Prevention of Future Deaths reports · 2020

Shanté Turay-Thomas

Regulation 28 report to prevent future deaths, reference 2020-0124, written 27 Jan 2020. A coroner writes one of these when an inquest reveals a risk that could cause further deaths unless something changes.

Date of report27 Jan 2020
Reference2020-0124
DeceasedShanté Turay-Thomas
CoronerMary Hassell
Coroner areaInner North London
CategoryEmergency Services related deaths · Community health care · Other related deaths
Sourcejudiciary.uk record · original PDF
Responses published9

The report

Text extracted from the PDF text layer. Reproduced verbatim, including the scan's own layout.

Regulation 28:  Prevention of Future Deaths report 

Shanté Andreé Marie TURAY-THOMAS (died 15.09.18) 

THIS REPORT IS BEING SENT TO: 

1.

2.

3.

4.

5.

Partner
Winchmore Hill Practice
808 Green Lanes
Winchmore Hill
London N21 2SA

GP Member and Chair
Enfield Clinical Commissioning Group (CCG)
116 Cockfosters Road
Barnet
London EN4 0DR

Head of Global Manufacturing
Bausch & Lomb UK Ltd
106 London Road
Kingston Upon Thames
Surrey KT2 6TN

Chief Executive
Medicines & Healthcare Products Regulatory Agency
(MHRA)
10 South Colonnade
Canary Wharf
London E14 4PU

Chief Executive
London Central & West Unscheduled Care Collaborative
(LCW UCC - NHS 111 service provider)
St Charles Hospital
Exmoor Street
London  W10 6DZ

1 

 Chief Executive 
Advanced Health & Care Ltd 
(Adastra 111 clinical patient management system provider) 
Ditton Park 
Riding Court Road 
Datchet 
Berkshire SL3 9LL 

Chief Medical Officer 
London Ambulance Service NHS Trust 
220 Waterloo Road 
London  SE1 8SD 

Chair 
Association of Ambulance Chief Executives (AACE) 
30 Great Guildford Street 
London SE1 0HS 

6. 

7. 

8. 

9. 

Chief Executive 
NHS Digital 
1 Trevelyan Square 
Boar Lane 
Leeds LS1 6AE 

10. Professor Stephen Powis 
National Medical Director 
NHS England & NHS Improvement 
Skipton House 
80 London Road 
London SE1 6LH 

11. Sir Andrew Dillon 
Chief Executive 
National Institute for Health & Care Excellence (NICE) 
10 Spring Gardens 
London SW1A 2BU 

12. The Rt Hon Matt Hancock MP 

Secretary of State for Health & Social Care 
Department of Health & Social Care 
39 Victoria Street 
London SW1H 0EU 

2 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 1 

CORONER 

I am:   Coroner ME Hassell 
           Senior Coroner  
           Inner North London 
           St Pancras Coroner’s Court 
           Camley Street 
           London  N1C 4PP 

2 

CORONER’S LEGAL POWERS 

I make this report under the Coroners and Justice Act 2009,  
paragraph 7, Schedule 5, and  
The Coroners (Investigations) Regulations 2013, 
regulations 28 and 29. 

3 

INVESTIGATION and INQUEST 

On 18 September 2018, I commenced an investigation into the death of 
Shanté Andreé Marie Turay-Thomas. The investigation concluded at the 
end of the inquest on 13 January 2020. I made a narrative determination 
made at inquest, a copy of which I attach. 

4 

CIRCUMSTANCES OF THE DEATH 

Shanté was allergic to nuts and on 18 September 2018 told her mother 
that she had eaten hazelnuts.  She died soon after of acute anaphylaxis. 

5 

CORONER’S CONCERNS 

During the course of the inquest, the evidence revealed matters giving 
rise to concern. In my opinion, there is a risk that future deaths will occur 
unless  action  is  taken.  In  the  circumstances,  it  is  my  statutory  duty  to 
report to you. 

The MATTERS OF CONCERN are as follows.  

1.  At the time of her death, Shanté was not receiving specialist care 
for her allergies.  However, her general practitioners (GPs) failed 
to appreciate this.  They assumed that she was being treated for 
her  allergies  at  the  transitional  asthma  clinic  to  which  she  had 
been  referred  following  her  paediatric  discharge.    This  was  an 
incorrect assumption.   

3 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 The GPs had not identified Shanté (who had a high BMI and was 
severely  allergic)  as  being  at  particularly  high  risk  from  her 
allergies and asthma, and had no awareness that they were the 
sole providers of Shanté’s allergy care. 

2.  Shanté’s  GPs  knew  that  she  should  carry  two  adrenaline  auto 
injector (AAI) pens at all times, and they may have mentioned this 
to her, but they failed to record this and they did not emphasise it 
to her.   

They failed to emphasise to Shanté and her family that the reason 
for carrying two pens is primarily because in the event of severe 
acute anaphylaxis, the very strong likelihood is that both pens will 
need to be administered, one five minutes after the other, to keep 
the patient alive until the arrival of an emergency ambulance.   

The  GPs  did  not  explore  with  Shanté  the  reason  for  her erratic 
requests for a pen.  They did not explore with her where she kept 
her pens.  They did not test her understanding of medical advice. 

3.  The Emerade AAI accompanying leaflet does include the advice 
that two pens should be carried at all times, but the advice is not 
re-iterated on the outside of the box.  Consideration will need to 
be given to whether this is the appropriate advice in all cases, but 
it seems worthwhile to review the issue as a whole.   

(I  assume  the  same  is  true  of  the  EpiPen  and  the  JEXT,  but  I 
heard no evidence about these at inquest.) 

4.  The Emerade AAI is sold singly.  It could be sold in boxes of two 

as the norm and only singly in the alternative.  

5.  When Shanté’s AAI was changed from an EpiPen to an Emerade, 
her GPs failed to reconsider the prescription and to increase her 
dose from 300mgs to 500mcgs.  

6.  Following the scriptswitch, the GPs failed to ask Shanté to come 
in to the surgery for training in use of the Emerade.  This would 
also  have  presented  an  ideal  opportunity  to  explore  Shanté’s 
understanding  of  the  use  of  her  pens  and  to  ensure  that  she 
understood she needed to carry two at all times. 

7.  The  GPs  relied  upon  the  advice  given  by  Enfield  Clinical 
Commissioning Group (CCG) that the scriptswitch was simply the 
replacement  of  one  branded  product  with  another  branded 
product of the  same drug/device.  This gave false reassurance.  
The  CCG  joint  formulary  committee  introduced  a  new  drug  for 
GPs, but then gave the wrong advice to accompany this.  

4 

 
 
 
 
 
 
 
 
 
 
 
 
 8.  The  CCG  failed  to  draw  prescribers’  attention  to  the  need, 
following scriptswitch from EpiPen to Emerade, to reconsider the 
dose and to prescribe the higher dose of 500mcgs for patients at 
higher risk (which would have included Shanté). 

9.  The  CCG  failed  to  inform  prescribers  that  the  Emerade  pen 
requires different training to the EpiPen because different AAIs do 
not operate in the same way.  In fact, the CCG gave the opposite 
advice. 

10. I did not hear evidence that there is any NICE (National Institute 
for Health &  Care  Excellence) guidance  on  the  point.   It  seems 
that the whole area would benefit from NICE review. 

11. It would also benefit from review by NHS England, particularly in 
terms of the guidance given to CCGs and regarding consideration 
of incentives (CQUIN indicators etc.) to educate and train in the 
management of allergies. 

12. The  Emerade  AAI  (and  I  assume  the  EpiPen  and  JEXT)  leaflet 
does  not  specifically  advise  that  training  from  a  healthcare 
professional  is  needed  in  how  to  use  this  particular  AAI  as 
opposed to any other. 

13. I heard that the gold standard of training for use of any AAI is to 
give  the  patient  the  relevant  pen  (whichever  that  patient  is 
prescribed)  containing  a  placebo  rather  than  adrenaline  and, 
following  appropriate  instruction,  ask  the  patient  actually  to 
administer a dose.   

I heard at inquest that the incidence of this standard of training (in 
any setting) is rare.  That may be for good reasons, but it seems 
that  revisiting  best practice  training  at  a national level would  be 
helpful. 

14. When  Shanté  became  ill  following  the  ingestion  of  nuts,  her 
mother rang NHS 111 and got through to the London Central & 
West  (LCW)  service.    However,  the  call  handler  incorrectly 
recorded Shanté’s location: he failed to untick a box and so her 
grandmother’s address was recorded as her location, rather than 
her mother’s address where she was staying at the time.   

In an example of good practice, this error was recognised by the 
clinician who later took over the call.  However, what nobody at 
LCW realised  was  that  the  Adastra  computer  system  would  not 
then update in real time for any screens save that of the particular 
clinician inputting the information.   

5 

 
 
 
 
 
 
 
 
 
 
 
 The staff at LCW have since been made aware of this and have 
been trained to walk over and look at the primary screen to check 
the address, but it is not clear to me that there is now a national 
understanding of that element of the system. 

15. During the course of the 111 call, a number of errors were made.  
These were the errors of LCW individuals.  When LCW audited 
the call in the first instance, the audit identified the problem with 
the address, but failed to recognise how badly the call had gone 
in other ways.  Without effective audit and recognition of failings, 
it is difficult to see how there can be effective improvement. 

16. The  individuals  making  these  errors  were  working  within  the 
context of NHS Digital’s categorisation of anaphylaxis as needing 
a category 2 ambulance rather than a category 1 ambulance, on 
the Adastra computer system that supports the LCW 111 service.   

This was the wrong categorisation and not the categorisation that 
the  call  would  have  received  if  999  had  been  called  and  the 
London Ambulance Service contacted in the first instance.  Acute 
anaphylaxis  is immediately  life  threatening  and  must  be  treated 
as a category 1.   

I  heard  at  inquest  that  NHS  Digital  has  since  changed  its 
categorisation.  However, I also heard that for those areas (I think 
approximately half the country, though this is not completely clear 
to me), where the 999 service and the 111 service are supported 
by different computer systems rather than the same system being 
common to both services, there could remain inconsistencies of 
categorisation between 999 and 111.   

Even where there are inconsistencies in categorisation, the 999 
service  will  not  re-categorise 
following  a  111  clinician’s 
categorisation, unless a 999 clinician has spoken to the patient, 
so  inappropriate  111  categorisation  will  not  be  safety  netted  by 
the 999 service.  This must be recognised and factored in. 

17. In terms of national training for 111 call handlers, the NHS Digital 
distance learning pack contains advice that is in part inadequate 
and in part wrong.  It does not give the crucial information that one 
dose of adrenaline, by whichever device it is administered, is very 
unlikely  to  be  sufficient  in  the  case  of  acute  anaphylaxis.    It 
contains a photograph to illustrate the use of an AAI, but in the 
photograph the device is held incorrectly. 

18. I  am  unclear  as 

the  Adastra  111  algorithm 
automatically  prompts  administering  a  second  AAI  five  minutes 
after the first if there has been no improvement, but it should. 

to  whether 

6 

 
 
 
 
 
 
 
 
 
 
 19. One of the errors made by the first 111 call handler was a failure 
to  ask  to  speak  direct  to  the  patient.    This  was  the  error  of  an 
individual.   

However, this is not the first time that the issue has been brought 
to  the  attention  of  NHS  Digital.    At  inquest, I  asked the  witness 
who  appeared  on  behalf  of  NHS  Digital,  and  indeed  had  been 
chosen by NHS Digital as the person best able to assist the court, 
if this had been an issue in the past.  He said no.  However, on 18 
December  2018,  Peter  Harrowing,  HM  Assistant  Coroner  for 
Avon,  sent  a  prevention  of  future  deaths  report  to  NHS  Digital 
following the inquest touching the death of David Longden.   

It was only when I asked the witness appearing on behalf of NHS 
Digital specifically about Coroner Harrowing’s report in respect of 
Mr Longden, pointing out that Coroner Harrowing had raised the 
need for NHS Digital to place greater emphasis on the call handler 
speaking to the patient, that the witness remembered that he had 
indeed seen that report.   

I choose to characterise this as a memory lapse rather than as an 
intention wilfully to mislead the court.  (A witness who lies whilst 
giving evidence on oath at inquest may be found in contempt of 
court  and  may  even  be  prosecuted  for  the  crime  of  perjury.)  
Nevertheless, if NHS Digital does not have a grasp of this sort of 
detail,  specifically  brought  to  its  attention  by  a  coroner  in  a 
prevention of future deaths report, it is difficult to see how there 
can be effective improvement. 

20. The  issues  within  this  prevention  of  future  deaths  report  are 
predominantly national issues, but I heard at inquest that there is 
no  person  with  named  accountability  for  allergy  services  and 
allergy provision at NHS England or the Department of Health as 
a whole. 

6 

ACTION SHOULD BE TAKEN 

In  my  opinion,  action  should  be  taken  to  prevent  future  deaths  and  I 
believe that you have the power to take such action.  

7 

YOUR RESPONSE 

You are under a duty to respond to this report within 56 days of the date 
of this report, namely by 16 March 2020.  I, the coroner, may extend the 
period. 

7 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Your  response  must  contain  details  of  action  taken  or proposed  to  be 
taken, setting out  the timetable for action. Otherwise you must explain 
why no action is proposed. 

8 

COPIES and PUBLICATION 

I have sent a copy of my report to the following. 

  HHJ Mark Lucraft QC, the Chief Coroner of England & Wales 
  Professor Chris Whitty, Chief Medical Officer (CMO) for England 
  Care Quality Commission for England (CQC) 
  Dr Clare Dollery, Executive Medical Director, Whittington Health 
  Professor Adam Fox, allergist (independent of Shanté’s care) 
 

 mother of Shanté Turay-Thomas 

I  am  also  under  a  duty  to  send  the  Chief  Coroner  a  copy  of  your 
response.  

The Chief Coroner may publish either or both in a complete or redacted 
or summary form. He may send a copy of this report to any person who 
he  believes  may  find  it  useful  or  of  interest.  You  may  make 
representations to me, the Senior Coroner, at the time of your response, 
about  the  release  or  the  publication  of  your  response  by  the  Chief 
Coroner. 

9 

DATE                                                 SIGNED BY SENIOR CORONER 

27.01.20 

8
Also filed under 2020-0124: 2020-0124-Association-of-Ambulance-Chief-Executives_Redacted.pdf
Association of Ambulance Chief Executives 
Metal Box Factory (GG322) 
30 Great Guildford Street 
London 
SE1 0HS 

E:  

T:  020 7118 0977 

W:  www.aace.org.uk 

3 March 2020 

BY EMAIL:  

M E Hassell 
Senior Coroner 
Inner North London 
St Pancras Coroner’s Court 
Camley Street 
London 
N1C 4PP 

Dear Mr Hassell 

REGULATION 28 REPORT:  ACTION TO PREVENT FUTURE DEATHS:  TURAY-THOMAS 

I am writing in response to the Regulation 28 report to prevent future deaths following the inquest 
into the death of Shanté Andreé Marie Turay-Thomas which you issued on 27th January 2020 to 
Anthony Marsh, Chair of the Association of Ambulance Chief Executives (AACE). We would like to 
clarify that Martin Flaherty is the Managing Director of AACE and is responding on behalf of 
Anthony Marsh and the Association. In addition, we have also liaised with both our National 
Medical Directors Group and NHSE. 

To clarify, AACE is a private company owned by the English Ambulance NHS Trusts. It exists to 
provide ambulance services with a central organisation that supports, coordinates and implements 
nationally agreed policy. Our primary focus is the ongoing development of the English ambulance 
services and the improvement of patient care. We are a company owned by NHS organisations 
and possess the intellectual property rights of the JRCALC UK ambulance service clinical practice 
guidelines. AACE is not constituted to mandate or instruct ambulance service however we do have 
national influence via the regular meetings of ambulance Chief Executives and Trust Chairs along 
with a network of national specialist sub-groups. 

I have read and considered your report. Although no specific action was requested of AACE the 
observations within section 16 of your report are those about which I can offer comment. 

In line with the evidence you heard, at the time of this incident there was an inconsistency in the 
way in which AMPDS and NHS Pathways categorised anaphylactic shock. Whilst the technicalities 
of how the respective triage systems operate would be more appropriately commented on by 
others, it is worth noting that the tools have fundamentally different architecture and methods of 
operation. They are always likely to produce differing outcomes however we have worked closely 
with NHS England and other partners to reduce the variation as far as possible. 

Specifically, in relation to anaphylactic shock we have previously made representation to NHS 
England through the now disbanded Ambulance Response Programme Delivery Group (ARP DG) 

Chairman:  Professor Anthony C Marsh QAM SBStJ DSci (Hon) MBA MSc MA FASI 
Managing Director:  Martin Flaherty OBE 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 and the Emergency Call Prioritisation Advisory Group (ECPAG) requesting that the clinical coding 
inconsistency between AMPDS and NHS Pathways should be resolved as a matter of urgency. 
ECPAG is the NHS England led group responsible for the governance, control and approval of any 
change to clinical code sets (aligning codes to response categories). NHS England subsequently 
tasked the Clinical Coding Group (a sub-group of ECPAG administered by NHS England) to 
resolve the inconsistency.  As you heard through evidence submitted to you at the inquest, the 
inconsistency was resolved, and NHS Pathways now has a route through which severe and 
sudden onset anaphylaxis is categorised as a Cat 1 response. 

Your report makes the observation that inconsistency may remain in parts of the country where 
NHS Pathways is in use by the 111 provider and AMPDS is in use by the ambulance trust. I do not 
believe that is the case and, having consulted with clinical and operational colleagues within AACE, 
I cannot conceive of a circumstance where an incidence of anaphylaxis would be categorised as 
Cat 1 by the 111 provider but result in a different categorisation by the ambulance trust.  Once 
categorised by 111 incidents are passed directly to the Computer Aided Dispatch (CAD) system of 
the ambulance trust bypassing any further call handling or other intervention. The incident would 
present as a Cat 1 to the ambulance despatcher who would allocate an ambulance response. 

Therefore, your subsequent observation that 999 call handlers should ‘safety net’ the 111 triage is I 
think answered by my clarification that the remaining inconsistency that you believed there to be 
between 111/999 system does not in fact exist in relation to anaphylaxis. More generally it is true to 
say that, as a matter of NHS England policy, 999 call handlers do not routinely re-triage (or ‘safety 
net’) triage conducted by 111. This policy was introduced through the Ambulance Response 
Programme (ARP) governance in order to ensure that patients originating in 111 did not end up in 
a repetitive ‘triage loop’ which would disadvantage them through delaying the dispatch of an 
ambulance resource. Ambulance trusts agreed with NHS England that this would be inappropriate 
and that 111 triage should be accepted as a given with incidents deemed as requiring an 
ambulance response being placed directly in the 999-dispatch system by 111. 

Incidents originating either in 999 or 111 may be subject to additional clinical assessment by 
clinicians in ambulance control rooms but this would be conducted on the merits of the incident 
irrespective of the origin of the call. This would typically be conducted where an ambulance service 
has a queue of incidents that are awaiting an ambulance to become available to dispatch. In those 
circumstances clinicians may review the queue of incidents to judge whether any may be suitable 
for clinical telephone assessment or to establish further information about the patient’s condition 
which in some cases may result in the category of the incident being changed. Although it is 
possible that Cat 1 incidents could be subjected to additional clinical assessment, the speed of 
dispatch to Cat 1 incidents means that it is very unlikely to happen in practice. 

I trust you will accept that I have considered your report carefully and have addressed those issues 
which I am able to and particularly those articulated in Item 16. 

Yours sincerely 

Martin Flaherty OBE 
Managing Director 

Chairman:  Professor Anthony C Marsh QAM SBStJ DSci (Hon) MBA MSc MA FASI 
Managing Director:  Martin Flaherty OBE

Responses

9 responses published against this report on judiciary.uk. A response is a body's written reply to the coroner's concerns; publication is at the discretion of the Chief Coroner's office, so an absent response does not mean nobody replied.

Response from Advanced (PDF)
Coroner ME Hassell 
Senior Coroner 
Inner North London 
St Pancras Coroner's Court 
Carnley Street 
London 
N1C 4PP 

8th February 2020 

, 

J I  P 

trwr  11p. S  1,mons. 

> Ditton Park 

Riding Court Road 
Oatchet 
Berkshire 
SL 3 9LL 

> +44 (0) 8 45  160 6162 
www oneadvanced.com 

Dear Ms Hassell, 

Re:  Prevention of Future Deaths report - Shante Andree Marie Turay-Thomas 

Thank you for your Regulation 28 Report to prevent future deaths, detailing your concerns arising from 

the inquest into the death of Miss Shante Turay-Thomas. 

I would  like  to  assure you  that all  matters related  to  patient safety  are  taken  extremely  seriously  by 

Advanced.  We employ a  highly skilled  Clinical Team who  ensure that  we  fully comply  with the  NHS 

requirements on clinical safety in the manufacture of software (DCB 0129). 

Correction (section 18) 

I would like to make a correction to section 18. The algorithms embedded into the Adastra application 

are  the  NHS  Pathways  111  algorithms,  and  Advanced  have no editorial  control  over the  content  of 

these  algorithms.  Thus  your comments  around  the  second  administration  of adrenaline need  to  be 

directed to the NHS Pathways team. 

Errors made in collecting the address (section 14) 

The  Adastra  application  allows  the  collection  of two  address  fields,  a  home  address  and  a  current 

location address. The home address is checked against the Personal Demographic Service (PDS), the 

national  record  of every  patient's  demographic  information. This  includes  the patient's  NHS  number 

and registered General Practitioner's details. 

The error made by.the call handler was not in failing to untick the boxes, as they should have remained 

ticked to record the correct home address, but in realising this address was not the same as the current 

location of the patient.  If the patient is not at the registered home address the current location address 

can be recorded, and this is where the ambulance will be sent if one is required. The Adastra application 

allows  the  address  fields  to be  changed,  but  the  system  has  to  depend  on  the  user ensuring  the 

recorded address details are correct with the caller.  It is important that the patient's registered home 

address is recorded as this allows access to the Summary Care Record (SCR) which will often provide 

information  on  allergies  and  drugs  taken - which  would  have  been  very  useful  in  this  case.  It  also 

allows the post event message to be sent to the GP, who can then provide onward care where needed. 

Advanced H03'1h and Care l 1m 1ed (trad ng os Advanced) 1s o compllny registered ,n Eng•and and Wa'es ln!9r oompany number 2939302, Whoso reg,sterod 
office ,. o ,non Park, R,d,ng Court Road  Oatchet. SL3 9LL 

 Access to the  PDS was developed into Adastra  in  2008  and there  has  been  no changes  to  how this 

functionality works  since.  During  that  time,  we  estimate  that  over  100  million  calls  have  been  taken 

through Adastra,  resulting  in  an  estimated  12 million requests  for ambulance despatch.  The Adastra 

application  sent the ambulance request to the correct address as displayed. There are  already alerts 

in the application so that when performing a 111  assessment using the NHS  Pathways algorithms, if a 

point is reached where an ambulance despatch is needed, the address is displayed in a pop up screen. 

The user must confirm with the caller that this is correct, before ticking to confirm the address is correct, 

to enable the electronic despatch request message to be sent.  If the application cannot determine the 

correct ambulance  service to  send  the  case  to,  or if the  address  details  are  not sufficient or clear,  a 

second  pop  up message is displayed. This pop up occupies almost the whole screen,  and cannot be 

cleared from the screen, until the user has confirmed what actions they have taken to resolve the issue. 

These have been in the application from 2013 and were introduced to try and avoid this situation arising 

{ie ambulances being sent to the wrong address) when the functionality was designed. 

This Is the flrsttlme Advanced has been made aware of the scenario outlined In this case.  This involved asecond 

user being asked verbally by the call handler to contact the ambulanee service and details from a different screen 

(that  was  not  the  screen  in  live  use)  were  used.  Within  two-working  days  of  being  made  aware  ofthe 

scenario outlined In this case, Advanced Issued a reminder to all call  handlers  not to manually circumvent the 

software process and this was followed by a software update being released to prevent this manual override re-

occurring ie a message is displayed if a second user opens the same user case, warning them that the address 

details may be out of date.  This was deployed to all customers. 

Subsequently,  and as a direct result of this case, we are currently making two changes to the Adastra 

application which  are  in  development.  These  changes  will  ensure  that the user will  have  to  actively 

select,  and  tick,  the  address  field  that  is  required  when  there  are  differences  between  the  locally 

recorded  home address and  the  PDS  registered  home address. Also if any changes are  made to the 

address during the course of the consultation they will  be saved to the database immediately and will 

then  be  available  contemporaneously  to  any  other  user,  on  any  other  screen.  We  anticipate  these 

changes will be available to all Adastra customers by the end of April 2020. 

I hope that these measures we are undertaking provide you with the appropriate level of assurance in 

relation to our commitment to clinical safety. 

There are two further areas we would like you to consider: 

Review of the NHS  111  Ambulance Request message 

As part of his verbal statement to the inquest, Dr Alex Yeates spoke of the message structure used by 

Adastra  and  all  ambulance  system  suppliers  which  was  developed  by  the  Health  and  Social  Care 

Information  Centre  in  2013.  This  message  structure  works  well  and  has  been  used  by  Adastra  to 

request many millions of ambulances, but it could be improved. One of the issues raised at the inquest 

was  that  when  LCW contacted the  London  Ambulance  Service,  they could  not find  the  case  on  the 

ambulance  system  as  they  did  not know the  CAD  reference  number for  it.  It  would  also  have been 

useful ifthere had been electronic updating of LCW to inform them that the initial Category II ambulance 

had been recalled. Advanced has spoken to several ambulance system suppliers and we are all happy 

to work with  NHS Digital to develop this standard further,  but it requires NHS  Digital to lead. 

Advanoed Computer Software Group  1  1 
Ditton Park, Riding Court Road, Oatchel, SL3 9LL. 

· 
l ·m'ted -15  8  oompany registered ,n England and  a es u  er comp 

W  I 

nd 

any  number 05965280  whose registered office is 

· 

 Independent review of clinical triage systems 

One of the points raised at the inquest was the inconsistency in categorisations between AMPDS used 

by the London Ambulance Service for their 999 calls, and the NHS Pathways clinical decision support 

triage system used by the 111  service. There are alternative clinical decision support triage systems in 

use within other urgent care settings. The recent Care Quality Commission recommendations published 

in their 'Getting to the  right care  in the  right way - digital triage  in  health services'  report,  states that 

there should be a 'fair test of clinical performance of digital triage solutions'.  We would like to see you 

recommend  that this  is  carried  out and  conducted  by an  independent body  and  not  NHS  Digital,  the 

providers of the NHS 111  Pathways system, as this could be seen as a conflict of interest. 

Yours sincerely 

ordon Wilson 

Chief Executive Officer> Advanced 

Advancod Compu1or Softwaro Gl0UP  Limited is a company registered in England 811d  Wales under company numbor OS96S280,  who$O rogisteted office is 
DittOn Park, Riding Court Road,  Dolchot, SL3 9LL.
Response from Bausch Lomb (PDF)
Arnold &Porter 

F AO Ms Mary Hassell 
Senior Coroner 
Inner North London 
St Pancras Coroner' s Court 
Carnley Street 
London NlC 4PP 

16 March 2020 

Our Ref: 
Your Ref: 

Dear Madam: 

Regulation 28:Prevention of Future Deaths Report, Shante Andree Marie Turay-
Thomas 

This  letter is  sent on  behalf of Bausch &  Lomb UK Ltd  (Bausch)  in  response to your 
report dated 27 January 2020 under paragraph 7, Schedule 5, ofthe Coroners and Justice 
Act 2009 and regulations 28  and 29 of the Coroners (Investigations) Regulations 2013. 

You  raise  four  matters  of  concern  in  relation  to  Bausch's  product,  Emerade  at 
paragraphs 3, 4,  12 and  13  of your report.  We respond to these below. 

Review of advice to  patients that two  pens should be carried at all times 

Paragraph 3 of your report states: 

"The Em erade AA! accompanying leaflet does include the advice that two pens 
should be carried at all times,  but the advice is not re-iterated on the outside of 
the  box.  Consideration will need to  be given to  whether this is the appropriate 
advice in all cases,  but it seems worthwhile to review the  issue as a whole. 

(I assume the same is true c~fthe  EpiPen and the JEXT,  but I heard no evidence 
about these at inquest.)" 

[]  Arnold  &  Porter Kaye Scholer (UK) LLP 

25  Old  Broad  Street. Tower 42  I  Lon don , EC2N 1 HQ  I  United  King dom  I  www.arnoldporter.com 
Arnold  & Porter  Kaye  Scholer (U K)  LLP  is  a  limited  liability  partn ership org anised  under the  laws  of the  State  of  New  York 
(F040202000739),  is  authorised an d regula ted  by the Solicito rs  Reg ulation Authority under ID 25938 2 and  practises as Arnold 
& Porter in  England and  Wales . Our professional rules  may be  accessed  at www.sra .org .uk . It is  an  affiliate of Arnold & Porter 
Kaye  Scholer LLP . a limited  liability pa rtnersh ip organised  under the laws  of th e State  of Delaware. A list of the fi rm's part ners 
and  thei r professional  qualificat ions  is open to  inspectio n at  the web  add ress  above.  The  VAT  registration  number of Arn old 
&  Porter Kaye  Scholer (UK)  LLP is  GB 697 436 088 . 

EU  210567506v1 

 Arnold  Porter 

16 March 2020 
Page 2 

Bausch believes, consistent with advice issued by  MHRA 1
,  that the available evidence 
supports  a  recommendation  that patients  who  are  at  risk  of anaphylaxis  should  carry 
two  adrenaline  auto-injectors  at  all  times  and  this  is  stated  in  the  Patient Information 
Leaflet for  Emerade.  In  contrast however,  guidance  issued  by  the  British Society  for 
Allergy and Clinical Immunology ("BSACI") states: 

"There  is  no good evidence that issuing two AAls is  necessary or cost-effective 
in  most cases. After an  episode  in  A&E,  awaiting proper risk assessment,  the 
normal practice would be to  issue one device" 2

. 

Bausch supports a review of the recommendations  issued by expert bodies to  consider 
whether a consistent approach is possible in the interests of patients. 

With respect to  information on the labelling of medicinal products, this is controlled by 
the Human Medicines Regulations 2012 and the addition of any text must be approved 
by the competent regulatory authority before it can be implemented.  When considering 
the addition of information not specified in  the  regulations, it  is  necessary  to  take  into 
account  whether  this  can  be  included  on  the  packaging  in  legible  form  without 
adversely impacting the essential information already required to be present. 

Despite the current inconsistency of views held by MHRA and the BSACI, Bausch will 
initiate discussions with  MHRA  in  relation  to  whether it  would  be  appropriate to  add 
advice  regarding  the  need  to  carry  two  pens,  similar  to  that  included  in  the  Patient 
Information Leaflet for Emerade, on the outer packaging of adrenaline auto-injectors. 

Emerade supplied singly or in  twin-packs 

Paragraph 4 of your report states: 

"The  Emerade AA] is sold singly.  It could be  sold in  boxes oftwo as the  norm 
and only singly in the alternative. " 

Emerade,  like other adrenaline auto-injectors  is  authorised for  supply as  twin-packs as 
well as single auto-injectors. 

There are a range of circumstances in which a single pen, rather than a twin-pack, may 
properly be prescribed and dispensed for a particular patient including: 

https://www .gov .uk/drug-safety-update/adrenaline-auto-injectors-updated-advice-after-

european-review 

2 

Ewan  P  et  al.  BASCI  Guideline:  Prescribing  an  Adrenaline  Auto-injector.  Clinical  and 
Experimental Allergy 2016; 46(10):  1258 

EU  210567506v1 

 Arnold&Porter 

16  March 2020 
Page 3 

•  Where the patient needs to  replace only a single pen which has been lost, 

damaged, used or has reached the end of its shelf-life; and 

•  Where the patient wishes to keep a further pen at,  say, school (in addition to 

the pens carried with him/her). 

Furthermore, while guidance consistently recommends the prescription of at  least two 
adrenaline  autoinjectors  for  patients with multiple risk factors  for  severe anaphylaxis, 
the position for other patients is  less clear. 

In circumstances where there will  always be a need for single adrenaline auto-injectors 
to  be  dispensed  in  some  situations  and  where  the  number  of auto-injectors  supplied 
depends on the prescription and not the pack size,  Bausch & Lomb suggests that, even 
if  packs  of  two  auto-injectors  are  available,  there  will  continue  to  be  a  parallel 
requirement for routine supply of packs containing a single auto-injector. 

Finally,  supply  of a  twin-pack containing two auto-injectors  may  not prevent  patients 
who should carry two auto-injectors, dividing the pack and keeping the pens in separate 
locations. Supply of a twin-pack cannot therefore replace the need for patient education 
in relation to the number of pens to  be carried at all times. 

Reference  to  training  by  a  healthcare  professional  in  the  patient  information 
leaflet 

Paragraph  12  of your report states: 

"The  Emerade  AA!  (and  I  assume  the  EpiPen  and JEXT)  leaflet  does  not 
specifically advise that trainingfrom a healthcare professional is needed in how 
to  use this particular AA! as opposed to  any other". 

However,  the  wording  of the  Patient  Information  Leaflet  for  Emerade  includes  the 
following  information,  specifically  advising  patients  that  training  from  a  doctor  or 
pharmacist is  required: 

"Always use  Emerade  exactly as your doctor  has  told you.  You  should check 
with your doctor or pharmacist ifyou are not sure". 

This  reflects  the  information  in  the  Summary  of Product Characteristics, which  is  the 
"datasheet"  provided for healthcare professionals: 

"All patients who  are prescribed Emerade  should be  thoroughly  instructed to 
understand the  indications.for the use and the correct method ofadministration 

EU  210567506v1 

 Arnold&Porter 

16  March 2020 
Page4 

(see  section 6. 6 [Which provides detailed instructions on administration]).  It  is 
strongly  advised  also  to  educate  the  patient's  immediate  associates  (e.g. 
parents, caregivers,  teachers) for the correct usage ofEmerade in case support 
is  needed in the emergency situation". 

The use of trainer pens 

Paragraph  13  of your report states: 

"I heard that  the  gold standard of training for  use  of any AA! is  to  give  the 
patient  the  relevant  pen  (whichever  that  patient  is  prescribed)  containing  a 
placebo rather than adrenaline and, following appropriate  instruction, ask the 
patient actually to  administer a dose. 

I heard at inquest that the  incidence ofthis standard oftraining (in  any setting) 
is rare.  That may be for good reasons,  but it seems that revisiting best practice 
training at a national level would be  helpful". 

Bausch distributes trainer pens  to  allergy  clinics and  further  supplies may be accessed 
through the  Emerade website  by  all  patients for practice purposes.  These trainer pens 
do  not include a placebo as this would limit use of the trainer pen to  single use and,  by 
administering an apparent "dose", would present a risk of confusion, with implications 
for patient safety. 

Bausch is  currently reviewing  the  design of its  trainer pens and plans to  incorporate a 
needle  cover  shield  extension  when  activated, 
to  more  closely  replicate  the  patient 
experience with the actual pen, albeit still  capable of being used  on multiple occasions 
for practice purposes. 

We  trust  that  this  letter  addresses  the  concerns  raised  in  your  report  in  relation  to 
adrenaline auto-injectors in general and Emerade in particular. Please contact us should 
you require any  further information. 

Yours faithfully, 

Arnold & Porter Kaye Scholer (UK) LLP 

EU  210567506v1
Response from Department of Health and Social Care (PDF)
From Helen Whately MP 
Minister of State for Care 

39 Victoria Street 
London 
SW1H 0EU 

020 7210 4850 

4 June 2020 

Our Ref: PFD-1202428 

Mary Elizabeth Hassell 
HM Senior Coroner, Inner North London 
St Pancras Coroners Court 
Camley Street 
London N1C 4PP 

Dear Ms Hassell 

Thank you for your correspondence of 27 January 2020 to Matt Hancock about the death 
of Shanté Turay-Thomas.  I am replying as my Ministerial portfolio covers Long Term 
Conditions, including allergies and I am grateful for the additional time in which to respond.      

I would like to start by saying how very sorry I was to read of the circumstances around the 
death of Shanté.  Her loss, at such a young age, must be extremely distressing for her 
family and loved ones and I offer my heartfelt condolences to them.  I agree that it is 
essential that we learn from this tragic event to prevent future deaths.   

Your report raises several serious matters of concern and has been sent to a number of 
health-related organisations at local and national level that have a role to play to ensure 
the safety of people with allergies.  My reply will focus on the response to your concerns 
by national level organisations.  However, I want to make clear that the failings relating to 
primary care and urgent unscheduled care services in north London as described in your 
report are unacceptable.  I expect the NHS to thoroughly review the circumstances of 
Shanté’s death and your findings and take robust action to put in place learnings to 
prevent such deaths from occurring again. 

Departmental officials have made enquiries with the organisations to which you issued 
your report to understand the system-wide response to the matters of concern.  I am 
assured that your concerns have been carefully considered.  I will not repeat the detail of 
the responses you will receive which are the responsibility of others.  However, I wish to 
highlight the actions being taken at a national level that are relevant to the concerns you 
have raised.  

I welcome the action agreed by the Commission on Human Medicines (that provides 
expert, independent advice to the Medicines and Healthcare products Regulatory Agency 
(MHRA)) to convene an Expert Working Group (EWG) on Adrenaline Auto-Injectors (AAIs) 
that will provide independent clinical advice on matters relating to the safe and effective 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 use of AAIs, including several of the issues highlighted in your report such as AAI product 
labelling; training in the use of AAIs; and whether AAIs should be available to the market in 
dual packs as the norm. 

In addition, the AAI EWG will help to inform a planned communication campaign being 
developed by the MHRA to raise awareness of the key messaging on the safe and 
effective use of AAIs.  As you will know, the key messages are:  

•  The need for patients to carry two AAIs with them at all times; and,  

• 

In cases of acute anaphylaxis, to administer a second AAI five to fifteen minutes 
after the first if there has been no sign of improvement.   

It is vital that these messages are understood by patients, their carers’ and healthcare 
professionals to prevent future tragedies.  

As to whether dual packs of AAIs should be available as the norm, I note that all AAI 
brands in the UK are authorised to supply AAIs in single or dual packs.  Companies are 
not required to market all pack configurations and at present, only Epipen is available to 
the market as a dual pack.  The MHRA intends to write to the manufacturers of Jext and 
Emerade to encourage the marketing of dual packs.   

While the MHRA agrees that the availability of AAIs in packs of two would support the aim 
of ensuring that patients always have access to two AAIs, there are important 
considerations, such as the need for flexibility.  The MHRA will seek the expert advice of 
the AAI EWG on this matter.  

On the emergency response to cases of anaphylaxis, I am advised that a clinical coding 
review, conducted in May 2019, led by NHS England and NHS Improvement (NHSEI), 
concluded that symptoms suggesting life-threatening anaphylaxis should receive a 
Category 1 response.  This change came into effect for NHS Pathways (used by NHS 111 
providers and around half of ambulance trusts in England) in October 2019 and equally 
applied to the Medical Priority Dispatch System (MPDS) used by other ambulance trusts.  I 
am advised that NHSEI has oversight of both the NHS Pathways and MPDS clinical 
decision support systems and the NHSEI-led Emergency Call Prioritisation Advisory Group 
provides governance and approval of any change to clinical code sets.  

In relation to NHS Pathways specifically, I am advised that there are established 
governance systems in place to capture, review and resolve issues relating to clinical 
coding (including from Prevention of Future Deaths reports); and to ensure latest clinical 
advice and guidance is reflected in call categorisation.  Independent clinical scrutiny of 
NHS Pathways is provided by a National Clinical Governance Group that includes 
representatives of medical Royal Colleges.  

The commissioning and management of general allergy services is a local matter.  For 
most patients (around 95 per cent) allergic diseases can be managed by primary or other 
non-specialist allergy services with routine therapies.  Approximately five per cent of 
patients with allergies require treatment in a secondary service, and of those, around 

 
 
 
 
 
 
 
 
 
 
 
 20,000 (0.1 per cent) require referral to a specialist centre.  These specialist services are 
commissioned at a national level by NHSEI to an agreed delivery specification.   

The National Institute for Health and Care Excellence (NICE) develops evidence-based 
guidance to support clinicians in managing allergy and related disorders.  Guidance is 
routinely subjected to an evidence surveillance exercise to establish whether an update is 
available.  However, guidance will be reviewed and updated at any time if important new 
evidence comes to light.    

NIICE has published clinical guidelines in this area covering food allergy in children and 
young people under 191; drug allergy2; anaphylaxis3; eczema4; and asthma5.  In addition, 
NICE has produced a range of technology appraisals for drugs and interventions to relieve 
symptoms of these conditions as well as diagnostic assessment tools. 

You will know from NICE’s response to your report that information is already contained in 
the British National Formulary (BNF) and the BNF for children (BNFc) that two AAIs should 
be prescribed; that patients should carry two AAIs at all times; and the importance of 
training in the use of the particular AAI prescribed.  Prescribers are expected to be alert to 
safety information in relation to the medicines they prescribe and to refer to the BNF and 
BNFc when making prescribing decisions with their patients.  NICE will consider how it can 
make more clear in its guideline on Anaphylaxis: assessment and referral after emergency 
treatment, the advice that two AAIs should be prescribed and that patients should carry 
two AAIs with them at all times.   

In relation to the training of primary healthcare professionals in the management of 
allergies, curricula and training is available through the relevant professional bodies (such 
as Royal College of Physicians and Royal College of General Practitioners (RCGPs)) and 
organisations, such as the British Society for Allergy & Clinical Immunology (BSACI), that 
provide educational, training and research resources.  I am advised that NHSEI has 
undertaken to bring the matters of concern in your report to the attention of Health 
Education England (HEE) and the RCGPs to consider what action might be taken to 
strengthen training in this area and where possible, NHSEI will help to facilitate the uptake 
of any new guidance or resource.    

We recognise the key role played in the wider primary care workforce by dietitians, to 
whom people with a variety of nutritional issues, including food allergies and intolerances, 
can be referred.     

Last year, a system of Primary Care Networks (PCNs), each formed from different GP 
practices working together, was established across England.  They are designed to bridge 
the historic divide between primary and community health services and provide an 

1 https://www.nice.org.uk/guidance/cg116 

2 https://www.nice.org.uk/guidance/cg183 

3 https://www.nice.org.uk/guidance/cg134 

4 https://www.nice.org.uk/guidance/cg57 

5 https://www.nice.org.uk/guidance/ng80 

 
 
 
 
 
 
 
 
                                                           
 
 
 
 
 
 extended range of community-based services.  PCNs are recruiting extra dietitians and 
other specialists, with funding being made available through an Additional Roles 
Reimbursement Scheme.  The NHS Long Term Plan6, published last year, has guaranteed 
a minimum investment of £4.5billion extra per year for primary medical and community 
care by 2023/24.  On top of this, an extra £1.5billion has been committed to general 
practice over the next four years for growing the workforce by 6,000 doctors and 26,000 
other staff.  

In relation to accountability for allergies, my Ministerial portfolio covers long term 
conditions, including allergies and the provision of allergy services.  Ministers are held to 
account by Parliament and that accountability is supported by several directorates within 
the Department of Health and Social Care that have relevance to policy on allergens.  This 
includes primary care; population health; oversight of the Food Standards Agency (FSA); 
environmental hazards and NHSEI sponsorship.  Although there is no single, named 
individual with oversight of all aspects of allergy policy, individuals and teams work closely 
together in the Department on all aspects of policy relating to allergies.   

Following the reforms initiated by the Health and Social Care Act 20127, NHSEI is 
responsible for clinical policy and strategy in the NHS in England (including for allergies) 
and NHSEI is held to account through the annual NHS mandate8.  NHSEI has a clinical 
reference group (CRG) for Specialised Immunology and Allergy Services9, that provides 
advice on specialised services and commissioning policies and quality standards. The 
CRG is chaired by a consultant immunologist, specialising in clinical Immunology and 
allergy and expert clinicians are among the membership.  I hope this provides assurance 
of the clinical leadership and policy oversight of allergy services.  

Although not directly relevant to the matters of concern in your report, you may wish to 
note that as part of its food hypersensitivity strategy, the FSA, which is responsible for 
protecting the health of the public in relation to food, is exploring how it can build a better 
picture of allergic consumer experience.  This includes exploring how to develop a 
reporting link that a range of stakeholders (e.g., consumers, businesses and potentially 
medical professionals) can use to facilitate the identification of emerging trends and where 
necessary, alert local authorities so that they can take appropriate investigative and 
enforcement action.   

To support this project and increase the data available to the FSA on food-related cases of 
anaphylaxis, including deaths and near fatal incidents, the Department is working to 
identify means of access to relevant data sets so they can be included for analysis.  

6 https://www.england.nhs.uk/long-term-plan/ 

7 http://www.legislation.gov.uk/ukpga/2012/7/contents/enacted 

8 
https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/803114/accountability-
framework-to-nhse-and-nhsi-2019-to-2020.pdf 

9 https://www.england.nhs.uk/commissioning/spec-services/npc-crg/blood-and-infection-group-f/f06/ 

 
 
 
 
 
 
 
                                                           
 
 
 
 
 Finally, I am aware that this is not the first time you have issued a Prevention of Future 
Deaths report following the inquest into the death of a child where there are concerns 
about the use of AAIs and the emergency response to cases of acute anaphylaxis.  

I am assured that national health organisations are taking important action to ensure the 
safety and quality of health services related to allergies and my response has referred to 
some of that work.  However, given the potential for systemic learning, Departmental 
officials have brought your concerns to the attention of the Healthcare Safety Investigation 
Branch (HSIB) that conducts independent investigations of patient safety concerns in 
NHS-funded care across England, to ask if it will consider the matters of concern in your 
report and whether these in part, or in whole, meet its criteria for national investigation.  
Due to the Covid-19 pandemic, HSIB’s consideration of this referral will be conducted 
when the situation allows. 

I am grateful to you for bringing these matters to my attention.  I hope this reply is helpful.  

HELEN WHATELY
Response from Enfield CCG (PDF)
Enfield CCG Medicines Management Team Action Update: Prevention of Future Death Notice - Shanté Andreé Marie TURAY-THOMAS 

Prevention of Future Death Notice Matters 
of Concern  

CCG Actions taken 

27th August 2015 - North Central London Joint Formulary Committee (NCL JFC) agreed to 
replace Epipen with Emerade on the joint formulary. 

7). The GPs relied upon the advice given by 
Enfield Clinical Commissioning Group (CCG) 
that the scriptswitch was simply the 
replacement of one branded product with 
another branded product of the same 
drug/device. This gave false reassurance. The 
CCG joint formulary committee introduced a 
new drug for GPs, but then gave the wrong 
advice to accompany this. 

1.  The  narrative  verdict  was  discussed  at  the  NCL  Medication  Safety  Officer  (MSO)  Local 
Network  on  17th  January  2020.  The  network  agreed  that  a  Medicines  Safety  Bulletin  on 
Adrenaline Auto Injectors (AAIs) would be distributed to GPs and other primary care healthcare 
professionals. The NCL Medicines Safety Bulletin on Adrenaline Auto Injectors (AAI) dated 24th 
January 2020 is attached as appendix 1. The bulletin was approved virtually by NCL Medication 
Safety Officer (MSO) Local Network following the meeting on 17th January 2020 and distributed 
to Enfield GP practices on 30th January 2020.  

The  Prevention  of  Future  Death  notice  will  be  discussed  at  the  next  NCL  Medication  Safety 
Officer (MSO) Local Network on 4th April 2020 to consider if there is anything further that should 
be done.  

failed 

8).The  CCG 
to  draw  prescribers’ 
attention  to  the  need,  following  scriptswitch 
from  EpiPen  to  Emerade,  to  reconsider  the 
dose  and  to  prescribe  the  higher  dose  of 
500mcgs  for  patients  at  higher  risk  (which 
would have included Shanté). 

9).The  CCG  failed  to  inform  prescribers  that 
the Emerade pen requires different training to 
the  EpiPen  because  different  AAIs  do  not 
operate  in  the  same  way.  In  fact,  the  CCG 
gave the opposite advice. 

2.  The  Prevention  of  Future  Deaths  report  has  been  shared  with  the  Chair  of  the  NCL  Joint 
Formulary  Committee  (JFC)  with  a  request  that  the  committee  ensures  that  they  address  the 
matters of concern at the next meeting on 20th April 2020.   

3. Information circulated to GPs by Enfield CCG:   

i) 16th January 2017 - Scriptswitch message: 

“Please switch EpiPen solution for injection auto-injectors 300 micrograms/0.3ml (1in 1,000) to 
Emerade solution for injection auto-injectors 300 micrograms/0.3ml (1 in 1,000). The Emerade® 
adrenaline auto-injector currently has the longest shelf life of all the Adrenaline auto-injector pens 
currently available 2.5 years (30 months) from date of manufacture compared to others that are 
18 months from date of manufacture. If patients are switched to Emerade, it is advised that they 
are offered training in the use of the device. An on-line instruction video is available at the official 
Emerade® website along with a ‘request button’ for free training”. 

  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 8 and 9 contd 

ii) 2nd Sept 2017 – newsletter to GP practices distributed 

The  Enfield  CCG  Newsletter  emailed  to  all  GP  practices  included  a  link  to  Adrenaline  auto-
-  https://www.gov.uk/drug-safety-
injectors  updating  advice  after  a  European 
update/adrenaline-auto-injectors-updated-advice-after-european-review.    The  link  references 
advice  to  healthcare  professionals  to  “ensure  that  people  with  allergies  and  their  carers  have 
been trained to use the particular auto-injector that they have been prescribed—technique varies 
between injectors”. 

review 

iii) 28th September 2018 – supply distribution alert 

Supply Disruption Alert was emailed to all GP surgeries   
http://nww.enfield.nhs.uk/Docs/Referral%20Forms/EpiPen%20supply%20issue.pdf 
The  alert  stated  “The  different  brands  of  adrenaline  auto-injectors  are  not  used  in  exactly  the 
same way and therefore specific training and advice is required for each of the devices- please 
see information on these alternative devices below”. This alert helped to reinforce the view that 
training was required if a patient was changed from one device to another. 

iv)  17th  July  2019  –  policy  statement  approved  by  Enfield  CCG  Medicines  Management 
Committee 

Policy Statement agreed by Enfield CCG Medicines Management Committee on the prescribing 
of adrenaline auto injectors (Appendix 2) was put on Enfield GP intranet and discussed with GPs 
at all the GP locality events. 

v) 19th July 2019 – scriptswitch message 

A  Scriptswitch  message  on  300mcg  AAIs  was  issued  stating  “Consider  prescribing  Emerade 
500mcg  if  risk  of  severe  anaphylaxis  (for  self-administration  patients  at  risk  of  severe 
anaphylaxis).  

vi) 30th January 2020 – Medicines Safety Bulletin 

Following receipt of the narrative verdict Medicines Safety Bulletin (Appendix 1- as above) was 
circulated to Enfield GP practices and community pharmacists, and, was uploaded on the Enfield 
GP intranet site. The bulletin advises GPs to review the dosage and ensure it is appropriate for 
the patient’s age and weight, and to seek specialist advice for high-risk patients. 

 
 
 
 
  
 
 
 
 
 
 
 8 and 9 contd.  

vii) 4th March 2020 - scriptswitch message update was issued advising practices: 

  All strengths of Emerade devices will be unavailable for the foreseeable future. 
  Review patients to ensure Adrenaline auto-injector (AAI) appropriate according to current 

guidelines 

  An alternative brand of AAI must be prescribed (Epipen or Jext)   
  Dose should be checked to ensure it is appropriate for the patient especially in children 

as they age and increase in weight or for patients whose condition changes. 

  Patients  should  be  given  robust  training  about  carrying  two  AAI  pens  with  them  at  all 
times,  and  administering  the  second  dose  5-15  minutes  after  the  first  if  their  condition 
does not improve. (Document in patient notes that this advice has been given). 

  Patient (or representative  where appropriate) must  be counselled  on  change  of  device 

and trained on how to use the new device. 

  For children - ensure an allergy action plan is in place and to ensure patient/carer can 

follow: https://www.bsaci.org/about/download-paediatric-allergy-action-plans 

viii)  

The Prevention of Future Deaths Report is included on the agenda of Enfield CCG’s Medicines 
Management Committee for discussion with GPs and pharmacists on 31st March 2020.  

ix) Ongoing Actions 

Matters of Concern and opportunities for shared learning from incidents are raised during face to 
face  practice  meetings with GPs,  community  pharmacists,  Primary  Care  Network  pharmacists 
and other prescribers; GP Locality meetings and GP protected learning time events.  

Enfield CCG has requested the Prevention of Future Deaths Notice is discussed at the following 
meetings to ensure that this incident can be shared and reviewed further and learning identified 
and shared across North Central London: 

  Medicines Safety Officers and Medical Devices Safety Officers (MDSO) meeting on 4th 
April 2020.  This meeting is attended by Medicines Safety Officers from primary care and 
hospital  trusts  and  is  chaired  by  a  GP;  this  group  will  help  share  learning  across 
healthcare sectors. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 8 and 9 contd. 

  Enfield Medicines Management Committee 31st March 2020. 
  North Central London Joint Formulary Committee on 20th April 2020. 

CCG  pharmacists  to  check  what  action  they  have  taken  regarding  the  NCL  Medicines  Safety 
Bulletin on Adrenaline Auto Injectors (AAI) have contacted all GP practices. This is being followed 
up with individual practice meetings, training meetings for GPs, training sessions for Primary Care 
Network pharmacists, and by CCG pharmacists working in practices. A record is in the process 
of being collated to capture actions by individual practices to ensure all patients regularly receive 
appropriate training in the use of their device and dosages have been reviewed.  This process is 
due to complete by 30th April 2020.  

An  analysis  of  CCG  actions  that  may  have  been  contributory  to  this  incident  were  lack  of 
governance  around  scriptswitch  messaging  and  newsletters  sent  to  practices  from  the  CCG 
Medicines  Management  team.  Following  the  review  of  this  case,  it  was  identified  that  a  more 
robust governance  and decision  making process  was needed.   Current measures to  stop  any 
reoccurrence include: 

  checking and approval of scriptswitch messaging by senior pharmacists 
  newsletters to be checked and approved by GP Clinical Lead 

NCL primary care group has been set up to review the local formulary and scriptswitch messaging 
to standardize messages across NCL CCGs. 

The CCG will now implement a post incident review and a report will be completed to ensure that 
all actions identified are implemented to prevent a recurrence of this nature.  This will include a 
review of governance processes and decision-making points This report will go to the Quality and 
Safety Committee, which will oversee any recommendations and ensure  implementation of all 
actions.
Response from London Ambulance Service (PDF)
London Ambulance Service  ,~t:kj 

NHS  Trust 

Chief Executive Office 
Headquarters 
220 Waterloo Road 
London 
SE1  8S0 

Tel:  0207 783 2046 

www.londonambulance.nhs.uk 

Your Ref: 
Our Ref : 4034 
20  March 2020 

HMC Mary Hassell 
Inner North London 
St Pancras Coroners Court 
Carnley Street 
London 
N1C 4PP 

Dear Ms Hassell 

Regulation 28:  Prevention of Future  Deaths Report for Shante Andree Marie Turay-
Thomas 

Thank you for your Regulation 28 Prevention of Future Deaths Report ("PFD") dated 27 January 
2020.  I would  like to take this opportunity at the outset of my letter to  offer my condolences to 
Ms Turay-Thomas's family. 

Following  the  conclusion  of  the  above  inquest,  the  London  Ambulance  Service  ("LAS") 
understands that the learned Coroner has raised  a query as to whether the LAS would  be  able 
to assist in relation to the latter two sub-paragraphs (3 and 4) of paragraph 16 of the PFD report. 
Those paragraphs provide as follows: 

"I  heard at this inquest that NHS  Digital has since  changed its  categorisation [of calls relating 
to acute anaphylaxis]. However,  I also heard that for those areas (I think approximately half the 
country,  though  this is  not completely clear to  me)  where  the  999 service and the  111  seryice 
are  supported by different computer systems rather than  the  same  system being common  to 
both services,  there could remain inconsistencies of categorisation between 999 and 111. 

Even  where  there  are inconsistencies in  categorisation,  the  999 service  will not re-categorise 
following  a  111  clinician's  categorisation,  unless a  999 clinician  has spoken  to  the  patient,  so 
inappropriate  111  categorisation  will  not be  safety netted by the  999  service.  This  must be 
recognised and factored in". 

The learned Coroner is correct that,  as per the evidence of the LAS as given in this inquest,  in 
parts of the  United  Kingdom  different triage  systems are in  use  by the  999  and  111  services. 
NHS  111  only  use  NHS  Pathways  (owned  and  designed  by  NHS  Digital)  whereas  UK 
Ambulance Trusts  can  use either NHS  Pathways  or MPDS  (owned  and  designed  by  Priority 
Dispatch Corp.  ("PDC")). 

Had  the  LAS triaged  the call  made by - th rough  its 999 system,  as the LAS  stated 
in  their evidence,  it  would  have been~ the  MPDS  algorithms  in  use  by the  LAS. 

1 

 The LAS  use the  MPDS  algorithm and  supporting  software on  licence from  PDC.  We refer the 
learned Coroner to Appendix 2 attached to these for further details of the MPDS algorithm itself 
and its operation. 

The  Response  Priority  for  each  MPDS  determinant  descriptor  - being  the  Category  of 
ambulance it requires - is set by ECPAG. This is because final decisions about categorisation 
are made  by ECPAG on  behalf of NHS  England.  It is,  therefore, this organisation (if any) who 
has the power to take the action the learned Coroner is seeking at sub-paragraph 3 above and, 
where possible, would  be capable of identifying any inconsistencies between the categories of 
ambulance assigned  to  dispositions within  the  NHS  Digital  system  on  the  one  hand  and  PDC 
on the other,  in  relation to acute anaphylaxis and  otherwise. 

The LAS, as an MPDS user for its 999 services and an NHS Pathways user for its 111  services, 
attend  the  relevant  user  groups  with  other  ambulance  trusts  and  fully  participate  in  such 
meetings, where there  is  the  opportunity to  share  data  back  and  forth,  discuss cases of note 
and share learning in order to make continual improvements through the feedback mechanisms 
to  NHS  Digital  and  the  Academy  at  PDC  process.  The  LAS  will  discuss the  findings  of PFD 
report  at  such  user groups.  Ultimately,  however,  changes to  how the  triage  system  operates 
are  a  matter for  the  International  Academies  of  Emergency  Dispatch  (IAED)  where  MPDS 
(owned by PDC) is concerned, or for NHS Digital under their national clinical governance group 
where NHS Pathways is concerned. 

From  the  LAS's  understanding  however,  there  are  limitations to  ECPAG's  ability to  align  the 
two systems - NHS  Pathways and  MPDS - as they operate differently and  have two different 
methodologies for reaching  a triage decision. 

In order to assist the learned Coroner, as indicated, the LAS attaches as Appendix 2- a detailed 
summary from the LAS formally setting out how it is UK Ambulance Services manage 999 calls 
and the national system for prioritisation of those calls. 

I hope this reply is  helpful in  explaining the actions undertaken by the LAS. 

Yours sincerely
Response from NHS Digital Redacted 1 (PDF)
1 Trevelyan Square 
Boar Lane 
Leeds LS1 6AE 

0300 303 5678 
0113 397 3614 

16th March 2020 

M E Hassel 
Senior Coroner 
Inner North London  
St Pancras Coroners Court 
Camely Street 
London 
N1C 4PP 

By Email: 

 Our Reference LT02120 

Dear Ms. Hassel  

Inquest into the death of Shanté Andreé Marie Turay-Thomas 

I am writing in response to the Prevention of Future Deaths (“PFD”) Report received from HM Coroner 
dated 21st January 2020.  This follows the death of Shanté Andreé Marie Turay–Thomas who sadly died 
on 15th September 2018. This was followed by an investigation and inquest which concluded on  13th 
January 2020.  Firstly, we would like to express our sincere condolences to the family of Ms Turay-Thomas. 

NHS Pathways is the clinical decision support software (CDSS) used by all 111 service providers, and 
some 999 ambulance trusts.  For information, we have included a short summary of the functions that 
NHS  Pathways  performs and  the governance  that underpins  it (containing background information 
on NHS Pathways) in Appendix A. 

HM Coroner  raised matters of concern numbered 1 - 20 in the PFD report.  Matters of concern 1 – 
14 and 20 are not applicable to NHS Pathways.   We set out below our response to matters of 
concern 15 to 19. 

www.digital.nhs.uk 
enquiries@nhsdigital.nhs.uk 

 
 
 
 
 
 
 
 
 
 
 
 
  
   
 
 
 
 
 1)  Matter of Concern 15 

During the course of the 111 call, a number of errors were made. These were the errors of LCW 
individuals. When LCW audited the call in the first instance, the audit identified the problem with 
the address, but failed to recognise how badly the call had gone in other ways. Without effective 
audit and recognition of failings, it is difficult to see how there can be effective improvement.  

NHS Digital has developed an extensive audit framework that applies to providers using NHS 
Pathways.  This was described in NHS Digital’s: supplementary second witness statement dated 
20th December 2019; submissions on conclusion dated 3 January 2020; and PFD submissions 
dated 17 January 2020.  

Clinical  providers  of  services  using  NHS  Pathways  (‘providers’)  must  enter  into  a  Licence 
Agreement  with  the  Secretary  of  State  for  Health  and  Social  Care,  including  requirements 
relating  to  implementation,  operation,  training  and  auditing.    This  licence  and  the  supporting 
materials are managed by NHS Digital. 

The  detailed  requirements  for  auditing  call  handlers  (including  clinicians)  establish  audit 
frequencies, and standards against which audits should occur.   These include criteria for auditing 
and requirements by way of qualification and training of those carrying out any audit.  NHS Digital 
also offers regular ‘audit levelling’ sessions to providers in order to achieve good and consistent 
practice.  

Ensuring the proper operation of the audit system and management of failed call audits  is the 
responsibility of the provider, within the framework provided by NHS Digital.  NHS Digital remains 
available to support providers in this area. 

2)  Matter of Concern 16 

The  individuals  making  these  errors  were  working  within  the  context  of  NHS  Digital’s 
categorisation  of  anaphylaxis  as  needing  a  category  2  ambulance  rather  than  a  category  1 
ambulance, on the Adastra computer system that supports the LCW 111 service.  

This was the wrong categorisation and not the categorisation that the call would have received 
if 999 had been called and the London Ambulance Service contacted in the first instance. Acute 
anaphylaxis is immediately life threatening and must be treated as a category 1.  

I heard at inquest that NHS Digital has since changed its categorisation. However, I also heard 
that for those areas (I think approximately half the country, though this is not completely clear to 
me), where the 999 service and the 111 service are supported by different computer systems 
rather than the same system being common to both services, there could remain inconsistencies 
of categorisation between 999 and 111.  

Even  where  there are  inconsistencies  in  categorisation,  the 999  service  will  not  re-categorise 
following a 111 clinician’s categorisation, unless a 999 clinician has spoken to the patient, so 
inappropriate  111  categorisation  will  not  be  safety  netted  by  the  999  service.  This  must  be 
recognised and factored in.  

 
 
 
 
 
 
 
 
 
 
 
 
 Ambulance  response  categorisation,  and  the  alignment  of  different  triage  systems,  is  not  the 
responsibility of NHS Digital. This was set out in the following submissions made on behalf of 
NHS Digital: 

• 
letter to HM Coroner dated 22 July 2019 
• 
letter to HM Coroner dated 9 October 2019 
•  witness statement dated 14 November 2019  
•  evidence given at inquest on 12 December 2019 
•  submissions on conclusion dated 3 January 2020 
•  PFD submissions dated 17 January 2020. 

NHS  Pathways  and  the  Medical  Priority  Dispatch  System  (MPDS)  –  (the  system  used  by 
ambulance trusts not using NHS Pathways) operate by prioritising patients based on symptoms. 
Although they work differently, neither is designed to make a diagnosis. NHS Digital does not 
have  oversight  or  detailed  knowledge  of  MPDS;  it  is  a  competing  system  provided  by  a 
commercial supplier. 

The categorisation of certain symptoms, which may occur in anaphylaxis, as requiring a category 
2 emergency ambulance within NHS Pathways was ratified in 2017 during NHS England’s review 
of  ambulance  standards  called  the  “Ambulance  Response  Program”.    This  program  was 
implemented  from  2nd  October  2017,  and  was  approved  by  the  Secretary  of  State  following 
extensive piloting.   

NHS Pathways’ ambulance response codes are also ratified by the National Ambulance Services 
Medical Directors (NASMED) (an advisory group consisting of medical director representatives 
from  all  ambulance  services  in  England,  Wales,  Scotland  and  Northern  Ireland)  and  the 
Emergency  Call  Prioritisation  Advisory  Group  (ECPAG)  (a  further  group  of  advisors  to  NHS 
England). 

Following  the  Ambulance  Response  Program,  NHS  England  led  (supported  by  NASMED  and 
ECPAG) a “clinical coding review” in May 2019, reviewing the category 1 ambulance response 
definition.    Consequently,  it  was  decided  that  symptoms  which  may  suggest  life-threatening 
anaphylaxis should receive a category 1 ambulance response and the necessary changes were 
made by NHS Pathways. These were beta tested in September 2019 and deployed nationally 
from October 2019. 

The  Ambulance  Response  Program  and  clinical  coding  review  applied  equally  to  MPDS, 
whose ambulance dispositions were similarly reviewed and ratified.    

Where an ambulance disposition is triggered by a 111 call, the NHS Pathways system is able 
to trigger ambulance dispatch (subject to local validation procedures which may be applied to 
category 3 and 4 ambulances).  It would not be clinically safe to require re-triage or validation 
for category 1 and 2 ambulances due to the additional time this would take.   As described in 
NHS Digital’s submissions the 111 and 999 questions and responses are identical where NHS 
Pathways is used (other than an additional first ‘nature of call’ question for 999).  MPDS is an 
entirely  different  product  with  different  operating  rationale  but  with  ambulance  codes  still 

 
 
 
 
 
 
 
 
 
 
 
 ratified in the same manner.  The 999 service is not designed or intended to be a ‘safety net’ 
for 111.  

NHS Digital contributes to ECPAG and will continue to support NHS England as far as it is 
able  with  the  complex  challenge  of  aligning  the  two  very  different  systems.    There  is  no 
evidence to suggest that prior to this incident, NHS Digital, through LCW or any other user of 
NHS Pathways, had been made aware of any issue or concern in respect of the ambulance 
response category for symptoms that may occur in anaphylaxis.  If NHS Digital became aware 
of such an issue then its response would include raising this with NHS England, NASMED and 
ECPAG. 

NHS England is the organisation charged with overseeing both  NHS Pathways and MPDS, 
and  has  the  remit  and  ability  to  review  potential  inconsistencies  or  change  ambulance 
categorisation. Accordingly, concerns regarding ambulance categorisation or inconsistencies 
between MPDS and NHS Pathways can only be properly answered by NHS England.  

3)  Matter of Concern 17 

In  terms  of  national  training  for  111  call  handlers,  the  NHS  digital  distance  learning  pack 
contains  advice  that  is  in  part  inadequate  and  in  part  wrong.    It  does  not  give  the  crucial 
information that one dose of adrenaline, whichever device it is administered, is very unlikely 
to be sufficient in the case of acute anaphylaxis.  It contains a photograph to illustrate he use 
of an AAI, but in the photograph the device is held incorrectly.  

As NHS Digital described in evidence and subsequent submissions, the distance learning pack 
is provided as a foundation of background information to all call handlers before they start training.  
It does not direct the progress of specific calls nor is it relied upon to communicate advice which 
a  call  handler  should  give.    It  would  not  be  clinically  safe  or  appropriate  for  non-clinical  call 
handlers to be required to exercise knowledge or judgment, or act unprompted by the system, in 
this way.  Call handlers are instead supported by the questions and care advice presented by the 
NHS  Pathways  system.  The  system  contains  ‘supporting  information’  to  help  call  handlers 
understand the clinical essence of what is being asked or advised. This is presented in ‘real-time’ so 
that  the  call  handler  has  the  required  information  in  front  of  them,  rather  than  having  to  rely  on 
memory. 

NHS Digital welcomed the evidence given by the expert witness, Professor Fox, at the inquest 
and  immediately  recognised  that  the  distance  learning  pack  could  be  improved,  assuring  the 
Coroner (in evidence and in the supplementary second witness statement dated 20th December 
2019) that a review would be undertaken, in consultation with Professor Fox, to address the points 
raised during his evidence.     

NHS Digital informed the Coroner in its PFD submissions, dated 17 January 2020, that the review 
had been completed and an amended version of the Distance Learning Pack had been produced 
and  released  to  all  organisations  that  use  NHS  Pathways  to  support  telephone  triage.  The 
amendments  were made  in  liaison  with  the expert witness.  In  summary the  following  changes 
have been made: 

a)  Anaphylaxis  is  now  described  as  “a  potentially  life-threatening  sudden,  very  severe 

 
 
 
 
 
 
 
 
 
 
 
 
 allergic reaction requiring urgent intervention”.  

b)  A  further  section  has  been  added  in  respect  of  AAIs  which  states  “A  person  with 
anaphylaxis needs emergency treatment with an injection of intramuscular adrenaline and 
people  who  have  previously  had  a  serious  allergic  reaction  often  have  an  adrenaline 
autoinjector (AAI) for use in case of future reactions. The AAI should be used as soon as 
signs of anaphylaxis appear. Further doses are needed at 5-minute intervals if there is no 
clear sign of recovery.” 

c)  Pictures of the Emerade, Jext, and Epipen AAI have been included. 

Information about the specific dose required in respect of each AAI has not been included. This 
is because the appropriate dose is a matter for the prescriber, not the call handler issuing system-
generated instructions on how to administer the medication. 

4)  Matter of concern 18 

I  am  unclear  as  to  whether  the  Adastra  111  algorithm  automatically  prompts  administering  a 
second AAI Five minutes after the first if there has been no improvement, but it should. 

NHS Pathways has always prompted call handlers to give instructions in respect of a second 
administration of AAI if there is no improvement, as NHS Digital stated in the PFD submissions, 
dated 17 January 2020.   This is NHS Pathways content and is not affected by the system into 
which NHS Pathways is embedded (e.g. Adastra in this case). 

Currently, NHS Pathways content suggests that a second AAI is administered after 10 to 15 
minutes if there has been no improvement in the patient’s condition.  NHS Digital has reviewed 
this  timeframe  and  it  is  being  amended  to  5  minutes  in  line  with  guidelines  from  the 
Resuscitation Council.   These changes will be made  in NHS Pathways Release 20, which 
was originally scheduled for deployment in May 2020, but has subsequently been delayed due 
to coronavirus.  

5)  Matter of concern 19 

One of the errors made by the first 111 call handler was a failure to ask to speak direct to the 
patient. This was the error of an individual.  

However,  this  is  not  the  first  time  that  the  issue  has  been  brought  to  the  attention  of  NHS 
Digital. At inquest, I asked the witness who appeared on behalf of NHS Digital, and indeed 
had been chosen by NHS Digital as the person best able to assist the court, if this had been 
an  issue  in  the  past.  He  said  no.  However,  on  18  December  2018,  Peter  Harrowing,  HM 
Assistant Coroner for Avon, sent a prevention of future deaths report to NHS Digital following 
the inquest touching the death of David Longden.  

It was only when I asked the witness appearing on behalf of NHS Digital specifically about 
Coroner Harrowing’s report in respect of Mr Longden, pointing out that Coroner Harrowing had 
raised the need for NHS Digital to place greater emphasis on the call handler speaking to the 
patient, that the witness remembered that he had indeed seen that report.  

 
 
 
 
 
 
 
 
 
 
 I choose to characterise this as a memory lapse rather than as an intention wilfully to mislead 
the  court.  (A  witness  who  lies  whilst  giving  evidence  on  oath  at  inquest  may  be  found  in 
contempt of court and may even be prosecuted for the crime of perjury.) Nevertheless, if NHS 
Digital does not  have a  grasp  of  this sort of detail,  specifically brought to  its  attention  by  a 
coroner in a prevention of future deaths report, it is difficult to see how there can be effective 
improvement. 

For accuracy the Prevention of Future Death Report that the Coroner refers to was that for 
Mrs Susan Longden and not Mr David Longden.     

Speaking to the Patient 

In NHS Digital’s supplementary second witness statement, dated 20th December 2019, the 
following was addressed: 

•  The importance of call handlers speaking to the patient where safe and appropriate is a 
fundamental principle and core competency for NHS Pathways use.  This is emphasised 
through initial training, call critiques, supervised practice, use of toolkits and completion of 
written and practical assessments.  

•  Speaking with the patient is also a competency indicator assessed during auditing of new 

and experienced call handlers.  

•  NHS Digital was not made aware of the inquest touching the death of Ms Susan Longden 
in advance and therefore was unable to assist Dr Harrowing in his inquiry.  NHS Digital’s 
response to Dr Harrowing explained (as in the above bullets) the emphasis placed on the 
importance of speaking to a patient. 

•  NHS Digital consider that the NHS Pathways training materials and licence requirements 
sufficiently address the need and importance of call handlers speaking directly to patients 
and  recognise  that  111  and  999  providers  should  continue  to  enforce  this  with  call 
handlers. 

• 

In  version  19.3.0  of  NHS  Pathways,  which  was  released  to  111  and  999  on  the  13th 
January 2020, the following changes were made to the NHS Pathways content to further 
enhance and support call handlers in attempting to speak to the patient: 

b) 

a)  “Inclusion of a new question for calls received from 3rd party callers asking, “Is it possible for 
me  to  speak  to  them?”,  which  will  follow  establishing  that  the  patient  is  conscious  and 
breathing.  This will force call handlers to ask this; and  
in questions asking, “[is the patient] so breathless that speaking more than a few words is 
impossible?” addition of a statement on the supporting information to remind call handlers to 
again try to speak with the patient, even if the 3rd party caller has refused in response to the 
question above. The supporting information now states (new language in capitals) “To find 
out if there are features of life-threatening breathing difficulty. THIS WILL BE EASIER TO 
ASSESS BY SPEAKING TO THE PATIENT.” 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
  Appendix A 

Function of NHS  Pathways 

BACKGROUND INFORMATION 

NHS  Pathways is a programme providing  the Clinical  Decision Support System  (CDSS)  used 
in NHS 111  and half of English ambulance services. This triage  system supports the remote 
assessment of over  16.7 million calls per annum. These calls are managed by non-clinical 
specially  trained  call  handlers  who  refer  the  patient  into  suitable  services  based  on  the 
patient’s  health  needs  at  the  time  of  the  call.  These  call  handlers  are  supported  by 
clinicians who are able to provide advice and guidance or who can take over the call 
if the situation requires it. The system is built around a clinical hierarchy,  meaning that life-
threatening  problems  assessed  at  the  start  of  the  call  trigger  ambulance  responses, 
progressing  through  to  less  urgent  problems  which  require  a  less  urgent  response  (or 
“ disposition”)  in other settings. 

Governance  of NHS Pathways 

The  safety of  the  clinical  triage  process endpoints  resulting  from a 111 or 999  assessment 
using  NHS  Pathways,  is overseen  by the National  Clinical  Governance  Group, hosted by the 
Royal  College  of General  Practitioners.  This  group  is  made  up  of  representatives  from  the 
relevant  Medical  Royal Colleges.  Senior  clinicians  from  the  Colleges  provide  independent 
oversight  and scrutiny of the  NHS Pathways clinical content. Changes to the NHS Pathways 
clinical content cannot be made unless there is a majority agreement at NGCC. 

Alongside  this  independent  oversight,  NHS  Pathways  ensures  its  clinical  content  and 
assessment protocols  are  concordant  with  the  latest  advice  from  respected  bodies  that 
provide  evidence  and  guidance  for  medical  practice  in  the  UK.    In  particular,  we  are 
concordant with the latest guidelines  from: 

• 
• 
• 

NICE  (National  Institute  for Health  and Clinical  Excellence) 
The UK Resuscitation Council 
The UK Sepsis Trust
Response from NHS England (PDF)
Coroner ME Hassell 

Senior Coroner 
Inner North London 
St Pancras Coroners Court 
Camley Street 
London  
NC1 4PP  

Dear Ms Hassell,  

National Medical Director  
NHS England & NHS Improvement  
Skipton House 
80 London Road 
London 
SE1 6LH 

27th May 2020 

Re: Regulation 28 Report to Prevent Future Deaths – Shante Turay-Thomas 
who died on 15 September 2018.  

Thank you for your Regulation 28 Report (hereafter the ‘report’) dated 29/01/2020 
concerning the tragic death of Shante Turay-Thomas on 15 September 2018. Firstly, 
I would like to express my deepest condolences to Shante’s family and all those 
touched by her death. 

Your report concludes that Shante’s death was a result of acute anaphylaxis 
following an allergic reaction to nuts she had consumed. Your report covers many 
complex issues and expresses a number of concerns directed towards several 
different organisations. In preparing our response to your report, NHS England and 
NHS Improvement (hereafter ‘NHSEI’) has worked closely with the Department of 
Health, Enfield Clinical Commissioning Group (CCG), NHS Digital and the Medicines 
and Healthcare products Regulatory Agency (MHRA) to inform their responses and 
ours. I am aware that these organisations will address the points most relevant to 
them, therefore we have focused on responding to the concerns you have directed 
toward NHSEI and that are relevant to NHSEI’s statutory duties.  

Following the inquest, you asked NHSEI to address your concern regarding 
guidance given to CCGs and regarding considerations of incentives (CQUINS 
indicators etc) to educate and train primary care staff on the management of 
allergies.  

NHSEI has an assurance role for CCGs and their commissioning of healthcare 
services. NHSEI does not have any direct role in offsetting the curriculum content for 
health professionals’ education and training. However, I can set out the following 
which sets out NHSEI’s work and commitment in this area: 

•  We impose a contractual obligation on healthcare providers to employ staff who 
are appropriately trained and to require staff to stay up to date with lifelong 
learning. NHSEI works closely with Health Education England (HEE) and the 
professional Royal Colleges, who would be best placed to consider whether 

NHS England and NHS Improvement 

 
 
 
 
 
 
 
 
 
 
 
 
 
 • 

under-graduate and professional training in relation to allergy management 
could be strengthened.  
I will ensure your report is sent to HEE and the Royal College of General 
Practitioners. Our commissioning teams will liaise directly with all relevant 
organisations to facilitate uptake of any new guidance and resource that would 
support better management of people with severe allergies.  

•  The use of incentives such as CQUIN have been set nationally for 2020/21 

contracts. CQUIN as a route to provide incentives will be under review in future 
years, and decisions will be made in due course for 2021/22.   

•  The British Society for Allergy & Clinical Immunology (BSACI) provide training 

for primary care staff across the country (workshops and educational 
meetings). Additional training of primary care staff / establishing training 
requirements would be via their professional organisations (e.g. Royal College 
of GPs), Health Education England and CCGs. 

•  Allergists and Immunologists are Fellows/Members of the Royal College of 
Physicians. Allergy and Immunology training is the responsibility of the RCP 
through the JRCPTB (Joint Royal Colleges of Physicians Training Board). A 
CCT (Certificate of Completion of Training) in allergy and immunology is 
available to support specialist allergy care. The Royal College of Physicians 
also runs an accreditation scheme – Improving Quality in Allergy Services. 
BSACI also provide training for a range of professionals.  

You also asked NHSEI to address your concern around there being no person with 
named accountability for allergy services and allergy provision. 

Allergy is a clinical speciality concerned with the identification and management of 
patients with a broad range of allergic conditions. These include asthma, eczema 
and acute food allergy. The patient group is broad, including paediatric and adult 
patients, and the acuity with which allergic reactions manifest is extremely varied. 

NHSEI has overall responsibility for the quality of allergy services across England.  
Most allergy services are commissioned locally through CCGs; however, to support 
patients with rare and complex conditions NHSEI directly commissions some 
specialised services. These include specialised allergy services through the Blood 
and Infection National Programme of Care. 

We have an established Clinical Reference Group (CRG) that provides clinical 
advice and leadership on the specialised immunology and allergy services. The 
relevant CRG is chaired by a consultant immunologist. Membership of the CRG is 
comprised of clinicians, commissioners, public health experts, patients and carers.   
They use their specific knowledge and expertise to advise NHS England on the best 
ways that specialised services should be provided. 

CRGs lead on the development of clinical commissioning policies, service 
specifications and quality standards. They also provide advice on innovation, horizon 
scanning, service reviews and guide work to reduce variation and deliver increased 
value. CRGs, through their Patient and Public Voice (PPV) members, also help 
ensure that any changes to the commissioning of specialised services involve 
patients and the public. 

 
 NHSEI also has a number of National Clinical Directors and National Specialty 
Advisers who are practicing clinicians from across England. They provide clinical 
leadership, advice, input and support across distinct areas of NHS conditions and 
services.  

We are deeply saddened by Shante’s death. We are grateful to have had the 
opportunity to respond to your concerns relevant to NHSEI. We will continue to work 
the with HEE, the professional Royal Colleges and the other organisations 
addressed in your report to keep abreast of any new guidance or resources that they 
produce that would support better management of people with severe allergies. We 
will consider whether any of our communication routes or commissioning levers can 
help with their uptake and embedding.  

Thank you for bringing this important patient safety issue to my attention and please 
do not hesitate to contact me should you need any further information. 

Yours sincerely, 

Professor Stephen Powis 
National Medical Director
Response from Nice (PDF)
N I CE Nationa l Inst itute for 

Hea lth and Care Excellence 

ij  'IV'/~ 

1 

0  MAR  2020 

10  Sprin g Gardens 

Lo ndon 
SW1A 2BU
United  Kin gdom 

+44 (0)300 323 0140 

02 March 2020 

M.E.  Hassell 
Senior Coroner 
Inner North London 
St Pancras Coroner's Court 
Carnley S treet 
London N 1C 4PP 

Our ref:  EH-304794 

Dear Ms  Hassell , 

I write  in  response  to  your correspondence,  dated  28  January 2020,  regarding  the  death  of 
Shante Andree Marie TURAY-THOMAS.  I was very sorry to read of Ms Turay-Thomas' death. 

We  have  considered  the  circumstances  surrounding  Ms  Turay-Thomas·  death,  and  the 
concerns  raised  in  your  report.  In  particular,  the  concerns  that  there  is  an  absence  of NICE 
guidance concerning : 

• 
• 
• 

the dose of adrenaline  in  different adrenaline  auto-injectors pens , 
the  number of pens that should be  carried,  and 
training  in  how to  use the devices 

We  note that you say in  point  10 of your report that  'the whole area would benefit from  NICE 
review'. 

Having  reviewed  your concerns , we consider that the  British National  Formulary (BNF) and 
the  BNF for Children (BNFc) already conta in detailed advice on  these aspects of care, 
including  the fo llowing  pieces  of MHRA/CHM advice from  2017 and 2019,  in  the section on 
Adreneline/Epinephrine (https://bnf.nice.org .uk/drug/adre nalineepinephrine.html) : 

'Adrenaline auto-injectors:  updated advice after European review (August 2017) 

With  intramuscular use 

Folfowing a European review of all adrenaline auto-injectors approved in the EU, the MHRA 
recommend that 2 adrenaline auto-injectors are prescribed, which patients should carry at all 
times. This is particularly important for patients with  allergic asthma, who are at increased 
risk of a severe anaphylactic re action. Patients with  allergies and their carers should be 
trained to  use the particular auto-injector they have been prescribed and encouraged to 
practise using a trainer device.  Patients are  advised to  check the expiry date  of the 
adrenaline auto-injectors and obtain  replacements before they expire. 

NICE 

www.nice.org.uk  I  nice@ nice.org.uk 

 'Adtenallne a,,,to./n/ecfD, w t action taken to supportsafety(Od.ober 2019} 

With lntratnuscular use 

Hea/tht:are professions/a.,. reminded to follow exiatlng advice on the uae ofadtenllllne 
auto-lnjectortl- also PrNorlblng and dispensing Information.  Patients should be 
enooursged to reed the MHRA advice sheet and to sign up for the Expiry alert NfVlce d 
their auto-ln}e«or device an the manuladurer's website. 
Healthcare professkJnala ahould IIIBo be awan, d recent alerta and letters Issued about 
adrenaline auto-injeal.ors Including Iha activation failure iaue wfttl Emerade9 aul.o-/njector 
pens and the fol.r-month extension ofexpiry dates ofCfJl'laln batchN ofEpipen8and J..-. 
The MHRA has produced e letter on the EmetadeCl>actlvatlon Issue, which ahoUld be 
provided to patients and their caws.. 

The BNF and BNFc are both joint publcalions of the British Medlcal Aasoclation and the 
Royal Pharmaceullcal Society, and both fomlularfea are accesslble from the NICE website. 
Prescriber& ant expected to refer to lnfonnation within the BNF and the BNFc to help inform 
prescribing decisions made with Individual patlenta and carers. This expectation Is aet out in 
the Gena-al Meclcal Council's publlcaUon on 'Qgqd practice In presatinq and managing 
rnedJclnes and deyices'. within the section lllled: Keeping up to date and prescribing safely. 

We have publshed a clinical guideline on food allergy In under 19s: asessment and 
diagnosis (CG118), which we undendand would haw been relevant to Ms Turay-Thomas' 
assessment and diaanosis. However, it does not cover the management of anaphylactlc 
reactions and we therefore consider it not drectly relevant to the concams regarding t.r 
death. 

We have alao published a clnical guideline on anaphylaxie: assessment and referral after 
emergency treatment (00134) and we have a quality standard on anaphylaxla (0S119). 
Both this guldellne and qually standard cover care alter emergency treatment for suspected 
anaphylaxls. Including assessment and referral to speclalst allergy services. That Is, they 
begin at the point In the clinical pathway immacllataly aftera health professional has started 
to manage a suspected anaphylactic raactlon. 

It Is not clear from Y0II" report whether Ma Turay-Thomas had ever e,cperlenced an 
anaphylactlc reaction before the one that caused her death, and it appears that she had died 
by the time any health professfonal attended her during the ratal episode. nis therefore 
undear whether either 00134 or QS119 would have been directly relevant to the Issues that 
contributed to her death. 

We do not conskler it appropriate for us to dupllcate the BNF advice raferrad to above. 
However, we will consider how best to make clear In CG134 the advice that 2 adrenaline 
auto-injectors should be prescribed, which patients should carry at 811 tfflls.
Response from Radcliffes Le Brasseur (PDF)
By Email 

85 Fleet Street  
London  
EC4Y 1AE 
www.rlb-law.com 

Tel +44 (0)20 7222 7040 
Fax +44 (0)20 7222 6208 
LDE 113 London Chancery Lane 
info@rlb-law.com 



16 March 2020 

Our Ref:  JCS1/FMB/900100.13112 

Your Ref:   

22904082v3 

Mrs Mary Hassell 
Senior Coroner 
Inner London North 

Sent by Email Only 

Dear Madam 

Inquest Touching the death of Ms Shante Turay-Thomas 
Response to Regulation 28 report 

We continue to be instructed to act on behalf of the three G.Ps that were interested persons in the above 
matter. 

We are writing in response to the Learned Coroner’s Prevention of Future Deaths Report dated 27 January 
2020, issued pursuant to Regulation 28 of the Coroner’s (Investigations) Regulations 2013. 

We note the matters of concern that the Coroner has identified.  As the Coroner is aware, we set out the 
action that has been taken by the Winchmore Hill Practice (“the Practice”) and that proposed, within the 
witness  statements  of  the  GPs  (particularly  the  supplementary  statement  of  Dr  Takla)  and  the  PFD 
submissions (enclosed.) We summarise these below and also set out the additional steps that have and will 
be taken in response to this matter as follows: 

1.  Following  Ms  Turay-Thomas’  death,  the  Practice  undertook  an  audit  in  October  2019  of  all 
patients who were being prescribed Emerade to ensure that the dosage was in accordance with 
the BNF based on the respective patients’ weight and age. Indeed, all patients who use AAI pens 
have had their doses reviewed. Patients have been contacted to ensure that they have the correct 
dose  and  appropriate  knowledge  about  the  use  and  storage  of  the  pen.  Letters  were  sent  to 
patients on Emerade, which included up to date advice from the MHRA in July 2019, December 
2019 and March 2020. In respect of patients taking Jext and Epipen, letters were sent to them in 
January 2020. 

2.  All  prescriptions  for  AAIs  are  now  dealt  with  as  acute  prescriptions,  as  opposed  to  repeat 
prescriptions.  This ensures that each prescription is scrutinised in detail to ensure that the type 
of pen and dose of adrenalin is appropriate for the patient.  On the face of any AAI prescription, 
it is expressly stated for the avoidance of any doubt that a patient should carry two AAI pens on 
their person at all times and ensure that they are familiar with the use of the pen. There is also 
some safety netting advice in the event of an emergency. We have enclosed a sample prescription 
to illustrate this change. In addition, each prescription is accompanied by an AAI brand specific 
letter to the patient providing the most important details about the AAI pens to enable a patient 
to use it safely and effectively. We have also enclosed a copy of a standard letter in this regard. 

Offices in London, Leeds and Cardiff 
Authorised and regulated by the Solicitor's Regulation Authority 
Firm SRA Number 184739 
A list of partners may be inspected at the above address 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Mrs Mary Hassell 

16 March 2020 

Page 2 



3.  The  Practice  has  taken  significant  steps  to  ensure  that  it  is  up  to  date  with  anaphylaxis 
management  and  the  use  of  AAI  pens.  It  arranged  in-house  training  for  all  clinical  staff  on  9 
December  2019,  which  was  delivered  by  a  Nurse, 
  This  training  included 
demonstrations as to how to use the three most common AAI pens, namely Emerade, JEXT and 
EpiPen, advice as to the different dosages available, instructions and demonstrations as to the 
different methods of administration for each brand. This was all based on the respective brands’ 
advice/instruction on their own explanatory posters and the demonstrations were done by Nurse 
  She  has  since  confirmed  that  she  has  also  seen  a  few  patients  who  have  attended  the 
Practice to obtain some training on the use of their AAI.  Following this training, the Practice has 
ensured that there are sample pens for all three brands available to the Practice nurses in order 
so that they are able to demonstrate to any patient who has been prescribed an AAI pen for the 
first time or those that are unsure as to how to use the pens the correct method of administration. 
The Practice actively encourages all patients who are prescribed AAIs to arrange an appointment 
with the practice nurse or pharmacist in order to obtain proper training in the use of the pen. 

4. 

In addition to the in-house training, mandatory basic life support training was held at the Practice 
on 27 February 2020.  This was delivered by an external provider, Mr
and Associates Healthcare Training and Education.   The basic life support training always includes 
a segment on anaphylaxis management.  However, the Practice contacted the training provider 
prior  to  the  course  in  order  to  emphasise  that  a  more  extensive  section  was  required  on 
anaphylaxis management.  We enclose a copy of the e-mail in this regard.  The basic life support 
training lasted two and a half hours in duration.  It included a 40 minute section on anaphylaxis 
management which encompassed recognising symptoms, the imperative of carrying two pens at 
all times and the different administration of the three types of pens available. 

 of 

5.  The Practice has received a number of alerts from the MHRA explaining defects pertaining to the 
Emerade  AAIs.  All  relevant  information  has  been  disseminated  to  patients  who  have  been 
prescribed Emerade. The Practice recently  received a further alert from the MHRA confirming 
that  all  Emerade  150mcg  AAIs  are  to  be  recalled  as  a  result  of  the  potential  for  the  pen  to 
malfunction. The Practice has ensured that all patients who were prescribed Emerade 150mcg 
have been switched to an alternative brand. Patients were sent letters with an alternative script 
on 9 March 2020. 

6. 

It was recognised by the Practice that some patients will not be inclined to attend the Practice in 
order to obtain advice and training in the use of an AAI pen.  To mitigate this, links to training 
videos and the relevant websites have been included within the letters to patients to encourage 
them to check that they are familiar with AAI self-administration and advice.  Furthermore, all 
clinical  staff  at  the  Practice  have  been  instructed  to  ensure  that  any  patient  who  attends 
(regardless of the purpose of their attendance) and who is being prescribed adrenalin has their 
AAI prescription as well as their understanding in relation to the use of the pen reviewed. This 
includes checking whether the patient is receiving appropriate care and support from secondary 
care. 

7. 

In addition, whenever a patient who has an allergy turns eighteen, the Practice will ensure that 
as part of any medication review, a check is undertaken as to whether the patient is appropriately 
transitioned into the care of adult allergy specialists if this is deemed necessary.  This will guard 

 
 
 Mrs Mary Hassell 

16 March 2020 

Page 3 



against  the  possibility  that  a  patient  is  left  without  specialist  allergy  care,  which  is  crucial  to 
safeguarding their health and well-being. 

8.  The Practice recruited an in-house pharmacist who commenced work in December 2020.   This 
appointment will prove conducive in ensuring that both patients and clinicians have up to date 
information in relation to AAI awareness.  The pharmacist is also on hand to provide training to 
any  patients  in  relation  to  the  correct  use  of  an  AAI.  There  is  accordingly  now  two  layers  of 
protection  in  that  guidance  is  offered  to  a  patient  when  prescribing  the  AAI  as  well  as  when 
dispensing it.  

9.  The Practice has undertaken a rigorous review of all patients that have been prescribed AAI pens 
so as to ensure that the correct dose and pen is being prescribed.  The Practice has nominated 
one  of  the  Pharmacists  to  act  as  the  ‘Practice  Anaphylaxis  Champion.’  This  role  will  include 
ensuring staff awareness, training and regular surveillance of appropriate prescribing practices 
and adherence to practice protocols. The Pharmacist will also contact all patients prescribed AAIs 
to undertake regular reviews of their condition, treatment and training. 

10. The Practice has cascaded the learning from this matter to other practices within the Primary 
Care  Network.  The  AAI  policy  together  with  patient  information  leaflets  and  messages  on 
prescriptions have been shared. 

11. The Practice has shared learning with the CCG medicine management team and the message on 

scriptswitch has been amended as follows: 

  All strengths of Emerade devices will be unavailable for the foreseeable future.  

  Review patient to ensure Adrenaline auto-injector (AAI) appropriate according to current 

guidelines  

  Alternative brand of AAI must be prescribed (Epipen or Jext)  

  Dose should be checked to ensure it is appropriate for the patient especially in children 

as they age and increase in weight, or for patients whose condition changes.  

  Patients  should be given robust  training about  carrying two AAI  pens with them at all 
times, and administering the second dose 5-15 minutes after the first if their condition 
does not improve. (Document in patient notes that this advice has been given). 

  Patient (or representative where appropriate) must be counselled on change of device 

and trained on how to use the new device.  

  For children - ensure an allergy action plan is in place and to ensure patient/carer can 

follow: https://www.bsaci.org/about/download-paediatric-allergy-action-plans  

  Any proposed changes to be made by CCG Pharmacist, will need to be approved by a 

Senior doctor at the practice. 

 
 Mrs Mary Hassell 

16 March 2020 

Page 4 



The Practice has also continued to liaise closely with the CQC and has co-operate fully with its investigation. 
The Practice will also provide the CQC with a copy of this response. 

The Practice is firmly of the view that the actions undertaken thus far and those proposed moving forward 
will serve to minimise the likelihood of any similar unfortunate incidents occurring in the future and will 
also enable them to provide an enhanced service to any patients requiring anaphylaxis management. There 
have been valuable lessons learned from the tragic death of Ms Turay-Thomas. 

Please do not hesitate to contact us if you have any queries. 

Yours faithfully 

RadcliffesLeBrasseur 

RadcliffesLeBrasseur

Related reports

Other reports by Mary Hassell

See all →

More reports categorised “Emergency Services related deaths”

See all →

Track Emergency Services related deaths

See every Prevention of Future Deaths report matching Emergency Services related deaths, and how often a new one appears.

What would an alert for this have sent me? Search the full text

Free to try — the preview shows the real matches and how many arrived in the last 12 months. Your first email alert is free.

These reports are published by the Chief Coroner's office at judiciary.uk and are © Crown copyright. The text here is reproduced from the published PDF so it can be searched. If something on this page is wrong, or you are a person named in it and want it reviewed, email drcjar@gmail.com and we will act promptly.