REGULATION 28:  REPORT TO PREVENT FUTURE DEATHS (1)

NOTE: This form is to be usedafter an inquest.

REGULATION 28 REPORT TO PREVENT FUTURE DEATHS

THIS REPORT IS BEING SENT TO:

1. The British Orthopaedic Association
2. Managing Director of Stryker (UK) Ltd

1

CORONER

I am Richard T Middleton, Assistant Coroner, for the Coroner Area of Dorset

2

CORONER’S LEGAL POWERS

I make this report under paragraph 7, Schedule 5, of the Coroners and Justice
Act 2009 and regulations 28 and 29 of the Coroners (Investigations) Regulations
2013.

3

INVESTIGATION and INQUEST

On the 14th December 2023, an investigation was commenced into the death of
Reginal Victor Smith, born on the 11th June 1933.

The investigation concluded at the end of the Inquest on the 10th January 2025.

The Medical Cause of Death was:

1a   Hypovolemic Shock

1b   Re operation of fractured neck of femur

1c    Fractured neck of femur

II

The  conclusion  of  the  Inquest  recorded  that  Reginald  Victor  Smith  died  as  a
consequence of a rare but recognised complication of a surgical procedure.

4

CIRCUMSTANCES OF THE DEATH

On  9/10/23  Mr  Smith  had  a  witnessed  fall  at  his  care  home.  He  was
admitted to Poole Hospital where he underwent surgery to repair a right
extra capsular neck of femur fracture on 12/10/23. The surgery involved
Mr Smith  being laid  on  a traction  table  and a jig  was  used to align the
fracture for screws to be inserted into a titanium nail which is placed into
the femur.

1

 On 24/10/23  he was discharged from hospital. On 5/12/23 he attended
Poole Hospital for a review appointment when X rays disclosed a failure
of  metalwork  inserted  on  12/10/23  and  he  was  readmitted  to  hospital.
On 7/12/23  Mr  Smith  underwent revision surgery.  It was  apparent  that
the hip screw was slightly off centre and being approximately 1mm-2mm
off centre did not  make  proper contact  with  the nail.  Following  surgery
his health deteriorated. Mr Smith received palliative care and he died in
hospital on 7/12/23.

5

CORONER’S CONCERNS

The MATTERS OF CONCERN are as follows:

1) During the inquest evidence was heard that:

i)

  Evidence  was  given  to  suggest  there  were  two  probable
reasons  for  the  hip  screw  not  correctly  passing  into  the
nail:

firstly, before the femoral nail and jig were inserted it may
be that the jig was loose and needed tightening or

secondly, the jig used might have been slightly bent

2) I have concerns with regard to the following:

i)

ii)

iii)

   Each  jig  is  used  many  times  in  surgery  having  been
sterilised  after  each  procedure.  It  is  hammered  into  the
thigh  bone  and  on  this  occasion  may  have  become
deformed over time.

   The  jig  was  sent  away  to  the  manufacturer  for  analysis
but  was  lost  and  so  no  information  was  available  to  the
court in relation to its integrity.

    There  is  no  quality  control  in  place  in  relation  to  the
examination of the jigs being used (other than when it is
assembled in  theatre  by a nurse) prior to surgery.  There
is  no  auditing/  spot  checks  in  relation  to  the  integrity  of
the jigs.

2

 6

ACTION SHOULD BE TAKEN

In  my  opinion  urgent  action  should  be  taken  to  prevent  future  deaths  and  I
believe you and/or your organisation have the power to take such action.

7

YOUR RESPONSE

You are under a duty to respond to this report within 56 days of the date of this
report, by 18th March 2025.  I, the coroner, may extend the period.

Your  response  must  contain  details  of  action  taken  or  proposed  to  be  taken,
setting out the timetable for action. Otherwise you must explain why no action
is proposed.

8

COPIES and PUBLICATION

I  have  sent  a  copy  of  my  report  to  the  Chief  Coroner  and  to  the  following
Interested Persons:

(1)  Mr Smith’s Family

I am also under a duty to send the Chief Coroner a copy of your response.

The  Chief  Coroner  may  publish  either  or  both  in  a  complete  or  redacted  or
summary  form.  He  may  send  a  copy  of  this  report  to  any  person  who  he
believes may find it useful or of interest. You may make representations to me,
the coroner, at the time of your response, about the release or the publication
of your response by the Chief Coroner.

9

Dated

Signed

21st January 2025

Richard T Middleton

3

Responses

2 responses published against this report on judiciary.uk. A response is a body's written reply to the coroner's concerns; publication is at the discretion of the Chief Coroner's office, so an absent response does not mean nobody replied.

Response from British Orthopaedic Association (PDF)
38-43 Lincoln’s Inn Fields
London WC2A 3PE
Registered Charity No. 106694
Company Registration No. 3482958
View our privacy policy at boa.ac.uk/privacy

10th April 2025

Mr Richard T Middleton,
Assistant Coroner for the Coroner Area of Dorset

Sent via e-mail to:

Coroner's Support Officer

Dear Mr Middleton,

Re. Reginald Victor Smith - Prevention of Future Deaths Report (ref: 35392141)

Thank you for granting the British Orthopaedic Association (BOA) an extension to respond to the
Regulation 28 Report concerning the death of Reginal Victor Smith (date of birth 11th June
1933).

Much of the delay in responding has been due to obtaining and viewing the X-rays available.

The scope of the information contained within the Regulation 28 Report and the available
imaging is not sufficient to review the procedure carried out by the surgeon.

We note the matters of concern raised in the Report on the evidence received by the inquest
suggest there were two probable reasons for the hip screw not correctly passing into the nail:





firstly, before the femoral nail and jig were inserted it may   be that the jig was loose and
needed tightening or
secondly, the jig used might have been slightly bent

We also note the specific concerns raised that:

1. Each jig is used many times in surgery having been sterilised after each procedure. It is
hammered into the thigh bone and on this occasion may have become deformed over
time.

2. The jig was sent away to the manufacturer for analysis but was lost and so no

information was available to the court in relation to its integrity.

3. There is no quality control in place in relation to the examination of the jigs being used
(other than when it is assembled in theatre by a nurse) prior to surgery. There is no
auditing/ spot checks in relation to the integrity of the jigs.

All implants are supplied with an accompanying document setting out the recommended
operative technique. The document offers guidance that the surgeon should heed but, as with
any such technical guide, each surgeon must consider the particular needs of each patient and
make appropriate adjustments when and as required.

 The BOA has drafted generic advice for trauma and orthopaedic surgeons on the need for
vigilance as to the condition and preparation of any jig used, as well as adherence to the
operative technique documentation accompanying each implant.

This draft guidance is currently being reviewed by the BOA and pertinent sub-specialty societies
before being published. The BOA will of course send you a copy. It is envisaged that the guidance
will be ready for publication by the end of May.

The BOA will also draft guidance as to the number and orientation of intraoperative imaging and
post-operative ‘check’ X-rays.

Once again, the BOA will of course send you a copy upon publication. It is envisaged that this
guidance too will be available by the end of May.

Do please contact me should you require any additional information.

Yours sincerely,

Head of Policy and Public Affairs
Response from Stryker UK Ltd (PDF)
The Coroner’s Office for the County of Dorset 
Civic Centre, Bourne Avenue 
Bournemouth, BH2 6DY 

Per email: coroner.service@bcpcouncil.gov.uk 

26 March 2025 

Dear Mr. Middleton, 

Re: Regulation 28 Report, REF: 35392141 

I write in response to the Regulation 28 Report to Prevent Future Deaths dated 21 January 2025, REF: 35392141, 
issued  by  the  Coroner  for  the  County  of  Dorset  (“the  Report”).  Stryker  UK  Ltd  (“Stryker”)  has  taken  the 
opportunity to carefully review the Report, investigated the reported adverse event and considered the Matters 
for Concern which were raised therein. The following outlines Stryker’s position and response to the concerns 
identified.  

I. 

Use of the Gamma Nail Distal Targeting System 

Concern Raised:  

Each jig is used many times in surgery having been sterilized after each procedure. It is hammered into the thigh 
bone and on this occasion may have become deformed over time.  

Stryker’s Response: 

The Distal Targeting System (“Targeting System”) is an instrument designed to assist healthcare professionals in 
the  precise  application  and  implantation  of  compatible  Stryker  implants,  such  as  the  Long  Gamma  Nail.  The 
instrument  should  only  ever  be  used  by  licensed  healthcare  professionals  who  are  qualified  by  appropriate 
training  methods  and  are  fully  familiar  with  the  instruments  intended  use  and  all  the  applicable  surgical 
techniques.  

The  Targeting  System  is  specifically  designed  and  intended  for  repeated  use.  Its  major  components  are 
comprised of carbon fiber composite material, ensuring high strength, durability, resistance to deformation and 
long-term  structural  integrity.    When  the  instrument  is  cleaned,  sterilized  and  maintained  according  to  the 
Instructions  for  Cleaning,  Sterilization,  Inspection  and  Maintenance  (“Maintenance  Instructions”),  it  remains 
undamaged and retains its shape, and its serviceable lifespan is not affected.  

Importantly and contrary to the concern raised, the Targeting System is not hammered into place. The Operative 
Technique  (“OT”)  explicitly  states  that  the  instrument  itself  is  not  designed  to  be  struck  and  notes  that  the 
insertion of the Long Gamma Nail into the femoral canal should progress smoothly, without excessive force. The 
only instance where a force may be applied is if dense bone is encountered and sufficient reaming has already 
been confirmed. In such instances, a specific strike plate is attached to the Nail Holding Screw component of the 
instrument and a designated slotted hammer may be used, with caution. Striking the Targeting System directly 
is  against  protocol,  and  the  OT,  Instructions  for  Use  (“IFU”)  and  Maintenance  Instructions  all  warn  that 
hammering the instrument may result in damage or breakage, rendering it unsuitable for further use.  

The  OT,  IFU  and  Maintenance  Instructions  require  the  healthcare  professional  to  ensure  and  verify  that  all 
components function correctly with each other and are free from damage. The Maintenance Instructions provide 

 
 
 
  
 clear directions for performing a functional check, stating that if any failure or damage is detected, the instrument 
must be replaced and must not be used as it may have reached the end of its serviceable life.  

Additionally, the OT and IFU emphasize that throughout the procedure, the instrumentation must be repeatedly 
checked  to  ensure  correct  angles,  optimal  alignment  and  secure  connections  between  the  implant  and  the 
instrument, to allow for proper and precise positioning and fixation of the implant.  

The  Targeting  System  is  designed  for  durability  and  proper  function  when  used  as  intended.  The  provided 
guidelines clearly outline correct handling, maintenance, and functional checks to prevent the device from being 
used in the unlikely event it has been damaged. This ensures the Targeting System remains fit for purpose when 
used according to the provided guidance.  

II. 

Location of the Distal Targeting System 

The following concern was raised: 

The jig was sent away to the manufacturer for analysis but was lost and so no information was available to the 
court in relation to its integrity. 

Stryker’s response: 

Stryker confirms that the Targeting System has not been returned to the manufacturer for analysis. Immediately 
upon  notification  of  the  adverse  event,  Stryker  initiated  a  Product  Investigation  (ref:  PI  3492425)  and 
coordinated with the hospital for collection of the instrument. Stryker was informed that the Kit containing the 
Targeting System was placed in the trauma loan instrument storeroom for retrieval, as per the standard process 
for the collection of devices. Despite Stryker making arrangements for the Kit’s collection, it could not be located 
in  the  relevant  storeroom.  Stryker  worked  with  the  hospital  in  searching  for  the  Kit  and  following  several 
attempts at locating the Kit and investigating its potential whereabouts, it remained unaccounted for.  

III.  Quality control of the Distal Targeting System 

The following concern was raised: 

There is no quality control in place in relation to the examination of jigs being used (other than when it is assembled 
in theatre by a nurse) prior to surgery. There is no auditing/spot checks in relation to the integrity of the jigs.  

Stryker’s response: 

Stryker  confirms  all  devices  and  instrumentation  undergo  vigorous  quality  control  measures  during 
manufacturing to ensure they are free from design, material or manufacturing defects. These stringent quality 
controls ensure that only compliant products are placed on the market.  

All instrumentation, including the relevant Targeting Device, is placed with the hospital under a Consignment 
Agreement. This Agreement explicitly assigns responsibility for appropriate maintenance, handling, inspection, 
sterilization  and  storage  of  the  instrumentation  to  the  hospital.  This  includes  pre-use  inspections  by  the 
operating surgical team.  

Stryker  remains  committed  to  supporting  hospital  and  healthcare  professionals  by  ensuring  access  to 
instrumentation in good working condition. However, as outlined in Section I above, the OT, IFU and Maintenance 
Instructions of the instrument outline the necessity of a functional check before each procedure and indicate that 
the responsibility for inspecting and verifying the integrity of the instrumentation prior to its use lies with the 
hospital and healthcare professionals using the device. These documents provide clear guidance on necessary 

 
 
 
 
 pre-use checks and quality controls, emphasizing that any instrument showing signs of wear or damage must not 
be used and should be replaced.  

In this case, according to the feedback received by Stryker when carrying out the investigation, all necessary 
checks  were  carried  out  and  no  deformation  or  damage  was  observed  on  the  relevant  Targeting  System. 
Additionally, prior to the revision surgery, a Stryker representative removed the Kit containing the Targeting 
System and carried out an inspection of the instrumentation, finding no quality or functional issues. 

Conclusion 

Stryker is committed to patient safety and product integrity. The Targeting System is designed and manufactured 
to withstand repeated use when handled and maintained according to the provided guidelines. The OT, IFU and 
Maintenance Instructions outline the necessary precautions, handling and inspection procedures to ensure the 
safe and successful use of the instrument.  

Stryker  acknowledges  the  concerns  raised  but  does  not  propose  any  additional  actions  at  this  time.  The 
safeguards in place, as outlined above, provide sufficient risk mitigation. Furthermore, the reported incident rate 
remains an extremely rare occurrence, with only 12 similar adverse events reported globally since 2013, despite 
over 2.6 million lag screws being distributed and implanted successfully in that same timeframe.  

Stryker  appreciates  the  opportunity  to  review  this  matter  and  provide  our  feedback.  We  remain  open  to 
continued engagement and are committed to upholding the highest standards of medical device safety.  

Should you have any questions in relation to this letter, do not hesitate to contact me.  

Sincerely,  

Aman Auluck 
Associate Manager, PMS 
Stryker UK, Ireland, and Northern Europe 

Stryker House, Hambridge Road 
Newbury, Berkshire, RG14 5AW 
Tel: +44 1635 262 476 
nby_qara@stryker.com 
www.stryker.com 

A AuluckElectronically signed by: A AuluckReason: I approve this documentDate: Mar 31, 2025 14:29 GMT+1 
 
 
 
 
 
 
 
 
 
 
 
 
 REF35392141 - Stryker Response - PI 3492425
and PI 3492428

Final Audit Report

2025-03-31

Created:

2025-03-31

By:

Status:

A Auluck (Aman.Auluck@stryker.com)

Signed

Transaction ID:

CBJCHBCAABAAhBe7Wlld4mzOS75FSNIXCI638-PYa2db

"REF35392141 - Stryker Response - PI 3492425 and PI 349242
8" History

Document created by Aman Auluck (Aman.Auluck@stryker.com)

2025-03-31 - 1:28:27 PM GMT- IP address: 176.25.79.12

Aman Auluck (Aman.Auluck@stryker.com) authenticated with Adobe Acrobat Sign.

Challenge: The user opened the agreement.

2025-03-31 - 1:29:08 PM GMT

Signer Aman Auluck (Aman.Auluck@stryker.com) entered name at signing as A Auluck

2025-03-31 - 1:29:45 PM GMT- IP address: 176.25.79.12

A Auluck (Aman.Auluck@stryker.com) authenticated with Adobe Acrobat Sign.

Challenge: The user completed the signing ceremony.

2025-03-31 - 1:29:46 PM GMT

Document e-signed by A Auluck (Aman.Auluck@stryker.com)

Signing reason: I approve this document

Signature Date: 2025-03-31 - 1:29:47 PM GMT - Time Source: server- IP address: 176.25.79.12

Agreement completed.

2025-03-31 - 1:29:47 PM GMT

Related reports

Other reports by Richard Middleton

See all →

More reports categorised “Hospital Death (Clinical Procedures and medical management) related deaths”

See all →

Track Hospital Death (Clinical Procedures and medical management) related deaths

See every Prevention of Future Deaths report matching Hospital Death (Clinical Procedures and medical management) related deaths, and how often a new one appears.

What would an alert for this have sent me? Search the full text

Free to try — the preview shows the real matches and how many arrived in the last 12 months. Your first email alert is free.

These reports are published by the Chief Coroner's office at judiciary.uk and are © Crown copyright. The text here is reproduced from the published PDF so it can be searched. If something on this page is wrong, or you are a person named in it and want it reviewed, email drcjar@gmail.com and we will act promptly.