Kate Robertson 
Assistant Coroner for North Wales (East and Central) 

REGULATION 28 REPORT TO PREVENT FUTURE DEATHS 

THIS REPORT IS BEING SENT TO:   

BOC Limited 

1 

CORONER 

I am Kate Robertson, Assistant Coroner for North Wales (East and Central)                     

2 

CORONER’S LEGAL POWERS 

I make this report under paragraph 7, Schedule 5, of the Coroners and Justice 
Act 2009 and regulations 28 and 29 of the Coroners (Investigations) Regulations 
2013. 

3 

INVESTIGATION and INQUEST 

On 21 December 2020 an investigation was commenced into the death of Ben 
Christopher Harrison following his death on 18 December 2020. The Inquest 
concluded on 10 May 2024 with a narrative conclusion. 

4 

CIRCUMSTANCES OF THE DEATH 

The circumstances of the death are as follows : 

Ben was aged 37 at the time of his death on 18 December 2020. He had known 
psychiatric issues. On 15 December 2020 and whilst a voluntary inpatient at the 
Ablett Psychiatric Unit, Glan Clwyd Hospital (part of the Betsi Cadwaladr 
University Local Health Board ‘BCUHB’) he was found in cardiac arrest with a 
ligature around his neck, 
resuscitated and oxygen cylinder utilised. The cylinder has two valves one on 
the top and one on the side, both of which must be opened before the cylinder 
will function. The valve on the side of the cylinder was not opened and so Ben 
was ventilated only on room air for approximately 5-10 minutes during the 
resuscitation process.  

. He was 

5 

CORONER’S CONCERNS 

During the course of the Inquest, the evidence revealed matters giving rise to 
concerns.   

In my opinion there is a risk that future deaths will occur unless action is taken. 
In the circumstances it is my statutory duty to report to you. 

Coroner's Office, County Hall, Wynnstay Road, Ruthin, LL15 1YN 

 
 
 
 
 
 
  
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 The MATTERS OF CONCERN are as follows.  – 

Evidence was heard during the Inquest that a CD Oxygen cylinder manufactured 
by BOC was used during the resuscitation of Ben. I was shown during the Inquest 
how the oxygen cylinder is operated. 

In order for the cylinder to release oxygen the valve on the side must have its tab 
removed and then the valve itself rotated until it is open with the valve at the top 
also  needing  to  be  opened.  In  Ben’s  resuscitation,  this  did not occur.  The  side 
valve had not been opened meaning that for 5-10 minutes Ben was ventilated on 
room air only. Once it was noted, it was immediately corrected.  

The evidence at Inquest was that having two valves was confusing for users and 
at times of high intensity and highly charged situations, even with training, those 
operating the cylinder may not necessarily recall that there are two valves to open. 
It  is  understood  that  more  pronounced  wording  has  been  included  on  the  side 
valve to attempt to alert users though this is not particularly pronounced.  

There  have  been 22  incidents  with oxygen cylinders  at  the  Health  Board since 
2014, including 2 since January 2024. There has been additional training for staff 
over  recent  years  as part of  their  ALS  /  ILS  training  including  specific  focus  on 
these cylinders and yet issues with the two valves on the cylinder remains. 

It  is  understood  that  BCUHB  have  referred  numerous  concerns  to  BOC  over 
recent  years  and  some  minor  amendments  have  been  made  to  the  cylinders. 
BCUHB  also  reported  to  the  Medicines  and  Healthcare  products  Regulatory 
Agency (MHRA) on 6 October 2022 under The Yellow Card Scheme. No response 
was formally received. 

I remain concerned that the CD Oxygen cylinders manufactured by BOC which, it 
is understood, supply most if not all Health Boards in Wales (under procurement 
processes) are unsafe for use in heightened / pressurised situations in that it is 
not overtly clear how the cylinders are to be operated with the confusion of the 
two valves. This is evidenced by very similar ongoing issues identified by BCUHB 
even with adequate training of staff. 

6 

ACTION SHOULD BE TAKEN 

In my opinion action should be taken to prevent future deaths and I believe you 
have the power to take such action. 

Coroner's Office, County Hall, Wynnstay Road, Ruthin, LL15 1YN 

  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 7 

YOUR RESPONSE 

You are under a duty to respond to this report within 56 days of the date of this 
report, namely 5 July 2024. I, Kate Robertson, the coroner, may extend the 
period. 

Your response must contain details of action taken or proposed to be taken, 
setting out the timetable for action. Otherwise, you must explain why no action is 
proposed. 

8 

COPIES and PUBLICATION 

I have sent a copy of my report to the Family of the Deceased, BCUHB and to 
the Chief Coroner. 

I have also sent a copy of the Report to the following for their information:- 

a.  Eluned Morgan, Health Minister for Wales  
     b.   Medicines and Healthcare Regulatory Agency 
     c.   Health Services Safety Investigations Body 
     d.  NHS Wales Executive 

I am also under a duty to send the Chief Coroner a copy of your response.  

The Chief Coroner may publish either or both in a complete or redacted or 
summary form. He may send a copy of this report to any person who he believes 
may find it useful or of interest. You may make representations to me, the 
coroner, at the time of your response, about the release or the publication of 
your response by the Chief Coroner. 

d 

Dated 10 May 2024 

Signature   
Assistant Coroner for North Wales (East and Central) 

Coroner's Office, County Hall, Wynnstay Road, Ruthin, LL15 1YN 

    |

Responses

1 response published against this report on judiciary.uk. A response is a body's written reply to the coroner's concerns; publication is at the discretion of the Chief Coroner's office, so an absent response does not mean nobody replied.

Response from Boc (PDF)
Ms Kate Robertson 
Assistant Coroner for North Wales (East and Central) 
Coroner’s Office 
County Hall 
Wynnstay Road 
Ruthin 
Wales 
LL15 1YN 

BOC Limited 
Forge 
43 Church Street West 
Woking 
Surrey 
UK 
GU21 6HT   

05.07.2024 

Response to Coroner’s Regulation 28 Report (Ref: 2024-0256) 

Dear Ms Robertson, 

Firstly, may I apologise for the late response to your Regulation 28 Report (Ref: 2024-0256) 

concerning the outcome of the Inquest into the death of Ben Christopher Harrison, on 18 December 

2020. Although we are now aware that you sent the report to BOC, our first sight of the document 

was in an e-mail sent to myself on the 20th of June from the MHRA, whereby a copy of your report was 

attached to the e-mail correspondence. Having received the e-mail from the MHRA, we checked all 

the likely recipients of this type of correspondence in BOC but could not find any evidence of its 

receipt. It would be very helpful if you could indicate to whom you sent your original note so that we 

can investigate as to why it was not forwarded to the appropriate personnel in BOC.  

I appreciate that you are looking for a timely response to the conclusions that you came to during the 

Inquest, but I felt it was important to provide you with some of the regulatory background concerning 

the design and supply of the portable CD Medical Oxygen cylinders fitted with an integral valve. It is 

unfortunate that BOC was not involved in your investigation during the Inquest, as we would have 

been able to provide you with this additional information at that time. 

To provide some history behind the development of the integral valved medical gas cylinder 

packages, BOC initially raised a variation to their long-standing Marketing Authorisation (PL 

0735/5000) back in 2000. The variation covered the supply of Medical Oxygen in a lightweight 

cylinder, fitted with a GCE CombiLite integral valve (referred to as a Valve with Integral Pressure 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 
 
 
 
 
 
 Regulator (VIPR)). The variation was submitted to the Medicines and Healthcare products Regulatory 

Agency (MHRA), who approved the new cylinder package for use.  

The new CD cylinder was designed to replace the existing D size Medical Oxygen cylinder, which 

consisted of a heavy steel cylinder shell fitted with a standard ‘pin index’ medical cylinder valve. This 

package required a separate pressure regulator to be fitted to the cylinders to control the flow of gas 

to the patient.  There were a number of benefits with the introduction of the new package, including 

a significant reduction in the package weight, and the ability to almost double the gas capacity of the 

cylinder by increasing the filling pressure from 137 bar(g) to 230 bar(g) . But the main benefits were 

to provide a cylinder which eliminated the user having to make high pressure connections to the 

valve and not requiring any additional equipment (such as a valve opening key) to enable the 

cylinder to be used for administration to the patient.  

Although there have been several changes to the CD cylinder since its introduction (including further 

increases to the cylinder pressure to increase capacity and adding the CombiLite integral valve (VIPR) 

to different cylinder sizes and products), the basic valve design is the same as the original valve 

fitted to the CD cylinder in 2000.  

The CombiLite valve is manufactured and tested by Gas Control Equipment (GCE) and is CE certified by 

their Notified Body (as a Class IIB medical device) to the Medical Device Regulations (MDR), as being 

compliant with the requirements set out in the relevant ISO standard (BS ISO 10524-3). 

As part of the requirements of the Marketing Authorisation, BOC must provide information to the end 

user of the cylinder package about how to handle and use the cylinder safely. This includes the 

agreed wording for the cylinder labelling, information in the Summary of Product Characteristics 

(SmPC) and the Patient Information Leaflet (PIL), which is designed to provide information to the 

patient in ‘easy to understand’ language. These documents are made available to Healthcare 

Facilities and to Homecare patients to inform them how to safely use the cylinder and to form part of 

the training requirements for its safe use. Note that the safety refers to both the safety of the person 

administering the gas to the patient and the patient safety aspects of oxygen therapy. 

In addition to the safety information that we are obliged to provide on the cylinder label, in the SmPC 

and in the PIL, BOC took the decision to prepare an illustrated ‘Instructions for Use’ (IFU) to provide 

additional information to make sure that sufficient information is made available to the end user to 

operate the valve both correctly and safely.  The IFU (copy attached) is provided ‘on-line’ as a 

suitable training document for both Healthcare Professionals and Homecare patients to access and 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 for Healthcare Facilities to use as additional training material, when training their staff. In addition to 

describing the correct method of operating the VIPR, the IFU also covers the cylinder handling 

requirements to administer oxygen safely.  

BOC also provides videos on-line to help show users how to safely use the product, both to ensure 

that the patient receives the correct gas for therapy and that the risks associated with delivering 

oxygen safely are met.  

It should also be noted that BOC has worked with the relevant UK and European Gas Associations 

(British Compressed Gases Association (BCGA) and the European Industrial Gases Association (EIGA)) 

to prepare ‘best practice’ guidance for both the industry and for end users (via their websites). These 

Guidance Notes were prepared to harmonise the information that the individual gas companies 

prepare to accompany their products and have been reviewed prior to publication by the relevant 

national Regulatory Authorities - the MHRA in the UK and the HPRA in Ireland.  Although the GCE 

valve used in the CD cylinder package is the most common type of VIPR used by BOC in medical gas 

cylinder packages, it is also commonly used by other gas suppliers in the UK and across Europe.  

As part of the certification process when the valve was initially introduced, GCE were required to 

carry out usability testing, to demonstrate that the user would not be ‘confused’ by the new design. It 

should be noted that when the GCE CombiLite valve was first introduced, it was ‘standard practice’ for 

all cylinder packages that a ‘shut-off’ valve would be required to be opened initially, before setting 

the flowrate to be administered to the patient. The development of the new type of cylinder valve 

simplified this process, providing a means to operate the valve without any additional equipment, as 

previously mentioned.  

Since the initial introduction of the integral valve, BOC has placed more than 600,000 medical gas 

cylinder packages fitted with VIPRs onto the UK and Irish markets. To date, BOC has supplied 

approaching 40 million Medical Oxygen cylinders to Healthcare Facilities throughout the UK and 

Ireland, as well as to Homecare patients, providing the opportunity for patients with respiratory 

disease to live a ‘normal’ life. As the cylinders are not supplied specifically for ‘single’ patient use, 

they are usually operated multiple times during each supply, indicating that this type of integral 

valve has been ‘opened’ to supply patients upwards of 200 million times over the last 24 years. 

Although BOC understands the need for the valve to be operated ‘intuitively’, especially during 

emergency situations, it is also important to note that it is imperative that users are appropriately 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 trained in the use of Medical Oxygen cylinders due to the nature of oxygen and the associated risks 

related to the potential for ignitions, if the user instructions are not correctly followed. 

Within the 40 million cylinders supplied, there have been a couple of serious incidents that have 

unfortunately led to the patient being killed or seriously injured. During the Coroner’s Inquests 

covering these incidents, the Coroner has praised BOC for the information they have developed to 

promote the safe use of the product and the relevant mitigation that we now recommend to avoid 

the outcome of an oxygen ignition when administering the gas. In all cases, BOC feels that it is 

imperative that anyone who is required to administer Medical Oxygen should be trained in the 

operation of the cylinder package to ensure that they minimise the risk of an oxygen fire occurring, as 

well as making sure that gas is flowing to the patient when it is being administered. Although the 

potential risks are high, providing the simple rules for the handling and setting up of the cylinder are 

followed, I would stress that Medical Oxygen cylinders have an extremely safe record. 

With respect to the incident that occurred at Glan Clwyd Hospital, I can confirm that BOC were not 

informed of the incident at the time it occurred. In response to the statement made in your 

Regulations 28 Report, where it is mentioned:  

BCUHB also reported to the Medicines and Healthcare products Regulatory Agency (MHRA) on 6 

October 2022 under The Yellow Card Scheme. No response was formally received.  

I can confirm that BOC did receive two separate Yellow Card reports from the MHRA in October 2022. 

Both Yellow Card reports were anonymous, in that the reporter ticked “no” on the report to the 

question “Can we send your personal contact details to the manufacturer”, meaning that BOC were 

not aware as to who had submitted the Yellow Card reports to the MHRA, so we were unable to 

formally respond to the reporter of either report.  BOC did however respond to the MHRA’s Defective 

Medicines Report Centre (DMRC) in response to both Yellow Card reports. 

Some details stated in one of the received Yellow Card reports were, however, identical to 

information that had been reported to BOC by Betsi Cadwaladr University Health Board (BCUHB) via 

email communications in September 2022. This led BOC to suspect that one of the Yellow Card reports 

had been submitted by someone from BCUHB but as no reporter details were submitted on the 

Yellow Card report, we were unable to confirm this, or report back to the reporter.  

Around the time of the two Yellow Card reports, BOC had been in discussion with the MHRA about the 

actions we had already taken, to emphasise the need to open the cylinder valve handwheel prior to 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 
 selecting the prescribed flow for the patient. This included a change to the cylinder valve handwheel 

tamper evident cover (which is fitted as a regulatory requirement over the cylinder valve handwheel 

as a tamper evident seal). The raised letters on the cover were pad printing in black to make it more 

obvious that the cover should be removed before use, making it more obvious that the cylinder valve 

handwheel should be opened to administer the gas.  When BOC supplies medical oxygen to 

Homecare patients, part of the procedure we have to follow is to ensure that either the patient or 

their ’carer’ are fully trained and assessed to make sure they understand the correct procedure to be 

used. With Healthcare Facilities, this is a little more difficult to organise (as we have no indications / 

records of the staff that may be required to use the cylinders). However, BOC does regularly offer a 

variety of training programmes for different aspects of handling Medical Oxygen, as well as offering 

‘free’ training for nursing staff to make sure that they are familiar with the correct procedures. With 

the cylinder package now having been in service for almost 25 years, the responsibility of ensuring 

that all staff are appropriately trained is down to the Healthcare Facility, as they are aware of the 

changes in the staffing levels and their need for retraining. 

From the discussions BOC has had with the staff at Glan Clwyd Hospital, I can confirm that BOC has 

provided them with some training, but this has been related to the management of the Medical Gas 

Pipeline System (MGPS). The training BOC provided was intended for ‘Training the Trainer’, aimed 

primarily at the engineering and portering staff responsible for operating the MGPS. Although it did 

cover some aspects of handling cylinders, this was related to cylinders used to supply the pipeline, 

rather than cylinders used at the patient’s bedside or when transferring patients between 

departments. BOC has offered to provide awareness training for nurses in their ‘mess room’, 

(avoiding any issues of taking nurses away from the wards), but this offer was declined by the 

hospital. 

It is noted in your Regulation 28 Report that the nurses have undergone ALS / ILS training with 

respect to administering Medical Oxygen in emergency situations. It is disconcerting that you also 

indicated that, despite being retrained, nurses are still making the mistake of not turning on the 

cylinder valve handwheel prior to administering the gas. It would be useful if you could provide the 

scope of the Life Support training to see whether it correctly covers the operation of the cylinder 

valve as well as the precautions to be taken when administering Medical Oxygen. BOC’s 

understanding of this training is that an essential part of any procedure involving the administration 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 of Medical Oxygen is that there is a requirement to ensure that a flow of gas has been established for 

the patient. 

As part of the range of cylinders BOC offers to customers, we now have available to Healthcare 

Facilities the IQD Medial Oxygen cylinder, based on the same lightweight cylinder, but fitted with a 

VIPR that has an electronic gauge. The gauge has a number of functions, including visual and audible 

alarms to warn the user when the cylinder has not been set up correctly. One of the alarms provides 

an indication as to when a flow has been selected without first opening the cylinder valve 

handwheel. I understand that BOC has offered this cylinder package to Glan Clwyd Hospital, but they 

have decided not to purchase it. I have attached for your information, a copy of the IFU for the IQD 

cylinder package so that you can see the functionality of the electronic gauge, but you will note that 

the instructions still cover the safe use and handling of Medical Oxygen cylinders. 

I can also confirm that BOC is currently in the process of working with two medical gas valve 

manufacturers who are developing a single operation Medical Gas VIPR, where the cylinder valve 

handwheel and flow selector functionality is incorporated into the same knob/valve. This work is in 

its final stages, with BOC due to conduct testing on the valves to ensure they operate in compliance 

with the relevant ISO standard. However, it should be noted that from a usability point of view, the 

introduction of this valve will need to be carefully controlled as it will require the user to follow 

‘different’ procedures compared to those for the existing valve design. 

As you can see from the information I have provided, BOC attaches great importance to constant 

product evaluation as part of our continuous improvement programme., In this regard, BOC has a 

number of projects currently under way to improve both the safe operation of the valve, as well as 

reviewing and improving patient safety. One issue we have with many users is that incidents with 

valves are not always reported to BOC, and with the MHRA’s Yellow Card procedure, it is not always 

possible to understand the full circumstances related to each incident. The 22 incidents at Glan Clwyd 

Hospital were only reported to BOC verbally, long after the event, and we have received no 

indications of the two cases that you have referenced in your Report has having occurred this year. 

As it appears that the appropriate personnel in BOC did not receive your Regulation 28 Report until 

late June, via the e-mail from the MHRA (which will require a separate investigation to see why your 

official Report was not received), we would request an extension to your time frame before providing 

the exact information you have requested. Although we have provided you with a description of the 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd. 

 
 
 plans that BOC has in place to address the issues raised in your Regulation 28 report, we would need 

more time to provide timescales for some of the actions, hence the request for an extension. It would 

be informative were BOC to have a formal meeting with you to go through our responses to the 

points you have raised in the Regulation 28 Report to clarify our position with respect to this 

unfortunate incident and to ensure that you are satisfied with our responses.  

Please accept my apologies for the lengthy response to your report, but, as you can see, there are 

many aspects related to the use and operation of our Medical Gas cylinder packages. We would 

welcome the opportunity to meet with you to go through the points raised and I look forward to your 

reply to this response. 

Yours sincerely 

Quality & Regulatory Affairs (QRA) Technical Manager 
BOC Limited 

BOC Limited, a subsidiary of Linde plc. Registered in England & Wales, company number 337663.  
Registered office: Forge, 43 Church Street West, Woking, Surrey, GU21 6HT, England.  
The BOC logo and the BOC word are trademarks or registered trademarks of Linde plc or its affiliates. Copyright © 2022. BOC Ltd.

Related reports

Other reports by Kate Robertson

See all →

More reports categorised “Hospital Death (Clinical Procedures and medical management) related deaths”

See all →

Track Hospital Death (Clinical Procedures and medical management) related deaths

See every Prevention of Future Deaths report matching Hospital Death (Clinical Procedures and medical management) related deaths, and how often a new one appears.

What would an alert for this have sent me? Search the full text

Free to try — the preview shows the real matches and how many arrived in the last 12 months. Your first email alert is free.

These reports are published by the Chief Coroner's office at judiciary.uk and are © Crown copyright. The text here is reproduced from the published PDF so it can be searched. If something on this page is wrong, or you are a person named in it and want it reviewed, email drcjar@gmail.com and we will act promptly.