there is nothing in the data to indicate that the information is inaccurate or 
was not correctly recorded. 

We set out below examples from the data downloaded where it shows that the PM100N device was recording the 
relevant information: 

Line of Data 

Date 

Time1 

Explanation 

82057 to 82300 

10.5.22 

01:05:36 
01:21:48 

to 

registered  at  various 
Interference 
intervals  which  is  consistent  with  the 
evidence of the mother of the Deceased 
that  she  put  him  back  in  his  crib 
(around  00:56)  and  fed  him  (around 
01:00). 

1 not adjusted for the PM100N device running16 minutes fast 

12 

 
 
 
 
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
  
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Line of Data 

Date 

Time1 

Explanation 

85379 

10.5.22 

04:46:32 

Sensor disconnected. 

85575 

10.5.22 

04:59:56 

the  evidence  of 
Consistent  with 
evidence  of  the  mother  and  father  of 
the  Deceased  that  they  disconnected 
the PM100N device & started CPR. 

PM100N  device  turned  off  according 
to  evidence  of  the  mother  of  the 
Deceased. 

85576 to 85592 

10.5.22 

12:15:20 
12:16:24 

to 

PM100N  device  switched  back  on. 
Please  see  photographs  of  PM100N 
device  taken  by  police, 
timed  at 
12:16:11. 

85593 

10.5.22 

13:24:20 

PM100N  device  was  taken  away  by 
NHS  Community  Nurse  –  possibly 
switched  on 
further 
information. Medtronic have not been 
informed  as  to  whether  or  not  the 
volume was adjusted on this occasion. 

test.  No 

to 

86318 

10.5.22 

14:56:44 
14:57:48 

to 

86348 

87562 

25.5.22 

12:19:08 

6.7.22 

10:29:48 

New readings for oxygen & pulse – it 
is assumed the sensor was attached to 
someone  to  test.  Medtronic  have  not 
been informed as to whether or not the 
volume was adjusted on this occasion. 

Switched  on 
Medtronic 

Switched  on 
Medtronic 

for  download  by 

for  download  by 

The PM100N device has been used safely in both the hospital and the home environment throughout the UK. Prior 
incidents concerning this model of device have been investigated and reported to regulators, including the MHRA, 
as appropriate. An Operator’s Manual is provided to healthcare professionals together with the devices. In addition, 
a Home Use Guide is also made available to healthcare providers which can be provided to patients or their carers 
by treating healthcare professionals.  Medtronic also offers training on the set-up and use of the devices to healthcare 
professionals for use in both hospital and home environments, so they are equipped to set up and use the machines 
and to demonstrate their use to patients and/or their carers. For these reasons, Medtronic believes that the PM100N 
device is suitable for home use. 

13 

 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Conclusion 

Based on the investigations carried out by Medtronic both before and following receipt of the Regulation 28: Report 
to  Prevent  Future  Deaths  dated  18  August  2023  and  the  subsequent  documents  received  on  9  October  2023,  it 
appears to Medtronic that the PM100N device used by the Deceased was functioning appropriately at all material 
times, that it was accurately recording data and was accordingly suitable for home use.  In light of these conclusions, 
Medtronic does not consider that modification or change is required to the device in question. 

12 January 2024 

14 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Appendix A 

Inquest touching the death of Devon Drew Turner. 

Photographs Taken at Reading University Hospital on 10 January 2024 

Photograph 1 – Service Menu 

15 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Photograph 2 – Alarm Priorities 

16 

 
 
 
 
 
 
 
 
 
 
 
 
 Photograph 3 – Alarm Volume 

17
Response from NHS England (PDF)
Katy Thorne KC 
Assistant Coroner  
Berkshire Coroner’s Office 
Reading Town Hall 
Blagrave Street 
Reading  
RG1 1QH 

Dear Ms Thorne 

National Medical Director  
NHS England  
Wellington House 
133-155 Waterloo Road  
London 
SE1 8UG 

18 October 2023  

Re: Regulation 28 Report to Prevent Future Deaths – Devon Drew Turner who 
died on 10 May 2022.  

Thank you for your Report to Prevent Future Deaths (hereafter “Report”) dated 18th 
August  2023  concerning  the  death  of  Devon  Drew  Turner  on  10th  May  2022.  In 
advance of responding to the specific concerns raised in your Report, I would like to 
express my deep condolences to Devon’s parents and wider family. NHS England are 
keen to assure the family and the coroner that the concerns raised have been listened 
to and reflected upon.  

On receipt of your Report, it was shared with colleagues from NHS England’s central 
Patient  Safety  (including  medical  device  specialists)  and  Children  &  Young  People 
teams as well as my regional Quality colleagues in the South East. The Patient Safety 
Team reviewed the concerns raised and have confirmed that they are not aware of 
similar incidents and that there are no known patient safety issues relating to SATS 
machine alarms in general.  

From  your  Report  the  underlying  issue  relates  to  the  specification  of  the  SATS 
machine  used,  its  alarm  settings  and  volume,  whether  it  was  suitable  for  use  in  a 
patient’s  home  and  whether  the  machine  in  this  case  was  working  properly.  These 
issues would fall under the remit of the Medicines and Healthcare products Regulatory 
Agency (MHRA), who are the regulator for all medical devices in the UK, who I note 
that you have also addressed your concerns to.  

We have been in touch with the MHRA regarding the concerns raised in Devon’s case. 
They have advised that they are in receipt of the manufacturer’s investigation report 
into 
the  Medtronic  Nellcor  PM100N  Bedside  SpO2  Monitor  (serial  number 
MBH1920704) that was being used to monitor Devon’s SATS at the time of his death 
and  that  Medtronic  advised  that  they  found  no  fault  with  the  device  during  an 
examination in the presence of the police. The alarm volume was set at its maximum 
while the pulse and key beep volumes were silenced.  I note your letter to Medtronic, 
dated 9 October 2023, stating that you are unable to accept some of the evidence in 
its  entirety,  and  MHRA  is  continuing  to  work  with  the  manufacturer  to  ensure  this 
matter is fully investigated and any identified actions are undertaken.  

MHRA advised that the operator’s manual for the device indicates it is appropriate for 
home  use  when  used  as  an  adjunct  in  patient  assessment  and  is  to  be  used  in 

                                                                                                                       
 
 
 
 
 
 
 
 
 
 
 
 
  
 conjunction  with  clinical  signs  and  symptoms.  There  are  several  warnings  in  the 
operating manual regarding the alarms, including, for example, ensuring the speaker 
is clear of any obstruction to prevent an inaudible alarm tone. Patients should also be 
kept under close surveillance when monitoring as it is possible, although unlikely, that 
radiated  electromagnetic  signals  from  sources  external  to  the  patient  and  the 
monitoring system can cause inaccurate measurement readings. It is advised not to 
rely entirely on the monitoring system's readings for patient assessment. NHS England 
has been asked to be sighted on MHRA’s response to you and will review any further 
considerations from the organisation.  

My  regional  colleagues  in  the  South  East  have  also  been  engaging  with  Berkshire 
Integrated Care Board (ICB) on the circumstances of this case, and NHS England will 
be seeking assurances that any identified local learnings are acted upon. I understand 
that they have been in touch with you separately on some of the concerns raised.  

I would also like to provide further assurances on national NHS England work taking 
place around the Reports to Prevent Future Deaths. All reports received are discussed 
by  the  Regulation  28  Working  Group,  comprising  Regional  Medical  Directors,  and 
other clinical and quality colleagues from across the regions. This ensures that key 
learnings and insights around preventable deaths are shared across the NHS at both 
a national and regional level and helps us pay close attention to any emerging trends 
that may require further review and action.  

Thank you for bringing these important patient safety issues to my attention and please 
do not hesitate to contact me should you need any further information. 

Yours sincerely, 

National Medical Director

Related reports

Other reports by Katy Thorne

See all →

More reports categorised “Child Death (from 2015)”

See all →

Track Child Death (from 2015)

See every Prevention of Future Deaths report matching Child Death (from 2015), and how often a new one appears.

What would an alert for this have sent me? Search the full text

Free to try — the preview shows the real matches and how many arrived in the last 12 months. Your first email alert is free.

These reports are published by the Chief Coroner's office at judiciary.uk and are © Crown copyright. The text here is reproduced from the published PDF so it can be searched. If something on this page is wrong, or you are a person named in it and want it reviewed, email drcjar@gmail.com and we will act promptly.